Ambu® aScope™ 5 Broncho 2.7/1.2 624001000USRecalls & reported problems

Ambu A/S · Single-Use Flexible Endoscope · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05707480157280), received between Nov 22, 2024 and Dec 8, 2025 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Material Split, Cut or Torn2
  • Physical Resistance/Sticking2

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A VATS SURGICAL PROCEDURE THAT OUR A5 2.7MM ENDOSCOPE DISTAL TIP BROKE OFF INTO THE PATIENTS ENDOTRACHEAL TUBE. ANESTHESIA TEAM WENT IN AND RETRIEVED THE PIECE THAT HAD BROKEN OFF. PATIENT CONDITION WAS NOT AFFECTED BY THE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23748072 ↗
  • InjuryDec 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A VATS SURGICAL PROCEDURE THAT OUR A5 2.7MM ENDOSCOPE DISTAL TIP BROKE OFF INTO THE PATIENTS ENDOTRACHEAL TUBE. ANESTHESIA TEAM WENT IN AND RETRIEVED THE PIECE THAT HAD BROKEN OFF. PATIENT CONDITION WAS NOT AFFECTED BY THE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23740214 ↗
  • MalfunctionJan 7, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A CLINICIAN ACCUSTOMED TO PIERCING THE WHITE RUBBER COVER OF THE WORKING CHANNEL OF ASCOPE 5 BRONCHO WITH A SYRINGE, IN ORDER TO ACCESS THE WORKING CHANNEL, REPORTED THAT A PIECE OF WHITE RUBBER DISLODGED FROM ASCOPE 5 BRONCHO AND FLUSHED DOWN IN THE WORKING CHANNEL DURING BRONCHOSCOPY. THE INCIDENT INVOLVED A PATIENT, BUT DID NOT LEAD TO PATIENT HARM, AS THE USER WAS ABLE TO SUCTION THE RUBBER PIECE UP INTO THE SPECIMEN TRAP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21092218 ↗
  • MalfunctionJan 7, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A CLINICIAN ACCUSTOMED TO PIERCING THE WHITE RUBBER COVER OF THE WORKING CHANNEL OF ASCOPE 5 BRONCHO WITH A SYRINGE, IN ORDER TO ACCESS THE WORKING CHANNEL, REPORTED THAT A PIECE OF WHITE RUBBER DISLODGED FROM ASCOPE 5 BRONCHO AND FLUSHED DOWN IN THE WORKING CHANNEL DURING BRONCHOSCOPY. THE INCIDENT INVOLVED A PATIENT, BUT DID NOT LEAD TO PATIENT HARM, AS THE USER WAS ABLE TO SUCTION THE RUBBER PIECE UP INTO THE SPECIMEN TRAP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21088770 ↗
  • MalfunctionNov 22, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED BY THE CUSTOMER THAT TWO ASCOPE 5 BRONCHO ULTRA-THIN SCOPE WOULD NOT ALLOW PASSAGE OF ERBE CRYOPROBE THROUGH WORKING CHANNEL. DURING A ROUTINE CRYO PROCEDURE WITH OUR A5 ULTRA-THIN SCOPE THE STAFF HAD NO PROBLEM PASSING THE PROBE THOUGH THE WORKING CHANNEL UNTIL THE END OF THE CHANNEL WHERE RESISTANCE WAS FELT, AND THE PROBE WOULD NOT PASS THROUGH THE END OF SCOPE RESULTING IN KINKING THE PROBE. THE STAFF THEN SWITCHED TO A NEW A5 ULTRA-THIN SCOPE FROM A DIFFERENT BOX WITH THE SAME LOT NUMBER AS WELL AS A NEW CRYOPROBE. THE PHYSICIAN ALSO REMOVED THE SCOPE FROM THE PATIENT MAKING SURE THE SCOPE WAS COMPLETELY STRAIGHT AND THE PROBLEM STILL OCCURRED AGAIN. AFTER THE SECOND TIME THE STAFF SWITCHED TO A REUSABLE OLYMPUS SCOPE, AND THEY HAD NO PROBLEM PASSING THE CRYOPROBE. THE SCOPES CAME FROM TWO DIFFERENT BOXES/ AREAS BUT HAD THE SAME LOT NUMBER. THE PROCEDURE IS CATEGORIZED AS OFF-LABEL, SINCE ASCOPE 5 BRONCHO WAS USED ON A PEDIATRIC PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20760784 ↗
  • MalfunctionNov 22, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED BY THR CUSTOMER THAT TWO SCOPE 5 BRONCHO ULTRA-THIN SCOPE WOULD NOT ALLOW PASSAGE OF ERBE CRYOPROBE THROUGH WORKING CHANNEL. DURING A ROUTINE CRYO PROCEDURE WITH OUR A5 ULTRA-THIN SCOPE THE STAFF HAD NO PROBLEM PASSING THE PROBE THOUGH THE WORKING CHANNEL UNTIL THE END OF THE CHANNEL WHERE RESISTANCE WAS FELT, AND THE PROBE WOULD NOT PASS THROUGH THE END OF SCOPE RESULTING IN KINKING THE PROBE. THE STAFF THEN SWITCHED TO A NEW A5 ULTRA-THIN SCOPE FROM A DIFFERENT BOX WITH THE SAME LOT NUMBER AS WELL AS A NEW CRYOPROBE. THE PHYSICIAN ALSO REMOVED THE SCOPE FROM THE PATIENT MAKING SURE THE SCOPE WAS COMPLETELY STRAIGHT AND THE PROBLEM STILL OCCURRED AGAIN. AFTER THE SECOND TIME THE STAFF SWITCHED TO A REUSABLE OLYMPUS SCOPE, AND THEY HAD NO PROBLEM PASSING THE CRYOPROBE. THE SCOPES CAME FROM TWO DIFFERENT BOXES/ AREAS BUT HAD THE SAME LOT NUMBER. THE PROCEDURE IS CATEGORIZED AS OFF-LABEL, SINCE A SCOPE 5 BRONCHO WAS USED ON A PEDIATRIC PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20757475 ↗

Other Bronchoscope (flexible or rigid) devices