Skip to content

Avenir 10CFR15030Recalls & reported problems

Shanghai Wallaby Medical Technologies Co., Inc. · Embolic Coil · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 06970908211355), received between May 20, 2024 and May 20, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Stretched” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2024Stretched; Stretched

    A HEADWAY 167 DUO WAS NAVIGATED UNDER HIGH TORTUOSITY, PHYSICIAN ATTEMPTED TO DEPLOY A AVENIR COIL((B)(6)). THERE WAS HIGH FLOW, CAUSING IT TO BE QUITE TURBULENT. AS THE COIL WAS DEPLOYED, FLOW WOULD CATCH THE COIL AND CAUSE IT TO COMPACT INTO ONE CORNER OF THE FISTULA CAVITY. AS THE PHYSICIAN CONTINUED TO PUSH COIL OUT, IT WOULD CONTINUE TO BECOME COMPRESSED INTO ONE CORNER OF THE FISTULA. THE PHYSICIAN THEN PROCEEDED TO RETRIEVE THE COIL. AFTER RETRIEVING/ PUSHING COIL OUT A FEW TIMES, THE COIL BECOME "BUNCHED" UP ON ITSELF, AND THE COIL STRETCHED UPON FINAL RETRIEVING/RETRACT. THE PHYSICIAN REQUIRED THE USE STENT RETRIEVER TO PULL BACK THE STRETCHED COIL. THERE WAS ON PATIENT HARM REPORTED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19356502 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page