Ambu® aScope™ 5 Broncho 2.7/1.2 624001000USRecalls & reported problems
Ambu A/S · Single-Use Flexible Endoscope · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05707480157280), received between Nov 22, 2024 and Dec 8, 2025 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Material Split, Cut or Torn2
- Physical Resistance/Sticking2
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryDec 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
DURING A VATS SURGICAL PROCEDURE THAT OUR A5 2.7MM ENDOSCOPE DISTAL TIP BROKE OFF INTO THE PATIENTS ENDOTRACHEAL TUBE. ANESTHESIA TEAM WENT IN AND RETRIEVED THE PIECE THAT HAD BROKEN OFF. PATIENT CONDITION WAS NOT AFFECTED BY THE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23748072 ↗InjuryDec 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
DURING A VATS SURGICAL PROCEDURE THAT OUR A5 2.7MM ENDOSCOPE DISTAL TIP BROKE OFF INTO THE PATIENTS ENDOTRACHEAL TUBE. ANESTHESIA TEAM WENT IN AND RETRIEVED THE PIECE THAT HAD BROKEN OFF. PATIENT CONDITION WAS NOT AFFECTED BY THE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23740214 ↗MalfunctionJan 7, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
A CLINICIAN ACCUSTOMED TO PIERCING THE WHITE RUBBER COVER OF THE WORKING CHANNEL OF ASCOPE 5 BRONCHO WITH A SYRINGE, IN ORDER TO ACCESS THE WORKING CHANNEL, REPORTED THAT A PIECE OF WHITE RUBBER DISLODGED FROM ASCOPE 5 BRONCHO AND FLUSHED DOWN IN THE WORKING CHANNEL DURING BRONCHOSCOPY. THE INCIDENT INVOLVED A PATIENT, BUT DID NOT LEAD TO PATIENT HARM, AS THE USER WAS ABLE TO SUCTION THE RUBBER PIECE UP INTO THE SPECIMEN TRAP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21092218 ↗MalfunctionJan 7, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
A CLINICIAN ACCUSTOMED TO PIERCING THE WHITE RUBBER COVER OF THE WORKING CHANNEL OF ASCOPE 5 BRONCHO WITH A SYRINGE, IN ORDER TO ACCESS THE WORKING CHANNEL, REPORTED THAT A PIECE OF WHITE RUBBER DISLODGED FROM ASCOPE 5 BRONCHO AND FLUSHED DOWN IN THE WORKING CHANNEL DURING BRONCHOSCOPY. THE INCIDENT INVOLVED A PATIENT, BUT DID NOT LEAD TO PATIENT HARM, AS THE USER WAS ABLE TO SUCTION THE RUBBER PIECE UP INTO THE SPECIMEN TRAP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21088770 ↗MalfunctionNov 22, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED BY THE CUSTOMER THAT TWO ASCOPE 5 BRONCHO ULTRA-THIN SCOPE WOULD NOT ALLOW PASSAGE OF ERBE CRYOPROBE THROUGH WORKING CHANNEL. DURING A ROUTINE CRYO PROCEDURE WITH OUR A5 ULTRA-THIN SCOPE THE STAFF HAD NO PROBLEM PASSING THE PROBE THOUGH THE WORKING CHANNEL UNTIL THE END OF THE CHANNEL WHERE RESISTANCE WAS FELT, AND THE PROBE WOULD NOT PASS THROUGH THE END OF SCOPE RESULTING IN KINKING THE PROBE. THE STAFF THEN SWITCHED TO A NEW A5 ULTRA-THIN SCOPE FROM A DIFFERENT BOX WITH THE SAME LOT NUMBER AS WELL AS A NEW CRYOPROBE. THE PHYSICIAN ALSO REMOVED THE SCOPE FROM THE PATIENT MAKING SURE THE SCOPE WAS COMPLETELY STRAIGHT AND THE PROBLEM STILL OCCURRED AGAIN. AFTER THE SECOND TIME THE STAFF SWITCHED TO A REUSABLE OLYMPUS SCOPE, AND THEY HAD NO PROBLEM PASSING THE CRYOPROBE. THE SCOPES CAME FROM TWO DIFFERENT BOXES/ AREAS BUT HAD THE SAME LOT NUMBER. THE PROCEDURE IS CATEGORIZED AS OFF-LABEL, SINCE ASCOPE 5 BRONCHO WAS USED ON A PEDIATRIC PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20760784 ↗MalfunctionNov 22, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED BY THR CUSTOMER THAT TWO SCOPE 5 BRONCHO ULTRA-THIN SCOPE WOULD NOT ALLOW PASSAGE OF ERBE CRYOPROBE THROUGH WORKING CHANNEL. DURING A ROUTINE CRYO PROCEDURE WITH OUR A5 ULTRA-THIN SCOPE THE STAFF HAD NO PROBLEM PASSING THE PROBE THOUGH THE WORKING CHANNEL UNTIL THE END OF THE CHANNEL WHERE RESISTANCE WAS FELT, AND THE PROBE WOULD NOT PASS THROUGH THE END OF SCOPE RESULTING IN KINKING THE PROBE. THE STAFF THEN SWITCHED TO A NEW A5 ULTRA-THIN SCOPE FROM A DIFFERENT BOX WITH THE SAME LOT NUMBER AS WELL AS A NEW CRYOPROBE. THE PHYSICIAN ALSO REMOVED THE SCOPE FROM THE PATIENT MAKING SURE THE SCOPE WAS COMPLETELY STRAIGHT AND THE PROBLEM STILL OCCURRED AGAIN. AFTER THE SECOND TIME THE STAFF SWITCHED TO A REUSABLE OLYMPUS SCOPE, AND THEY HAD NO PROBLEM PASSING THE CRYOPROBE. THE SCOPES CAME FROM TWO DIFFERENT BOXES/ AREAS BUT HAD THE SAME LOT NUMBER. THE PROCEDURE IS CATEGORIZED AS OFF-LABEL, SINCE A SCOPE 5 BRONCHO WAS USED ON A PEDIATRIC PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20757475 ↗
Other Bronchoscope (flexible or rigid) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page