Skip to content

CarboClear Pedicle Screw System Pedicle Screw - Non Cannulated (REF PNNS5535)Recalls & reported problems

Carbofix Orthopedics Ltd · Non cannulated pedicle screw (oncology) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290110200657), received between Jan 19, 2020 and Jan 19, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2020Fracture; Fracture

    THE CASE INVOLVED A PROSTATE CANCER PATIENT WITH METASTATIC LESION AT HIS UPPER THORACIC LEVELS. THE BONE WAS VERY STIFF. AS THE SURGEON WANTED TO USE A 5.5MM SCREW, CARBOFIX REPRESENTATIVE WHO ATTENDED THE SURGERY RECOMMENDED TO USE A 5.5MM BONE TAP. THE SURGEON DECIDED TO USE A 5.0MM TAP. TOWARDS THE END OF THE SCREW INSERTION, IT BROKE. THE SURGEON REMOVED THE REMAINDER OF THE SCREW FROM THE PEDICLE, SKIPPED A LEVEL, AND SUCCESSFULLY CONCLUDED THE SURGERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9604850 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page