Skip to content

ORTHOPANTOMOGRAPH OP 3D Pro OP300-1 (REF 901002)Recalls & reported problems

PaloDEx Group Oy · Class 2

All models of ORTHOPANTOMOGRAPH OP 3D Pro (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Oct 28, 2021Class IIOpen, Classified

    It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

    Z-2413-2024 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

Other X-Ray, Tomography, Computed, Dental devices

All X-Ray, Tomography, Computed, Dental devices →