CarboClear Pedicle Screw System Pedicle Screw - Non Cannulated (REF PNNS5545)Recalls & reported problems

Carbofix Orthopedics Ltd · Non cannulated pedicle screw (oncology) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07290110200640), received between Jul 4, 2019 and Jun 18, 2020 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture3
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2020Fracture; Fracture

    DURING A SURGERY WITH THE CARBOCLEAR PEDICLE SCREW SYSTEM, FOR THE TREATMENT OF ONCOLOGICAL PATIENT, A SCREW BROKE WHILE BEING INSERTED INTO THE PEDICLE; THE SCREW WAS REMOVED AND SURGERY WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10166994 ↗
  • MalfunctionJan 7, 2020Fracture; Fracture

    DURING FIXATION OF T6-T11 VERTEBRAE OF ONCOLOGICAL PATIENT IN (B)(6), ONE OF THE SCREWS AT T9 WAS INSERTED IN A MANNER THAT THE SCREW TULIP TOUCHED THE PEDICLE. THIS CAUSED SIGNIFICANT RESISTANCE THAT LED TO EXCESSIVE TORQUE AND BREAKAGE OF THE SCREW. THE PHYSICIAN DECIDED THE CONSTRUCT IS STRONG ENOUGH (LEAVING THE SCREW IN SITU, WITH FIXATION OF TWO LEVELS ABOVE AND TWO LEVELS BELOW THE AFFECTED VERTEBRAE).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9556734 ↗
  • MalfunctionJul 4, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A SURGERY IN (B)(6), A PEDICLE SCREW DISASSEMBLED (THE "TULIP" DISCONNECTED FROM THE SCREW); THE SCREW WAS GRIPPED BY A PLIER AND REMOVED (UNSCREWED).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8761996 ↗
  • MalfunctionJul 4, 2019Fracture; Fracture

    DURING A SURGERY WITH THE CARBOCLEAR PEDICLE SCREW SYSTEM IN (B)(6), FOR THE TREATMENT OF ONCOLOGICAL PATIENT, A SCREW BROKE WHILE BEING INSERTED INTO THE PEDICLE; THE SCREW WAS SUCCESSFULLY REMOVED WITH A DRILL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8761995 ↗