aScope™ 4 Broncho Slim 3.8/1.2 476101000Recalls & reported problems
Ambu A/S · aScope 4 Broncho Slim · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05707480157792), received between Apr 29, 2025 and Apr 30, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
DURING PROCEDURE INVOLVING A PATIENT, THE TIP OF ASCOPE 4 BRONCHO SLIM BROKE OFF INSIDE OF DOUBLE LUMEN TUBE. THE PATIENT OUTCOME WAS NOT AFFECTED. THE USED PROVIDED CLARIFICATION, THAT THE SCOPE GOT STUCK IN THE BRONCHIAL LUMEN, BUT IT DID NOT LEAD TO ANY HARM FOR THE PATIENT AND NO PROLONGED VENTILATION ISSUES WAS OBSERVED. THE CUSTOMER FURTHER CLARIFIED THAT THE REASON THE DISTAL TIP BROKE OFF INSIDE THE TUBE WAS BECAUSE THE SCOPE GOT STUCK IN THE TUBE. WHEN THE SCOPE BROKE INSIDE THE TUBE, THE USER SWAPPED OUT OF THE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21936402 ↗MalfunctionApr 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
DURING PROCEDURE INVOLVING A PATIENT, THE TIP OF ASCOPE 4 BRONCHO SLIM BROKE OFF INSIDE OF DOUBLE LUMEN TUBE. THE PATIENT OUTCOME WAS NOT AFFECTED. THE USER PROVIDED CLARIFICATION, THAT THE SCOPE GOT STUCK IN THE BRONCHIAL LUMEN, BUT IT DID NOT LEAD TO ANY HARM FOR THE PATIENT AND NO PROLONGED VENTILATION ISSUES WAS OBSERVED. THE CUSTOMER FURTHER CLARIFIED THAT THE REASON THE DISTAL TIP BROKE OFF INSIDE THE TUBE WAS BECAUSE THE SCOPE GOT STUCK IN THE TUBE. WHEN THE SCOPE BROKE INSIDE THE TUBE, THE USER SWAPPED OUT THE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925195 ↗
Other Bronchoscope (flexible or rigid) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page