The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05707480146017), received between Dec 30, 2024 and Sep 16, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Suction Failure2
Patient problems
- Foreign Body In Patient2
- Hypoxia2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 16, 2025Suction Failure; Suction Failure
THE ASCOPE DID NOT SUCTION DURING USE. THIS MALFUNCTION CAUSED A DELAY IN RESOLVING ATELECTASIS IN A PATIENT. TO CONTINUE CARING FOR THE PATIENT, THE MEDICAL TEAM HAD TO URGENTLY REPLACE THE FIBERSCOPE WITH ANOTHER DEVICE OF THE SAME MODEL AND LOT, WHICH WORKED PROPERLY. ADDITIONAL INFORMATION RECEIVED ON 18AUG2025: DURING A BRONCHIAL CLEARANCE FIBROSCOPY, THE FIBROSCOPY EQUIPMENT STOPPED SUCTIONING. A COMPLETE CHECK OF THE SUCTION LINE WAS CARRIED OUT, WITHOUT REVEALING ANY UPSTREAM FAILURE. THE PATIENT, ALREADY EXPERIENCING DESATURATION, REQUIRED A CHANGE OF FIBROSCOPE. A DEVICE FROM THE SAME BATCH WAS THEN USED AND FUNCTIONED CORRECTLY. HOWEVER, THIS INCIDENT LED TO A TEMPORARY WORSENING OF DESATURATION, AS WELL AS A DELAY IN CLEARING THE OBSTRUCTION IN A PATIENT WITH ACUTE LOBAR PNEUMONIA.
Patient problems: Hypoxia
MAUDE report 23064060 ↗InjurySep 16, 2025Suction Failure; Suction Failure
THE ASCOPE DID NOT SUCTION DURING USE. THIS MALFUNCTION CAUSED A DELAY IN RESOLVING ATELECTASIS IN A PATIENT. TO CONTINUE CARING FOR THE PATIENT, THE MEDICAL TEAM HAD TO URGENTLY REPLACE THE FIBERSCOPE WITH ANOTHER DEVICE OF THE SAME MODEL AND LOT, WHICH WORKED PROPERLY. ADDITIONAL INFORMATION RECEIVED ON 18AUG2025: DURING A BRONCHIAL CLEARANCE FIBROSCOPY, THE FIBROSCOPY EQUIPMENT STOPPED SUCTIONING. A COMPLETE CHECK OF THE SUCTION LINE WAS CARRIED OUT, WITHOUT REVEALING ANY UPSTREAM FAILURE. THE PATIENT, ALREADY EXPERIENCING DESATURATION, REQUIRED A CHANGE OF FIBROSCOPE. A DEVICE FROM THE SAME BATCH WAS THEN USED AND FUNCTIONED CORRECTLY. HOWEVER, THIS INCIDENT LED TO A TEMPORARY WORSENING OF DESATURATION, AS WELL AS A DELAY IN CLEARING THE OBSTRUCTION IN A PATIENT WITH ACUTE LOBAR PNEUMONIA.
Patient problems: Hypoxia
MAUDE report 23060420 ↗InjuryDec 31, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
CUSTOMER REPORTED, THAT WHILE PERFORMING AN ALVEOLAR LAVAGE DURING BRONCHIAL FIBROSCOPY, RUBBER OCCLUSIVE MEMBRANE OF THE FIBERSCOPE'S OPERATING CHANNEL RUPTURED AND PART OF IT FELL INTO PATIENT'S BRONCHIAL TREE. THEY WERE NOT ABLE TO RECOVER IT. PATIENT WAS ADMITTED TO INTENSIVE CARE UNIT FOR RESPIRATORY DISTRESS AND UNDERGONE FIBRESCOPY ON THE SAME DAY. PATIENT CONTINUED TO BE HOSPITALIZED IN INTENSIVE CARE AND ACTIONS WERE TAKEN IN THE HOSPITAL TO MANAGE THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 21047001 ↗InjuryDec 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
CUSTOMER REPORTED, THAT WHILE PERFORMING AN ALVEOLAR LAVAGE DURING BRONCHIAL FIBROSCOPY, RUBBER OCCLUSIVE MEMBRANE OF THE FIBERSCOPE'S OPERATING CHANNEL RUPTURED AND PART OF IT FELL INTO PATIENT'S BRONCHIAL TREE. THEY WERE NOT ABLE TO RECOVER IT. PATIENT WAS ADMITTED TO INTENSIVE CARE UNIT FOR RESPIRATORY DISTRESS AND UNDERGONE FIBRESCOPY ON THE SAME DAY. PATIENT CONTINUED TO BE HOSPITALIZED IN INTENSIVE CARE AND ACTIONS WERE TAKEN IN THE HOSPITAL TO MANAGE THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 21033956 ↗
Other Bronchoscope (flexible or rigid) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page