Skip to content

CarboClear Pedicle Screw System Rod - Type 4Recalls & reported problems

Carbofix Orthopedics Ltd · Shaped Rod of the CarboClear Pedicle Screw System · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07290109336367), received between Apr 29, 2021 and Feb 25, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fracture1

Patient problems

  • Back pain1
  • Paralysis1
  • Spinal Cord Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 25, 2025Break; Break

    INITIALLY, A SURGERY WITH THE CARBOCLEAR X PEDICLE SCREW SYSTEM TO TREAT AN ONCOLOGICAL 61-YEAR-OLD PATIENT WAS PERFORMED ON (B)(6) 2022, IN CANADA. THE CRBOCLEAR X SYSTEM WAS UTILIZED IN ORDER TO STABILIZE T5 - T10 VERTEBRAE WHILE T8 WAS REMOVED AND A VBR DEVICE WAS USED. ON (B)(6) 2025, CARBOFIX WAS INFORMED BY ITS CANADIAN DISTRIBUTOR THAT WHILE THE PATIENT WAS "WORKING ON HIS CAR" HE NOTICED PAIN. IMAGING WAS CONDUCTED AND REVEALED A BROKEN ROD. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2025, WHERE THE BROKEN ROD (CARBOCLEAR SHAPED ROD TYPE 4: DIAMETER - 6.0MM, LENGTH - 285MM) WAS REPLACED WHILE EXPANDING THE CONSTRUCT WITH NEW RODS FROM T4 TO T11. TWO (2) RODS AND 4 SCREWS (CARBOCLEAR X SYSTEM COMPONENTS) WERE USED.

    Patient problems: Back pain

    MAUDE report 21456097 ↗
  • InjuryApr 29, 2021Fracture; Fracture

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

    Patient problems: Paralysis; Spinal Cord Injury

    MAUDE report 11744190 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page