QinFlow Blood and Fluid Warmer QiF-01 Base Unit (BU)Recalls & reported problems

Quality In Flow Ltd · QinFlow Blood and Fluid Warmer QiF-01 Base Unit (BU). The BU controls the performance of the system and the outflow fluid temperature. The BU contains firmware (SW) and electronics (HW), a graphic display and audio speaker for user notifications. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07290016740011), received between May 22, 2018 and Apr 23, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Operational Issue” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Operational Issue1
  • Fail-Safe Design Failure1
  • Fail-Safe Problem1
  • Fluid/Blood Leak1
  • Protective Measures Problem1

Patient problems

  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2019Fluid/Blood Leak

    THE USER PROVIDED THE FOLLOWING DESCRIPTION: THEY USED THE UNIT FOR WARMING BLOOD. THE UNIT WORKED WELL FOR ~5 MINUTES. THEN THEY NOTICED BLOOD WAS LEAKING FROM THE CONNECTOR OF THE DISPOSABLE UNIT AND IT FELT TOO HOT. NO ALERT WAS PROVIDED BY THE UNIT. NO SMELL OF SMOKE / BURN. THEY IMMEDIATELY CUT THE WARMING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8540249 ↗
  • MalfunctionMay 22, 2018Fail-Safe Design Failure; Device Operational Issue; Fail-Safe Problem; Protective Measures Problem; Fail-Safe Design Failure; Device Operational Issue; Fail-Safe Problem; Protective Measures Problem

    DURING USER VALIDATION TESTING OF THE QIF BLOOD AND FLUID WARMER BASE UNIT AND COMPACT DISPOSABLE UNIT (CDU) BY U.S. AIR FORCE AFMESA, TESTERS RECORDED A HIGH IV FLUID OUTPUT TEMPERATURE OF 75°C (167°F) ON ONE OF THE UNITS. THE OUTPUT TEMPERATURE WAS MEASURED USING A CALIBRATED THERMOCOUPLE. AT THE SAME TIME, THE SUBJECT BASE UNIT DISPLAYED A "MALFUNCTION TOUT 74°C" NOTIFICATION IN ADDITION TO A CONTINUOUS SOUND ALERT. THE TESTERS SUSPECTED THAT THE CDU WAS CAUSING THE HIGH TEMPERATURE PROBLEM AND TESTED IT WITH A DIFFERENT BASE UNIT AND BATTERY USING THE SAME PARAMETERS. THE OUTPUT TEMPERATURE WAS NORMAL AT 38°C. THE SUBJECT BASE UNIT WAS ALSO TESTED WITH A DIFFERENT BATTERY AND CDU. FOR THIS LAST TEST, AS SOON AS THE BASE UNIT WAS TURNED ON TO START THE RUN, THE CDU HOUSING BEGAN TO MELT AND THE TESTERS STOPPED THE TEST.

    Patient problems: No Information

    MAUDE report 7531909 ↗

Other Warmer, Blood, Non-Electromagnetic Radiation devices