Swabable Vial Adapter 8070101Recalls & reported problems

West Pharma. Services Il, . Ltd · VA 13mm Swabable - VF · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 07290108240498), received between Mar 4, 2025 and Sep 24, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Contamination” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 2 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination13
  • Contamination /Decontamination Problem12
  • Contamination of Device Ingredient or Reagent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Contamination /Decontamination Problem

    The customer, Dompe farmaceutici SpA, contacted the device manufacturer on 25Aug26 to report during packaging of the Swabable Vial Adapter (SVA) with the customer's drug product, foreign particles were detected within the primary packaging of the devices and also within the sealing area of the primary packaging. This deficiency was found prior to use. The device had not yet entered into the distribution stage to the end user (i.e. being put into service), there was no health impact to the end user.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882976 ↗
  • MalfunctionJul 29, 2026Contamination; Contamination /Decontamination Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50069903 ↗
  • MalfunctionJun 16, 2026Contamination

    THE CUSTOMER, DOMPE FARMACEUTICI SPA, CONTACTED THE DEVICE MANUFACTURER ON 21MAY2026 TO REPORT THAT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THIS DEFICIENCY WAS FOUND PRIOR TO USE. AS THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25519773 ↗
  • MalfunctionJun 10, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, (B)(6), CONTACTED THE DEVICE MANUFACTURER ON 06MAY2026 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25457700 ↗
  • MalfunctionMay 19, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, DOMPE FARMACEUTICI SPA, CONTACTED THE DEVICE MANUFACTURER ON 22APR2026 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THE CUSTOMER CONFIRMED THAT THE AFFECTED DEVICES ARE SEGREGATED, NOT FOR USE. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25217723 ↗
  • MalfunctionApr 27, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, (B)(6), CONTACTED THE DEVICE MANUFACTURER ON 09APR2026 TO REPORT DURING PACKAGING OF THE SWABBLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THE CUSTOMER CONFIRMED THAT THE AFFECTED DEVICES ARE SEGREGATED, NOT FOR USE. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24999055 ↗
  • MalfunctionApr 15, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, DOMPE FARMACEUTICI SPA, CONTACTED THE DEVICE MANUFACTURER ON 19MAR2026 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THE CUSTOMER CONFIRMED THAT THE AFFECTED DEVICES ARE SEGREGATED, NOT FOR USE. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885765 ↗
  • MalfunctionApr 2, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, DOMPE FARMACEUTICI SPA, CONTACTED THE DEVICE MANUFACTURER ON 09MAR2026 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THE CUSTOMER CONFIRMED THAT THE AFFECTED DEVICES ARE SEGREGATED, NOT FOR USE. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24766066 ↗
  • MalfunctionMar 24, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, (B)(6), CONTACTED THE DEVICE MANUFACTURER ON 24-FEB-2026 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24680855 ↗
  • MalfunctionFeb 25, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, (B)(6), CONTACTED THE DEVICE MANUFACTURER ON 03-FEB-2026 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24444173 ↗
  • MalfunctionFeb 11, 2026Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, (B)(6), CONTACTED THE DEVICE MANUFACTURER ON 16JAN2026 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. THIS DEFICIENCY WAS FOUND PRIOR TO USE. THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), AND THEREFORE THERE WAS NO HEALTH IMPACT TO THE END USER. SAMPLES HAVE BEEN REQUESTED FROM THE CUSTOMER AND WILL FORM PART OF THE INVESTIGATION ONCE RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24314829 ↗
  • MalfunctionDec 23, 2025Contamination; Contamination /Decontamination Problem; Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, (B)(6), CONTACTED THE DEVICE MANUFACTURER ON (B)(6) 2025 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. FOREIGN PARTICLES WERE DETECTED IN MULTIPLE DEVICES FROM LOT K717 OF THE SWABABLE VIAL ADAPTERS (SVA). THE CUSTOMER CONFIRMED THAT THE AFFECTED DEVICES ARE SEGREGATED, NOT FOR USE. AS THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23878120 ↗
  • MalfunctionDec 11, 2025Contamination; Contamination /Decontamination Problem; Contamination; Contamination /Decontamination Problem

    THE CUSTOMER, (B)(6), CONTACTED THE DEVICE MANUFACTURER ON (B)(6) 2025 TO REPORT DURING PACKAGING OF THE SWABABLE VIAL ADAPTER (SVA) WITH THE CUSTOMER'S DRUG PRODUCT, FOREIGN PARTICLES WERE DETECTED WITHIN THE PRIMARY PACKAGING OF THE DEVICES AND ALSO WITHIN THE SEALING AREA OF THE PRIMARY PACKAGING. FOREIGN PARTICLES WERE DETECTED IN MULTIPLE DEVICES FROM LOT K716 OF THE SWABABLE VIAL ADAPTERS (SVA). THE CUSTOMER CONFIRMED THAT THE AFFECTED DEVICES ARE SEGREGATED, NOT FOR USE. AS THE DEVICE HAD NOT YET ENTERED INTO THE DISTRIBUTION STAGE TO THE END USER (I.E. BEING PUT INTO SERVICE), THERE WAS NO HEALTH IMPACT TO THE END USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23778662 ↗
  • MalfunctionMar 4, 2025Contamination; Contamination of Device Ingredient or Reagent; Contamination; Contamination of Device Ingredient or Reagent

    THE CUSTOMER, DOMPÉ FARMACEUTICI, CONTACTED THE DEVICE MANUFACTURER ON 04FEB2025 TO REPORT THAT DURING DISPENSING OF THE CUSTOMER'S DRUG FOR OPHTHALMIC USE, A FOREIGN BODY WAS NOTICED INSIDE A PIPETTE (DEVICE WHICH IS USED FOR THE ADMINISTRATION OF THE DRUG) AFTER RECONSTITUTION WITH THE WEST VIAL ADAPTER. NO ADVERSE EVENT HAS BEEN REPORTED AS THE DOSE RELATED TO THE REPORTED EVENT HAS NOT BEEN ADMINISTERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21518537 ↗

Other Set, I.V. Fluid Transfer devices

All Set, I.V. Fluid Transfer devices →