Vial Adaptor 8072045Recalls & reported problems
West Pharma. Services Il, . Ltd · Vial Adapter 20mm FLL w/ Ribs - VF · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290108240412), received between Jun 12, 2024 and Jun 12, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Contamination” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination1
- Contamination /Decontamination Problem1
- Device Contaminated During Manufacture or Shipping1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2024Contamination; Contamination /Decontamination Problem; Device Contaminated During Manufacture or Shipping; Contamination; Contamination /Decontamination Problem; Device Contaminated During Manufacture or Shipping
ON 13MAY2024, THE CUSTOMER, FERRING PHARMACEUTICALS, CONTACTED WEST PHARMA. SERVICES, IL LTD. (WEST IL), TO REPORT THAT DURING ASSEMBLY OF FERRING'S COMBINATION PACK, A VIAL ADAPTER, WAS OBSERVED WITH A BLUE FOREIGN PARTICULATE MATTER WITHIN THE PRIMARY STERILE PACKAGE. SUBSEQUENTLY, THE CUSTOMER OBSERVED A SECOND VIAL ADAPTER WITH ANOTHER BLUE FOREIGN PARTICULATE MATTER WITHIN THE PRIMARY STERILE PACKAGE. BOTH VIAL ADAPTERS ARE FROM LOT # F700. THE WEST PHARMA. VIAL ADAPTERS ARE CO-PACKED WITH THE DRUG FIRMAGON, THIS MEDICATION IS TO BE ADMINISTERED SUBCUTANEOUSLY AS AN INJECTION. BATCH RECORDS FOR LOT# F700 WERE REVIEWED AND THERE WERE NO NON CONFORMANCES FOUND. QUALITY CONTROL INSPECTIONS WERE CONDUCTED ACCORDING TO PROCEDURES DURING PRODUCTION, NO ISSUES WERE NOTICED. THE INCOMING INSPECTION RECORDS OF THE TYVEK AND NEEDLES WERE REVIEWED, NO ISSUES WERE NOTICED. LOT #F700 WAS MANUFACTURED ACCORDING TO RELEVANT PROCEDURES TESTED BEFORE RELEASE AND SHIPPED ACCORDING TO SPECIFICATIONS. RETAINED SAMPLES FROM LOT #F700 WERE 100% VISUALLY INSPECTED BY THE CONTRACT MANUFACTURER ACCORDING TO THEIR INTERNAL PROCEDURE, NO FINDINGS WERE OBSERVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19520328 ↗
FDA clearance and classification
Cleared through 510(k) K963583 on Oct 7, 1996 (decision: Substantially Equivalent). Submitted by Medimop Medical Projects, Ltd. as “Mixject Dispensing Pin/With Detachable Vial Holder/With Detachable Vial Holder And Preattached Needle”. The FDA review took 31 days. FDA record ↗
The FDA regulates product code LHI (set, i.v. fluid transfer) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Vial Adaptor 8072045
Has the Vial Adaptor 8072045 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07290108240412), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Vial Adaptor 8072045?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 12, 2024. That report lists device problems “Contamination” (1), “Contamination /Decontamination Problem” (1) and “Device Contaminated During Manufacture or Shipping” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Vial Adaptor 8072045?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Vial Adaptor 8072045 FDA approved?
It was cleared through 510(k) K963583 on Oct 7, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Vial Adaptor 8072045, and is it still sold?
The FDA's device database lists West Pharma. Services Il, . Ltd as the company responsible for it (brand Vial Adaptor, model 8072045). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Set, I.V. Fluid Transfer devices
- EXACTAMIX H938740Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall1,345 reports
- EXACTAMIX H938738Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall1,172 reports
- EXACTAMIX H938739Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall960 reports
- EXACTAMIX H938741Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall708 reports
- EXACTAMIX H938737Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall597 reports
- EXACTAMIX H938724Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall510 reports
- EXACTAMIX H938173Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls258 reports
- EXACTAMIX H938176Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls228 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page