Piccolo Composite Nailing System Proximal Femur Nail (REF PPFR11420)Recalls & reported problems
Carbofix Orthopedics Ltd · Nail indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric, high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, malunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07290109336534), received between May 6, 2020 and Mar 25, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K151010, K153536); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
Patient problems
- Bone Fracture(s)1
- Hip Fracture1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 25, 2026Fracture
ON MARCH 16, 2026, THE COMPANY BECAME AWARE OF A CASE INVOLVING ITS PICCOLO COMPOSITE PF NAIL. BASED ON THE INFORMATION PROVIDED TO THE COMPANY, THE DEVICE WAS IMPLANTED IN AN ONCOLOGIC PATIENT WITH FEMORAL METASTASES ON (B)(6) 2024, FOLLOWING A FALL RESULTING IN A PATHOLOGICAL FRACTURE OF THE FEMUR. APPROXIMATELY FIVE MONTHS POST-IMPLANTATION, THE NAIL FRACTURED AT OR NEAR THE LAG SCREW INTERFACE DUE TO NON-UNION. ON (B)(6) 2025, THE NAIL WAS REMOVED AND REPLACED WITH A TOTAL HIP PROSTHESIS.
Patient problems: Bone Fracture(s); Nonunion/Delayed-union Bone Fracture
MAUDE report 24686360 ↗InjuryMay 6, 2020Fracture; Fracture
A PROXIMAL FEMUR NAIL WAS IMPLANTED TO TREAT A PATHOLOGIC FRACTURE (ONCOLOGICAL PATIENT) IN (B)(6). MORE THAN 9 MONTHS POST-OPERATION, NAIL BROKE AT THE SUB-TROCHANTERIC AREA AND WAS REPLACED WITH A METAL NAIL IN A REVISION SURGERY.
Patient problems: Hip Fracture
MAUDE report 10028610 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K151010, K153536. One submission often covers many models, so check the product and lot details in each record.
- Z-1991-2016 · May 10, 2016 · TerminatedPiccolo Composite Tibia and Femoral Nailing Systems - Ball Tip Guide Wire for fixation of fractures that occur in and between the proximal and distal third of the long bones being
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page