Skip to content

aScope™ 4 Broncho Large Sampler Set 5.8/2.8 478501000Recalls & reported problems

Ambu A/S · aScope 4 Broncho Large Sampler Set · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05707480146017), received between Dec 30, 2024 and Sep 16, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Suction Failure2

Patient problems

  • Foreign Body In Patient2
  • Hypoxia2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2025Suction Failure; Suction Failure

    THE ASCOPE DID NOT SUCTION DURING USE. THIS MALFUNCTION CAUSED A DELAY IN RESOLVING ATELECTASIS IN A PATIENT. TO CONTINUE CARING FOR THE PATIENT, THE MEDICAL TEAM HAD TO URGENTLY REPLACE THE FIBERSCOPE WITH ANOTHER DEVICE OF THE SAME MODEL AND LOT, WHICH WORKED PROPERLY. ADDITIONAL INFORMATION RECEIVED ON 18AUG2025: DURING A BRONCHIAL CLEARANCE FIBROSCOPY, THE FIBROSCOPY EQUIPMENT STOPPED SUCTIONING. A COMPLETE CHECK OF THE SUCTION LINE WAS CARRIED OUT, WITHOUT REVEALING ANY UPSTREAM FAILURE. THE PATIENT, ALREADY EXPERIENCING DESATURATION, REQUIRED A CHANGE OF FIBROSCOPE. A DEVICE FROM THE SAME BATCH WAS THEN USED AND FUNCTIONED CORRECTLY. HOWEVER, THIS INCIDENT LED TO A TEMPORARY WORSENING OF DESATURATION, AS WELL AS A DELAY IN CLEARING THE OBSTRUCTION IN A PATIENT WITH ACUTE LOBAR PNEUMONIA.

    Patient problems: Hypoxia

    MAUDE report 23064060 ↗
  • InjurySep 16, 2025Suction Failure; Suction Failure

    THE ASCOPE DID NOT SUCTION DURING USE. THIS MALFUNCTION CAUSED A DELAY IN RESOLVING ATELECTASIS IN A PATIENT. TO CONTINUE CARING FOR THE PATIENT, THE MEDICAL TEAM HAD TO URGENTLY REPLACE THE FIBERSCOPE WITH ANOTHER DEVICE OF THE SAME MODEL AND LOT, WHICH WORKED PROPERLY. ADDITIONAL INFORMATION RECEIVED ON 18AUG2025: DURING A BRONCHIAL CLEARANCE FIBROSCOPY, THE FIBROSCOPY EQUIPMENT STOPPED SUCTIONING. A COMPLETE CHECK OF THE SUCTION LINE WAS CARRIED OUT, WITHOUT REVEALING ANY UPSTREAM FAILURE. THE PATIENT, ALREADY EXPERIENCING DESATURATION, REQUIRED A CHANGE OF FIBROSCOPE. A DEVICE FROM THE SAME BATCH WAS THEN USED AND FUNCTIONED CORRECTLY. HOWEVER, THIS INCIDENT LED TO A TEMPORARY WORSENING OF DESATURATION, AS WELL AS A DELAY IN CLEARING THE OBSTRUCTION IN A PATIENT WITH ACUTE LOBAR PNEUMONIA.

    Patient problems: Hypoxia

    MAUDE report 23060420 ↗
  • InjuryDec 31, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER REPORTED, THAT WHILE PERFORMING AN ALVEOLAR LAVAGE DURING BRONCHIAL FIBROSCOPY, RUBBER OCCLUSIVE MEMBRANE OF THE FIBERSCOPE'S OPERATING CHANNEL RUPTURED AND PART OF IT FELL INTO PATIENT'S BRONCHIAL TREE. THEY WERE NOT ABLE TO RECOVER IT. PATIENT WAS ADMITTED TO INTENSIVE CARE UNIT FOR RESPIRATORY DISTRESS AND UNDERGONE FIBRESCOPY ON THE SAME DAY. PATIENT CONTINUED TO BE HOSPITALIZED IN INTENSIVE CARE AND ACTIONS WERE TAKEN IN THE HOSPITAL TO MANAGE THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21047001 ↗
  • InjuryDec 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER REPORTED, THAT WHILE PERFORMING AN ALVEOLAR LAVAGE DURING BRONCHIAL FIBROSCOPY, RUBBER OCCLUSIVE MEMBRANE OF THE FIBERSCOPE'S OPERATING CHANNEL RUPTURED AND PART OF IT FELL INTO PATIENT'S BRONCHIAL TREE. THEY WERE NOT ABLE TO RECOVER IT. PATIENT WAS ADMITTED TO INTENSIVE CARE UNIT FOR RESPIRATORY DISTRESS AND UNDERGONE FIBRESCOPY ON THE SAME DAY. PATIENT CONTINUED TO BE HOSPITALIZED IN INTENSIVE CARE AND ACTIONS WERE TAKEN IN THE HOSPITAL TO MANAGE THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21033956 ↗

Other Bronchoscope (flexible or rigid) devices

All Bronchoscope (flexible or rigid) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page