aScope™ 4 Broncho Large 5.8/2.8 478001000Recalls & reported problems

Ambu A/S · Flexible Endoscope · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 05707480135530), received between Jun 4, 2021 and Sep 2, 2026 — classified as 12 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 4 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component10
  • Break6
  • Entrapment of Device2
  • Defective Device1
  • Material Fragmentation1
  • Mechanical Jam1
  • Mechanical Problem1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Foreign Body In Patient4
  • Unintended Extubation4
  • Airway Obstruction2
  • Cardiac Arrest2
  • Anxiety1
  • Hypoxia1
  • Insufficient Information1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjurySep 2, 2026Detachment of Device or Device Component

    During a bronchoscopy procedure using an aScope 4 Broncho Large 5.8/2.8, the distal tip of the device detached. A second bronchoscopy was performed using a reusable bronchoscope to retrieve the detached part with biopsy forceps. The involved device has been retained by the hospital for investigation if required. Patient outcome: Examination prolonged, with no serious consequences for the patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50570454 ↗
  • InjurySep 2, 2026Detachment of Device or Device Component

    During a bronchoscopy procedure using an aScope 4 Broncho Large 5.8/2.8, the distal tip of the device detached. A second bronchoscopy was performed using a reusable bronchoscope to retrieve the detached part with biopsy forceps. The involved device has been retained by the hospital for investigation if required. Patient outcome: Examination prolonged, with no serious consequences for the patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50564938 ↗
  • InjuryAug 7, 2026Detachment of Device or Device Component

    According to the reported customer complaint:Parts of the bronchoscope have come loose inside the tube. The client informed us by email; sales rep contacted them by phone.The bronchoscope will be picked up on Monday, and sales rep will speak with the senior physician in person. The patient was intubated with an 8.0 tube and then underwent bronchoscopy. During the procedure, the thoracic surgeon noticed that the bronchoscope "got stuck" in the tube. Upon withdrawing it, the thoracic surgeon discovered that the tip was still lodged in the tube. The patient was then reintubated; he was not harmed.The materials were forwarded to the purchasing department. The bronchoscope was inserted via a duckbill cuff with integrated suction and broke off inside the tube during the procedure. Fortunately, the patient was not harmed. The incident was detected due to unstable ventilation, allowing the patient to be reintubated immediately.

    Patient problems: Unintended Extubation

    MAUDE report 50174412 ↗
  • InjuryAug 7, 2026Detachment of Device or Device Component

    According to the reported customer complaint: Parts of the bronchoscope have come loose inside the tube. The client informed us by email; sales rep contacted them by phone. The bronchoscope will be picked up on Monday, and sales rep will speak with the senior physician in person. The patient was intubated with an 8.0 tube and then underwent bronchoscopy. During the procedure, the thoracic surgeon noticed that the bronchoscope "got stuck" in the tube. Upon withdrawing it, the thoracic surgeon discovered that the tip was still lodged in the tube. The patient was then reintubated; he was not harmed.The materials were forwarded to the purchasing department. The bronchoscope was inserted via a duckbill cuff with integrated suction and broke off inside the tube during the procedure. Fortunately, the patient was not harmed. The incident was detected due to unstable ventilation, allowing the patient to be reintubated immediately.

    Patient problems: Unintended Extubation

    MAUDE report 50172087 ↗
  • InjuryFeb 10, 2026Detachment of Device or Device Component

    AS PER THE CUSTOMER, PATIENT WAS ENDOSCOPED FEW TIMES ON THIS OCCASION THERE WAS A PIECE OF PLASTIC IN PATIENTS BRONCHUS WHICH THEY REMOVED, IT LOOKED LIKE A PIECE OF OUR SCOPE. THE DOCTOR DISCOVERED A WHITE CIRCULAR PIECE OF PLASTIC IN PATIENT`S TRACHEA WHEN HE WAS DOING A FOLLOW-UP BRONCHOSCOPY, SAME SCOPE HAD BEEN USED ON THE SAME PATIENT FEW HOURS BEFORE. THE PIECE OF PLASTIC WAS WHITE AND LOOKED LIKE A PIECE FROM THE END OF THE SCOPE. UNFORTUNATELY THE SCOPE AS WELL AS THE PIECE WAS THROWN AWAY AND CUSTOMER DOES NOT KNOW THE LOT NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24309430 ↗
  • InjuryFeb 10, 2026Detachment of Device or Device Component

    AS PER THE CUSTOMER, PATIENT WAS ENDOSCOPED FEW TIMES ON THIS OCCASION THERE WAS A PIECE OF PLASTIC IN PATIENTS' BRONCHUS WHICH THEY REMOVED, IT LOOKED LIKE A PIECE OF OUR SCOPE. THE DOCTOR DISCOVERED A WHITE CIRCULAR PIECE OF PLASTIC IN PATIENT`S TRACHEA WHEN HE WAS DOING A FOLLOW-UP BRONCHOSCOPY, SAME SCOPE HAD BEEN USED ON THE SAME PATIENT FEW HOURS BEFORE. PIECE OF PLASTIC WAS WHITE AND LOOKED LIKE PIECE FROM THE END OF THE SCOPE. UNFORTUNATELY, SCOPE AS WELL AS THE PIECE WAS THROWN AWAY AND CUSTOMER DOES NOT KNOW THE LOT NUMBER EITHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24301740 ↗
  • MalfunctionNov 19, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    WHILE WITHDRAWING THE BRONCHOSCOPE WITHOUT RESISTANCE, THE DISTAL TIP REMAINED LODGED IN THE ENDOTRACHEAL TUBE. THE INTENSIVE CARE PATIENT SUBSEQUENTLY HAD TO BE REINTUBATED. THE PATIENT WAS DEEPLY SEDATED, DID NOT STRAIN, AND THE BRONCHOSCOPY WAS PERFORMED WITHOUT DIFFICULTY AND WITH GOOD VISIBILITY. PATIENT OUTCOME WAS NOT AFFECTED.

    Patient problems: Unintended Extubation

    MAUDE report 23593929 ↗
  • MalfunctionNov 18, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    WHILE WITHDRAWING THE BRONCHOSCOPE WITHOUT RESISTANCE, THE DISTAL TIP REMAINED LODGED IN THE ENDOTRACHEAL TUBE. THE INTENSIVE CARE PATIENT SUBSEQUENTLY HAD TO BE REINTUBATED. THE PATIENT WAS DEEPLY SEDATED, DID NOT STRAIN, AND THE BRONCHOSCOPY WAS PERFORMED WITHOUT DIFFICULTY AND WITH GOOD VISIBILITY. PATIENT OUTCOME WAS NOT AFFECTED.

    Patient problems: Unintended Extubation

    MAUDE report 23579167 ↗
  • MalfunctionNov 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A BRONCHOSCOPY FOR SECRETION RETENTION, THE DISTAL END OF THE BRONCHOSCOPE WAS TORN OFF. THE ENDOSCOPE WAS DISPOSED OF. ONLY THE DISTAL END WAS RETAINED AND WILL BE SENT IN. PATIENT OUTCOME NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533586 ↗
  • MalfunctionNov 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A BRONCHOSCOPY FOR SECRETION RETENTION, THE DISTAL END OF THE BRONCHOSCOPE WAS TORN OFF. THE ENDOSCOPE WAS DISPOSED OF. ONLY THE DISTAL END WAS RETAINED AND WILL BE SENT IN. PATIENT OUTCOME NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23531395 ↗
  • InjuryAug 18, 2022Break; Mechanical Problem; Mechanical Problem

    AN ASCOPE 4 BRONCHO LARGE WAS USED ON PATIENT WITH A SIZE 7.5 MM INTUBATION TUBE. THE DISTAL END OF THE BRONCHOSCOPE BROKE OFF DURING REMOVAL OF THE SCOPE. THE BROKEN OFF PIECE HAD TO BE REMOVED FROM THE PATIENTS LUNG. PATIENT OUTCOME WAS NOT AFFECTED BY THE INCIDENT AS THE PIECE WAS SUCCESFULLY RETRIEVED BY THE PULMONOLOGISTS ON AN EMERGENCY BASIS.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15253271 ↗
  • MalfunctionAug 18, 2022Mechanical Jam; Mechanical Jam

    AN ASCOPE 4 BRONCHO LARGE WAS USED ON PATIENT WITH A SIZE 7.5 MM INTUBATION TUBE. THE DISTAL END OF THE BRONCHOSCOPE BROKE OFF DURING REMOVAL OF THE SCOPE. THE BROKEN OFF PIECE HAD TO BE REMOVED FROM THE PATIENTS LUNG. PATIENT OUTCOME WAS NOT AFFECTED BY THE INCIDENT AS THE PIECE WAS SUCCESFULLY RETRIEVED BY THE PULMONOLOGISTS ON AN EMERGENCY BASIS.

    Patient problems: Foreign Body In Patient

    MAUDE report 15247249 ↗
  • MalfunctionJan 12, 2022Break; Entrapment of Device

    AN ASCOPE 4 BRONCHO WAS USED IN COMBINATION WITH A SHILEY EVAC ORAL TRACHEAL TUBE. WHEN THE ASCOPE 4 BRONCHO LARGE WAS PULLED OUT, THE TIP OF THE BRONCHOSCOPE TEARED OFF. THE BENDING PIECE GOT STUCK IN THE TUBE AND CAUSED AN AIRWAY OBSTRUCTION. THE USER HAS CONFIRMED THAT THE TRACHEAL TUBE WAS BENT DURING THE PROCEDURE.

    Patient problems: Airway Obstruction; Foreign Body In Patient

    MAUDE report 13238803 ↗
  • MalfunctionJan 11, 2022Material Fragmentation; Material Fragmentation

    AN ASCOPE 4 BRONCHO WAS USED IN COMBINATION WITH A SHILEY EVAC ORAL TRACHEAL TUBE. WHEN THE ASCOPE 4 BRONCHO LARGE WAS PULLED OUT, THE TIP OF THE BRONCHOSCOPE TEARED OFF. THE BENDING PIECE GOT STUCK IN THE TUBE AND CAUSED AN AIRWAY OBSTRUCTION. THE USER HAS CONFIRMED THAT THE TRACHEAL TUBE WAS BENT DURING THE PROCEDURE.

    Patient problems: Airway Obstruction

    MAUDE report 13225304 ↗
  • InjurySep 3, 2021Physical Resistance/Sticking

    AMBU BRONCHOSCOPE SIZE LARGE BECAME STUCK ON XLT DISTAL TRACHEOSTOMY DISPOSABLE INNER CANNULA DURING BRONCHOSCOPY. IT WAS REPORTED THAT THE BRONCHOSCOPE GOT STUCK DURING BRONCHOSCOPY ON THE INNER CANNULA ON XLT DISTAL TRACHEOSTOMY AND THE PATIENT EXPIRED. DURING INVESTIGATION, THE NURSE CLARIFIED THAT THE BRONCHOSCOPE DID NOT CAUSE THE DEATH OF THE PATIENT, AS THE SCOPE WAS SUCCESSFULLY REMOVED WHILE THE PATIENT WAS ALIVE. THE NURSE FURTHER CLARIFIED, THAT IT WAS THE PATIENT'S UNDERLYING CONDITIONS THAT LED TO HIS UNFORTUNATE DEATH. THE DEVICE IS THEREFORE NOT RELATED TO THE DEATH OF THE PATIENT. HOWEVER, THE INCIDENT DID LEAD THE PATIENT TO HAVE AN ANXIETY CARDIAC ARREST.

    Patient problems: Anxiety; Cardiac Arrest

    MAUDE report 12422315 ↗
  • InjurySep 1, 2021Entrapment of Device; Entrapment of Device

    AMBU BRONCHOSCOPE SIZE LARGE BECAME STUCK ON XLT DISTAL TRACHEOSTOMY DISPOSABLE INNTER CANNULA DURING BRONCHOSCOPY. IT WAS REPORTED THAT THE BRONCHOSCOPE GOT STUCK DURING BRONCHOSCOPY ON THE INNER CANNULA ON XLT DISTAL TRACHEOSTOMY AND THE PATIENT EXPIRED. DURING INVESTIGATION, THE NURSE CLARIFIED THAT THE BRONCHOSCOPE DID NOT CAUSE THE DEATH OF THE PATIENT, AS THE SCOPE WAS SUCCESSFULLY REMOVED WHILE THE PATIENT WAS ALIVE. THE NURSE FURTHER CLARIFIED, THAT IT WAS THE PATIENT'S MANY OTHER CONDITIONS THAT LED TO HIS UNFORTUNATE DEATH. THE DEVICE IS THEREFORE NOT RELATED TO THE DEATH OF THE PATIENT. HOWEVER, THE INCIDENT DID LEAD THE PATIENT TO HAVE AN ANXIETY CARDIAC ARREST.

    Patient problems: Cardiac Arrest

    MAUDE report 12406198 ↗
  • InjuryJul 25, 2021Break

    AN ASCOPE 4 BONCHO LARGE WAS USED ON A PATIENT WITH A SIZE 8 INTUBATION TUBE. THE DISTAL END OF THE BRONCHOSCOPE BROKE OFF DURING REMOVAL OF THE SCOPE. THE BROKEN OFF PIECE HAD TO BE REMOVED FROM THE PATIENT'S AIRWAYS. PATIENT OUTCOME WAS NOT AFFECTED BY THIS INCIDENT AS THE PIECE WAS SUCCESSFULLY RETRIEVED AT BEDSIDE.

    Patient problems: Foreign Body In Patient

    MAUDE report 12222997 ↗
  • MalfunctionJul 25, 2021Break; Defective Device

    AN ASCOPE 4 BONCHO LARGE WAS USED ON A PATIENT WITH A SIZE 8 INTUBATION TUBE. THE DISTAL END OF THE BRONCHOSCOPE BROKE OFF DURING REMOVAL OF THE SCOPE. THE BROKEN OFF PIECE HAD TO BE REMOVED FROM THE PATIENT'S AIRWAYS. PATIENT OUTCOME WAS NOT AFFECTED BY THIS INCIDENT AS THE PIECE WAS SUCCESSFULLY RETRIEVED AT BEDSIDE.

    Patient problems: Insufficient Information

    MAUDE report 12222996 ↗
  • InjuryJul 6, 2021Break

    DURING A FIBERSCOPY IN THE ICU THE CONTROL LEVER OF AN ASCOPE 4 BRONCHO BROKE. THE HEALTHCARE PROFESSIONALS IMMEDIATELY TRIED ANOTHER ASCOPE 4 BRONCHO FROM THE SAME BATCH, BUT THE CONTROL LEVER OF THAT SCOPE ALSO BROKE. DUE TO THE PROLONGED PROCEDURE, THE PATIENT DESATURATED. THE ENDOSCOPY WAS THEN PERFORMED WITH A FUNCTIONAL ENDOSCOPE. PATIENT SATURATION LEVEL RETURNED TO NORMAL. NO LONG TERM SERIOUS HEALTH DETERIORATION FOR THE PATIENT WERE REPORTED.

    Patient problems: Hypoxia

    MAUDE report 12119768 ↗
  • InjuryJun 4, 2021Break

    DURING A FIBERSCOPY IN THE ICU THE CONTROL LEVER OF AN ASCOPE 4 BRONCHO BROKE. THE HEALTHCARE PROFESSIONALS IMMEDIATELY TRIED ANOTHER ASCOPE4 BRONCHO FROM THE SAME BATCH, BUT THE CONTROL LEVER OF THAT SCOPE ALSO BROKE. DUE TO THE PROLONGED PROCEDURE, THE PATIENT DESATURATED. THE ENDOSCOPY WSA THEN PERFORMED WITH A FUNCTIONAL ENDOSCOPE. PATIENT SATURATION LEVEL RETURNED TO NORMAL. NO LONG TERM SERIOUS HEALTH DETERIORATION FOR THE PATIENT WERE REPORTED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 11940894 ↗

Other Bronchoscope (flexible or rigid) devices