Skip to content

GN Hearing 21655200Recalls & reported problems

Gn Hearing A/S · EM,NC-HA-UPNG8 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05708296213375), received between Jan 11, 2024 and Jul 29, 2026 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • Detachment of Device or Device Component1
  • Inadequacy of Device Shape and/or Size1
  • Material Disintegration1
  • Patient-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Ear Pain1
  • Ear infection1
  • Infectious Otitis Externa1
  • Unspecified Ear or Labyrinth Problem1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2026Detachment of Device or Device Component

    It was reported that a wax filter grommet fell out of an encased receiver. The grommet fell out while the end-user was handling the device, it is unknown what the user was doing with the device at the time of the event. The wax filter grommet fell off outside of the user's ear, and hence no harm resulted from the event. No further follow up is expected.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50068614 ↗
  • MalfunctionJun 26, 2026Material Disintegration

    IT HAS BEEN REPORTED THAT A HEARING AID USER HAD MATERIAL PEELING ON AN ENCASED EARMOLD AFTER 2 WEEKS OF USE. THE USER REPORTED THAT SHE WAKES UP IN THE MORNING FEELING LIKE SHE HAS DEBRIS IN HER EAR. IT HAS NOT BEEN CONFIRMED WHETHER THE USER DOES HAVE MATERIAL FROM THE SHELL IN HER EAR, AND SHE HAS NOT SOUGHT MEDICAL ATTENTION. THE USER STOPPED WEARING THE EARMOULD AND A REMAKE WAS ORDERED. IT WAS REPORTED THAT THERE WAS NO HARM TO THE USER. NO FURTHER FOLLOW UP IS AVAILABLE OR EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25633454 ↗
  • MalfunctionFeb 25, 2026Fracture

    ESTIMATED DATE OF INCIDENT (B)(6) 2026. IT WAS REPORTED THAT THE SHELL OF A RECEIVER-IN-EAR EARMOULD BROKE INSIDE IN THE USER'S EAR, AND A PIECE WAS STUCK IN THE CANAL. THE USER WAS INSTRUCTED BY THE PROVIDER TO SEE A MEDICAL DOCTOR/EAR-NOSE-THROAT DOCTOR, AFTER THE PROVIDER COULD NOT REMOVE IT. NO HARM HAS BEEN REPORTED FOLLOWING THE INCIDENT. FURTHER FOLLOW UP IS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24441596 ↗
  • InjuryOct 18, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    ESTIMATED DATE OF INCIDENT (B)(6) 2024. IT WAS REPORTED THAT THE USER WAS FIT WITH HEARING AID IN (B)(6) AT FOLLOW UP APPOINTMENT (B)(6) 2024 SHE REPORTED THAT SHE HAD NOT BEEN ABLE TO WEAR DEVICES DUE TO DISCOMFORT, PAIN AND SUSPECTED THERE MAY BE AN INFECTION AS SHE GOT FLUID FROM THE EAR. THE USER WENT TO SEE HER GENERAL PRACTITIONER WHO SUSPECTED EXTERMAL EAR INFECTION AND PRESCRIBED BROAD SPECTRUM DROPS THAT WOULD ADRESS BACTERIAL AND FUNGAL. NO FURTHER FOLLOW UP IS EXPECTED.

    Patient problems: Infectious Otitis Externa

    MAUDE report 20479762 ↗
  • MalfunctionOct 11, 2024Fracture; Fracture

    ESTIMATED DATE OF THE EVENT: 11SEP2024. IT WAS REPORTED THAT AN NCASED EARMOLD BROKE. NO HARM TO THE USER WAS REPORTED. FURTHER FOLLOW UP IS EXPECTED. 06NOV2024: NO FURTHER FOLLOW UP INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20425646 ↗
  • MalfunctionSep 20, 2024Fracture; Fracture

    ESTIMATED DATE OF INCIDENT (B)(6)2024 IT WAS REPORTED THAT AN NCASED EARMOULD HAS CRACKED. NO OBJECT REQUIRED REMOVAL, AND USER STOPPED USING THE DEVICE AFTER NOTICING THE CRACK, WHEN HE SAID HE DIDN'T HEAR AS WELL FOR A FEW DAYS. THERE WAS NO HARM TO THE USER FOLLOWING THE INCIDENT. NO FURTHER FOLLOW UP IS AVAILABLE OR EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20264795 ↗
  • MalfunctionMay 8, 2024Fracture; Fracture

    ESTIMATED DATE OF INCIDENT (B)(6) 2024 IT WAS REPORTED THAT THE EARMOULD BROKE. THERE IS NO REPORT OF ANY HARM TO THE USER. NO FURTHER INFORMATION IS AVAILABLE, OR EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19266112 ↗
  • InjuryJan 11, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    DATE OF INCIDENT: (B)(6) 2023 IT WAS REPORTED THAT THE USER EXPERIENCES PAIN AND SWELLING ON ONE SIDE, IN THE EXTERNAL EAR CANAL. SYMOTOMS REPORTED WERE RED, SWOLLEN, PAINFUL. REPORTED THAT WITH REMOVAL OF THE DEVICE IMPROVED ISSUE. BUT ALSO REPORTED THAT THE ISSUE STARTED AFTER A BOUT OF LARYNGITIS. FIT WITH EARMOLD ON (B)(6) 2023 SYMPTOMS WITH THE EAR BEGAN ON SATURDAY ON (B)(6) 2023 REPORTED LARYNGITIS ONSET ON (B)(6) 2023 USER IS REPORTABLY ALLERGIC TO SURGICAL GLUE. THE USER WAS SEEN IN URGENT CARE TO ADDRESS CONCERNS WITH THE EAR, AND WAS PRESCRIBED AMOXICILLIN A 10-DAY COURSE 2/DAY AM AND PM. NO FURTHER FOLLOW UP HAS BEEN RECEIVED AND IS EXPECTED.

    Patient problems: Ear Pain; Ear infection; Unspecified Ear or Labyrinth Problem

    MAUDE report 18491010 ↗

Other Hearing aid, air conduction with wireless technology devices

All Hearing aid, air conduction with wireless technology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page