0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 06972855530022), received between Mar 3, 2025 and Nov 19, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Arcing2

Patient problems

  • Unspecified Respiratory Problem3
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2025Arcing; Arcing

    ON (B)(6), 2025 THIS FACILITY HAD A MALFUNCTION IN A LEASED OXYGEN CONCENTRATOR, MANUFACTURED BY DIRECT SUPPLY MANUFACTURING INC., ATTENDANT 5L OXYGEN CONCENTRATOR, MODEL: A5LC-1, THE CONCENTRATOR BEGAN TO PRODUCE SMOKE, FILLING ROOM N112 AND CORRIDOR WITH SMOKE SETTING OFF THE FIRE ALARM. THE STAFF REMOVED THE CONCENTRATOR FROM THE BUILDING TO THE REAR SERVICE PARKING LOT. AT THAT TIME IT STARTED PRODUCING FLAMES AND THE STAFF EXTINGUISHED THE MACHINE WITH A FIRE EXTINGUISHER. 2 RESIDENTS AND 8 STAFF WERE TRANSPORTED TO THE HOSPITAL OUT OF PRECAUTION AND RELEASED.

    Patient problems: Insufficient Information

    MAUDE report 23590460 ↗
  • MalfunctionSep 13, 2025Arcing; Arcing

    UNIT WAS SMOKING. CAUGHT ON FIRE. PUT OUT WITH A FIRE HOSE. PUT IN THE PARKING LOT. AT (B)(6) 2025 AT 11:10AM, CALLED TO REPORT AN INCIDENT THAT ALLEGEDLY INVOLVED THE FOLLOWING PRODUCT OR DEVICE: OXYGEN CONCENTRATOR. REPORTED THAT THE FOLLOWING OCCURRED AT (B)(6) 2025 AT 9:05AM: ALLEGEDLY THE POWER SWITCH WAS TURNED ON AND INSIDE FIRE STARTED. UNIT WAS SMOKING AND THEY PUT IT OUT WITH WATER HOSE AND HAVE PUT IT OUTSIDE IN THE PARKING LOT AWAY FROM THE BUILDING. THERE WERE NO INJURIES ACCORDING TO REPORTER. THE PRODUCT HAS BEEN TAKEN OUT OF USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23046007 ↗
  • MalfunctionMar 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    A CONCENTRATOR CAUGHT FIRE IN A MANNER CONSISTENT WITH HOUR COUNTER FAILURE. NO INJURIES OCCURRED.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 21501583 ↗
  • MalfunctionMar 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    A CONCENTRATOR CAUGHT FIRE IN A MANNER CONSISTENT WITH HOUR COUNTER FAILURE. NO INJURIES OCCURRED.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 21501582 ↗
  • MalfunctionMar 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    A CONCENTRATOR CAUGHT FIRE IN A MANNER CONSISTENT WITH HOUR COUNTER FAILURE. NO INJURIES OCCURRED.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 21501580 ↗

Other Generator, Oxygen, Portable devices

All Generator, Oxygen, Portable devices →