Skip to content

ORTHOPANTOMOGRAPH OP 3D Pro recallZ-2413-2024 · Palodex Group Oy · initiated Oct 28, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Palodex Group Oy or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Palodex Group Oy is preparing a Class IlIl field correction to update the firmware of ORTHOPANTOMOGRAPH OP 3D Pro of the 27 defective units and these devices will be retested for Pan ADC reproducibility. All affected persons will be notified of this defect in accordance with 21 CFR 806.10.

Recall details

Recall number
Z-2413-2024
Status
Open, Classified
Initiated by the company
Oct 28, 2021
Posted by the FDA
Aug 12, 2024
Recalling firm
Palodex Group Oy
Quantity
103
Product code
OAS
FDA submissions
K163423
Distribution
US Nationwide-Worldwide

Product description

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

Code information (models, lots, serial numbers)

DI: 06430035872725, 06430035872732, 06430035872749, 06430035872756, 06430035872763, 06430035872770

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page