LungVision System CoreRecalls & reported problems
Body Vision Medical Ltd · The LungVision System is designed to enable users to segment previously acquired 3D CT data sets and overlay and register these 3D segmented data sets with live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures. The mobile system cart is an optional unit to place the LUNGVISIONTM Main Unit and router, also used for storing all the system components: LV Board, Tablet, DVD reader, Keyboard, Mouse and cables. · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07290017652085), received between Jul 27, 2022 and Dec 5, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Computer Software Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Computer Software Problem2
- Unexpected Shutdown1
Patient problems
- Unintended Radiation Exposure2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 5, 2022Unexpected Shutdown; Unexpected Shutdown
DURING THE PROCEDURE THE COMPUTER POWER WAS REPEATEDLY LOST. THE USER LOST LESION IMAGE AND LOCATION AND COULD NOT CONTINUE SAMPLING FROM THE ACCURATE LOCATION. BIOPSIES WERE TAKEN BUT THE CASE WAS EVENTUALLY ABORTED AND NOT FINISHED AS PLANNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15914881 ↗MalfunctionJul 27, 2022Computer Software Problem; Computer Software Problem
AFTER SUCCESSFULLY COMPLETING THE FIRST PROCEDURE OF THE DAY, DR. (B)(6) WAS WAITING FOR THE PROCEDURE APPLICATION TO FINISH PROCESSING THE CASE DATA AND CLOSE. WHILE WAITING FOR THIS PROCESS TO COMPLETE THE CLINICIAN PERFORMED THE HARD SHUTDOWN OF THE TABLET AND THEN RESTARTED IT WITHOUT RESTARTING THE WHOLE SYSTEM, INCLUDING TABLET AND MAIN UNIT, AS RECOMMENDED BY VENDOR. THIS CAUSED THE TABLET AND MAIN UNIT TO BECOME OUT OF SYNC IN THE SUBSEQUENT TWO CASES THAT DAY. DURING CASES 2 AND 3, FALSE CABT ERROR MESSAGES WERE DISPLAYED DURING MAIN CARINA REGISTRATION. THE CLINICIAN WAS ABLE TO CONTINUE THROUGH THE ERROR MESSAGES AND PROCEEDED TO MARK THE MAIN CARINA. AFTER THIS, THE LESION OVERLAY WAS INITIALLY DISPLAYED ON THE FLUORO VIDEO DURING PROCEDURE; HOWEVER, AS SOON AS THE CLINICIAN MADE HIS FIRST ISOCENTER ADJUSTMENT, THE OVERLAY DISAPPEARED AND NEVER CAME BACK. HE TRIED MULTIPLE VIEWING ANGLES IN AN ATTEMPT TO GET THE OVERLAY TO REAPPEAR BUT IT DID NOT. THUS, DUE TO THE SOFTWARE ERROR MESSAGES AND THE INABILITY TO TROUBLESHOOT, THE CLINICIAN ABANDONED THE USE OF THE LUNGVISION SYSTEM AND UTILIZED THE MONARCH AURIS SYSTEM TO COMPLETE THE PROCEDURE. THOUGH THE DEVICE SOFTWARE M
Patient problems: Unintended Radiation Exposure
MAUDE report 15114368 ↗MalfunctionJul 27, 2022Computer Software Problem; Computer Software Problem
AFTER SUCCESSFULLY COMPLETING THE FIRST PROCEDURE OF THE DAY, DR. HOGARTH WAS WAITING FOR THE PROCEDURE APPLICATION TO FINISH PROCESSING THE CASE DATA AND CLOSE. WHILE WAITING FOR THIS PROCESS TO COMPLETE THE CLINICIAN PERFORMED THE HARD SHUTDOWN OF THE TABLET AND THEN RESTARTED IT WITHOUT RESTARTING THE WHOLE SYSTEM, INCLUDING TABLET AND MAIN UNIT, AS RECOMMENDED BY VENDOR. THIS CAUSED THE TABLET AND MAIN UNIT TO BECOME OUT OF SYNC IN THE SUBSEQUENT TWO CASES THAT DAY. DURING CASES 2 AND 3, FALSE CABT ERROR MESSAGES WERE DISPLAYED DURING MAIN CARINA REGISTRATION. THE CLINICIAN WAS ABLE TO CONTINUE THROUGH THE ERROR MESSAGES AND PROCEEDED TO MARK THE MAIN CARINA. AFTER THIS, THE LESION OVERLAY WAS INITIALLY DISPLAYED ON THE FLUORO VIDEO DURING PROCEDURE; HOWEVER, AS SOON AS THE CLINICIAN MADE HIS FIRST ISOCENTER ADJUSTMENT, THE OVERLAY DISAPPEARED AND NEVER CAME BACK. HE TRIED MULTIPLE VIEWING ANGLES IN AN ATTEMPT TO GET THE OVERLAY TO REAPPEAR BUT IT DID NOT. THUS, DUE TO THE SOFTWARE ERROR MESSAGES AND THE INABILITY TO TROUBLESHOOT, THE CLINICIAN ABANDONED THE USE OF THE LUNGVISION SYSTEM AND UTILIZED THE MONARCH AURIS SYSTEM TO COMPLETE THE PROCEDURE. THOUGH THE DEVICE SOFTWARE
Patient problems: Unintended Radiation Exposure
MAUDE report 15114135 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.