IceFORCE® 2.1 CX Cryoablation Needle FPRPR3604Recalls & reported problems

Boston Scientific Corporation · 2.1mm Cryoablation Needle · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07290104831027), received between Dec 22, 2023 and Oct 23, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Cooling” (2 reports).

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 18, 2020Class IICompleted

    Complaint trend regarding needle shaft gas leaks.

    Z-0658-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jun 4, 2020Class IITerminated

    The needle surface deteriorated due to unexpected electrolysis during procedure.

    Z-2677-2020 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Cooling2
  • Thermal Decomposition of Device2
  • Difficult to Remove1
  • Insufficient Heating1

Patient problems

  • Burn(s)2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 23, 2024Thermal Decomposition of Device; Insufficient Cooling; Thermal Decomposition of Device; Insufficient Cooling

    IT WAS REPORTED FROST OCCURRED ON THE DEVICE AND THE PATIENT EXPERIENCED A SKIN BURN. TWO CRYOABLATION NEEDLES WERE CHOSEN FOR THE CRYOABLATION TREATMENT OF A RENAL TUMOR LOCATED IN THE RIGHT KIDNEY. UNDER COMPUTED TOMOGRAPHY (CT) IMAGING, THE CORRECT PLACEMENT OF THE NEEDLES WAS CONFIRMED. THE FREEZE BEGAN, BUT ABOUT HALFWAY THROUGH THE FIRST FREEZE, THE MACHINE STOPPED FREEZING INTERMITTENTLY. THEY STOPPED THE FREEZE CYCLE AND TOOK THE NEEDLES OUT OF THE MACHINE AND THEN ATTEMPTED TO PLUG THEM BACK IN TO SEE IF IT WOULD RESOLVE THE ISSUE. HOWEVER, THE NEEDLES WOULD NOT GO BACK INTO THE PREVIOUS SLOTS. THE NEEDLES WERE REPOSITIONED IN NEW SLOTS, AND THE FREEZE CYCLE WAS STARTED AGAIN. DURING THE FREEZE, IT WAS NOTICED THAT THE NEEDLES WERE NOT GIVING A TEMPERATURE READING. WHEN THE NEEDLES WERE OBSERVED FOR ANY ICE FORMATION, IT WAS NOTED THAT ONE OF THEM DID NOT LOOK RIGHT. THE MACHINE WAS SHUT OFF IMMEDIATELY. WHEN EXAMINING THE PATIENT, THE SKIN WAS WHITE AND COLD TO THE TOUCH. THE SKIN AROUND THE NEEDLE WAS WARMED UP BY PLACING A HAND OVER THE SKIN. THEY WAITED ABOUT 20 MINUTES AND THEN THE PHYSICIAN TOOK THE 2 CRYOABLATION NEEDLES OUT OF THE KIDNEY. THEY DID A FOLLOW UP CT SC

    Patient problems: Burn(s)

    MAUDE report 20514143 ↗
  • InjuryOct 15, 2024Thermal Decomposition of Device; Insufficient Cooling; Thermal Decomposition of Device; Insufficient Cooling

    IT WAS REPORTED FROST OCCURRED ON THE DEVICE AND THE PATIENT EXPERIENCED A SKIN BURN. TWO CRYOABLATION NEEDLES WERE CHOSEN FOR THE CRYOABLATION TREATMENT OF A RENAL TUMOR LOCATED IN THE RIGHT KIDNEY. UNDER COMPUTED TOMOGRAPHY (CT) IMAGING, THE CORRECT PLACEMENT OF THE NEEDLES WAS CONFIRMED. THE FREEZE BEGAN, BUT ABOUT HALFWAY THROUGH THE FIRST FREEZE, THE MACHINE STOPPED FREEZING INTERMITTENTLY. THEY STOPPED THE FREEZE CYCLE AND TOOK THE NEEDLES OUT OF THE MACHINE AND THEN ATTEMPTED TO PLUG THEM BACK IN TO SEE IF IT WOULD RESOLVE THE ISSUE. HOWEVER, THE NEEDLES WOULD NOT GO BACK INTO THE PREVIOUS SLOTS. THE NEEDLES WERE REPOSITIONED IN NEW SLOTS, AND THE FREEZE CYCLE WAS STARTED AGAIN. DURING THE FREEZE, IT WAS NOTICED THAT THE NEEDLES WERE NOT GIVING A TEMPERATURE READING. WHEN THE NEEDLES WERE OBSERVED FOR ANY ICE FORMATION, IT WAS NOTED THAT ONE OF THEM DID NOT LOOK RIGHT. THE MACHINE WAS SHUT OFF IMMEDIATELY. WHEN EXAMINING THE PATIENT, THE SKIN WAS WHITE AND COLD TO THE TOUCH. THE SKIN AROUND THE NEEDLE WAS WARMED UP BY PLACING A HAND OVER THE SKIN. THEY WAITED ABOUT 20 MINUTES AND THEN THE PHYSICIAN TOOK THE 2 CRYOABLATION NEEDLES OUT OF THE KIDNEY. THEY DID A FOLLOW UP CT SC

    Patient problems: Burn(s)

    MAUDE report 20450922 ↗
  • MalfunctionDec 22, 2023Insufficient Heating; Difficult to Remove; Insufficient Heating; Difficult to Remove

    IT WAS REPORTED THAT THERE WAS RESISTANCE WHEN REMOVING THE NEEDLE. THIS ICEFORCE 2.1 CX NEEDLE WAS SELECTED FOR USE IN A CRYOABLATION FOR RENAL CELL CARCINOMA. DURING THE PROCEDURE, THERE WAS AN ITHAW MALFUNCTION DURING ACTIVE THAWING. THE PHYSICIAN WAITED SEVERAL MINUTES BEFORE ATTEMPTING TO REMOVE THE NEEDLE, BUT UPON REMOVAL THEY FELT RESISTANCE AND WORRIED ABOUT INTERNAL BLEEDING. THE FIELD INDICATED THAT THE RESISTANCE WAS LIKELY DUE TO NOT ENOUGH TIME ALLOWED FOR PASSIVE THAW. BLEEDING WAS NOT EVIDENT ON POST-CT SCAN, AND FURTHER FOLLOW-UP REGARDING THE PATIENT OUTCOME DID NOT REVEAL ANY PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18387579 ↗

Other Unit, Cryosurgical, Accessories devices

All Unit, Cryosurgical, Accessories devices →