Rhythm Healthcare P2Recalls & reported problems
Qingdao Kingon Medical Science and Technology Co., Ltd · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 06971882890017), received between Jun 8, 2026 and Jun 8, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Defective Alarm” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Alarm1
- Gas/Air Leak1
- Undercorrection1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2026Defective Alarm; Gas/Air Leak; Undercorrection
TECHNICAL AUDIT OF RHYTHM P2 POC PLATFORM CONFIRMS A SYSTEMIC 'SILENT FAILURE' MODE WHERE DEVICES DELIVER SUB-THERAPEUTIC OXYGEN (DOCUMENTED AS LOW AS 42% O2) WITHOUT TRIGGERING MANDATORY SAFETY ALARMS.FORENSIC TESTING IDENTIFIED AN 'OSCILLATION LOOPHOLE' IN THE ALARM LOGIC: THE DEVICE REQUIRES 600S OF CONTINUOUS LOW O2 TO ALARM, BUT ERRATIC INTERNAL SENSOR READINGS (FLUCTUATING 40%-90%) RESET THE TIMER, PREVENTING ISO 80601-2-69 ALERTS.NIST-TRACEABLE TESTING CONFIRMS A 10-30% POSITIVE BIAS IN ONBOARD DIAGNOSTICS. MANUFACTURER'S (B)(6) 2023 FORWARD ONLY FIRMWARE FIX IS DOCUMENTED AS INEFFECTIVE IN 2024 PRODUCTION UNITS.THIS DEFECT PRESENTS A RISK OF SERIOUS INJURY OR DEATH TO OXYGEN-DEPENDENT PATIENTS DUE TO LOSS OF DETECTABILITY OF HAZARDOUS CONDITIONS.P2#1 (SN# (B)(6)) ONBOARD READING: VARIABLE (52-71%); ANALYZER READING: 47.9%; NO LOW O2 WARNING! P2#2 (SN# (B)(6)) ONBOARD READING: VARIABLE (62-91%); ANALYZER READING: 60%; NO LOW O2 WARNING! P2#3 (SN# (B)(6)) ONBOARD READING: VARIABLE (66-91%); ANALYZER READING: 78.8%; LOW O2 WARNING P2#4 (SN# (B)(6)) ONBOARD READING: VARIABLE (83-91%); ANALYZER READING: 74.2%; NO LOW O2 WARNING! P2#5 (SN# (B)(6)) ONBOARD READING: VARIABLE (75-92
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 25438418 ↗
Other Generator, Oxygen, Portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.