Skip to content

SeQure SQ28_LB_150Recalls & reported problems

Accurate Medical Therapeutics Ltd · SeQure Microcatheter - 2.8 Fr., 150 cm, used with Large Beads · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290017268088), received between Mar 22, 2024 and Mar 22, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 22, 2024Break; Break

    THIS CASE WAS REPORTED BY A FACILITY IN (B)(6) ON 06 MARCH 2024. FACILITY REPORTS THAT SEQURE 2.8FR-150CM MICROCATHETER NORMAL PRE-PROCEDURE PREPARATION DONE. MICROCATHETER FLUSHED WITH SALINE. ADVANCED GUIDEWIRE WITHOUT RESISTANCE. STARTED EMBOLIZATION WITH EMBOZENES (250 MICRONS AND THEN 400 MICRONS). CONTOUR PARTICLES (355 -500) MIXED WITH CONTRAST MEDIA (IMERON 300) AND SALINE. APPROXIMATELY 10ML IMERON INJECTED AT A 80/20 DILUTION WITH SALINE. NO EVENT OBSERVED AT THIS STAGE. SEQURE 2.8FR-150CM MICROCATHETER WAS STILL INSERTED INSIDE THE PATIENT WHEN A SMALL ARTERIAL VASOSPASM WAS REALISED (INJECTED 0.25 MG VERAPAMIL INTRA-ARTERIALLY TO COUNTER THE SPASM). PROCEDURE WAS ALMOST DONE WHEN THE DOCTOR DECIDED TO INJECT MORE CONTOUR PARTICLES. SEQURE 2.8FR-150CM MICROCATHETER WAS FLUSHED INSIDE OF THE PATIENT WITH SALINE AND MET WITH RESISTANCE. DOCTOR ASKED FOR A GUIDEWIRE TO ADVANCE AND OPEN THE BLOCKAGE. DOCTOR THOUGHT THE GUIDEWIRE HAD ADVANCED AS THE RESISTANCE WAS RELIEVED. SAW ON IMAGING THAT THE TIP OF THE SEQURE 2.8FR-150CM MICROCATHETER HAD BROKEN OFF. AT THIS STAGE, EVERYTHING WAS REMOVED. PREPARATION FOR A PROGRADE CATHETER AND INJECTED CONTOUR PARTICLES (300-550) WITHO

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18959258 ↗

Other Catheter, Intravascular, Diagnostic devices

All Catheter, Intravascular, Diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page