IceFORCE™ 2.1 CX L 90° Cryoablation Needle FPRPR3618Recalls & reported problems
Boston Scientific Corporation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07290104831089), received between Aug 26, 2019 and May 18, 2020 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
2 FDA recall records name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 18, 2020Class IICompleted
Complaint trend regarding needle shaft gas leaks.
Z-0659-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jun 4, 2020Class IITerminated
The needle surface deteriorated due to unexpected electrolysis during procedure.
Z-2683-2020 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available5
- Defective Device1
Patient problems
- Muscle Spasm(s)5
- Cramp(s) /Muscle Spasm(s)1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2020Defective Device; Defective Device
IT WAS REPORTED THAT THE PATIENT EXPERIENCED MUSCLE STIMULATION DURING ACTIVE THAW (ITHAW). PATIENT UNDERWENT A RENAL CRYOABLATION PROCEDURE WITH 4 ICEFORCE AND 1 ICEFORCE LONG NEEDLES. DURING THE ITHAW STEP OF THE PROCEDURE, THE PATIENT TOLD THE NURSE THAT THE "THING ON THEIR BACK" WAS CAUSING HIM TO MOVE. THE MUSCLE STIMULATION STOPPED IMMEDIATELY WHEN THE ITHAW FEATURE WAS TURNED OFF. THE PHYSICIAN CONTINUED THE CASE USING PASSIVE THAW, AND THE CASE WAS SUCCESSFULLY COMPLETED WITH NO REPORTED PATIENT INJURY OR FURTHER COMPLICATIONS. THE NEEDLE CAUSING THE MUSCLE STIMULATION WAS NOT IDENTIFIED, AS THE PHYSICIAN DID NOT WANT TO CAUSE ANOTHER SHOCK TO THE PATIENT WHILST TRYING TO IDENTIFY THE INDIVIDUAL NEEDLE CAUSING THE PROBLEM.
Patient problems: Cramp(s) /Muscle Spasm(s); Muscle Spasm(s)
MAUDE report 10066145 ↗MalfunctionFeb 12, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
TWO NEEDLES WERE USED DURING A CRYOABLATION CASE. THIS MDR IS BEING SUBMITTED FOR THE SECOND OF THE NEEDLES USED DURING THE CRYOABLATION PROCEDURE. DURING THE ACTIVE THAW CYCLE TOWARDS THE END OF THE PROCEDURE, ONE OF THE TWO NEEDLES CAUSED MUSCLE STIMULATION. THE CASE WAS COMPLETED, WITH NO REPORTED INJURY TO PATIENT. THE USER IS UNSURE WHICH OF THE NEEDLES CAUSED THE MUSCLE STIMULATION. THE NEEDLES WILL BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION.
Patient problems: Muscle Spasm(s)
MAUDE report 9703434 ↗MalfunctionFeb 12, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
TWO NEEDLES WERE USED DURING A CRYOABLATION CASE. THIS MDR IS BEING SUBMITTED FOR THE FIRST OF THE NEEDLES USED DURING THE CRYOABLATION PROCEDURE. DURING THE ACTIVE THAW CYCLE TOWARDS THE END OF THE PROCEDURE, ONE OF THE TWO NEEDLES CAUSED MUSCLE STIMULATION. THE CASE WAS COMPLETED, WITH NO REPORTED INJURY TO PATIENT. THE USER IS UNSURE WHICH OF THE NEEDLES CAUSED THE MUSCLE STIMULATION. THE NEEDLES WILL BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION.
Patient problems: Muscle Spasm(s)
MAUDE report 9703425 ↗MalfunctionJan 3, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THREE NEEDLES WERE USED FOR A RENAL CRYOABLATION PROCEDURE. THE NEEDLES WERE TESTED WITH NO ISSUES. DURING THE FIRST FREEZE CYCLE, ONE OF THE NEEDLE'S CAUSED THE PATIENT TO TWITCH. THE CASE WAS COMPLETED AS EXPECTED. THE NEEDLES WILL BE RETURNED FOR INVESTIGATION.
Patient problems: Muscle Spasm(s)
MAUDE report 9547949 ↗MalfunctionJan 2, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A PATIENT UNDERWENT CRYOABLATION FOR A RENAL CASE WITH FIVE ICEFORCE NEEDLES. THE PATIENT WAS UNDER CONSCIOUS SEDATION. DURING ACTIVE THAW CYCLE, THE PATIENT EXPERIENCED MUSCLE SPASM. PASSIVE THAW WAS USED TO COMPLETE THE CASE AS EXPECTED WITH NO PATIENT INJURY. IT IS UNKNOWN WHICH NEEDLE CAUSED THE STIMULATION; THEREFORE ALL FIVE NEEDLES WILL BE RETURNED FOR INVESTIGATION.
Patient problems: Muscle Spasm(s)
MAUDE report 9538114 ↗MalfunctionAug 26, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING THE FREEZE CYCLE OF A CRYOABLATION PROCEDURE, FROST COLLECTED ON THE SHAFT OF THE NEEDLE. THE PATIENT'S SKIN WAS PROTECTED AND THE CASE WAS COMPLETED WITH NO INJURY TO THE PATIENT. THE NEEDLE WAS RETURNED FOR INVESTIGATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8932681 ↗
Other Unit, Cryosurgical, Accessories devices
- IceFORCE® 2.1 CX 90° Cryoablation Needle FPRPR3602Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls4 deaths274 reports
- IceRod® 1.5 CX 90° Cryoablation Needle FPRPR3533Boston Scientific Corporation · Unit, Cryosurgical, Accessories1 recall168 reports
- IcePearl® 2.1 CX 90° Cryoablation Needle FPRPR3601Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls78 reports
- IceFORCE® 2.1 CX Cryoablation Needle FPRPR3604Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls3 reports
- IcePearl™ 2.1 CX L 90° Cryoablation Needle FPRPR3617Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls1 report
- IceRod® He-Thaw Prostate Cryoablation Kit / Presice® System FPRPR4570Galil Medical Ltd · Unit, Cryosurgical, Accessories2 recalls0 reports
- IceRod® i-Thaw® Prostate Cryoablation Kit / Presice® System FPRPR4070Galil Medical Ltd · Unit, Cryosurgical, Accessories2 recalls0 reports
- Three IcePearl® 2.1 CX Prostate Cryoablation KitVisual-ICE® System FPRPR3608Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page