Skip to content

CarboClear X Pedicle Screw System Pedicle Screw - CannulatedRecalls & reported problems

Carbofix Orthopedics Ltd · Cannulated pedicle screw · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290110209896), received between May 27, 2022 and May 27, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2022Improper or Incorrect Procedure or Method; Detachment of Device or Device Component; Improper or Incorrect Procedure or Method; Detachment of Device or Device Component

    CARBOCLEAR X PEDICLE SCREW SYSTEM WAS USED IN A REVISION SURGERY FOR THE TREATMENT OF DEFORMITY SCOLIOSIS FOR A METAL ALLERGY PATIENT (THE ORIGINAL SURGERY WAS PERFORMED WITH METAL PEDICLE SCREW SYSTEM OF ANOTHER COMPANY; CARBOCLEAR X SYSTEM OR ANY OTHER IMPLANT OF CARBOFIX WERE NOT USED IN THE FIRST OPERATION). A LONG CONSTRUCT WAS INVOLVED (T4 - ALA). TITANIUM SCREWS WERE REPLACED WITH CARBOCLEAR X SCREWS. DUE TO THE LENGTH OF THE RODS REQUIRED, THE SURGEON USED PRE-BENT TITANIUM RODS OF ANOTHER COMPANY. DURING INSERTION OF A CARBOCLEAR X SCREW INTO A LOW LUMBAR VERTEBRA, THE SCREW' TULIP WAS DISASSEMBLED FROM THE SCREW. NO TAPPING WAS CONDUCTED PRIOR TO SCREW INSERTION INTO THE VERTEBRA. THE DISASSEMBLED SCREW WAS REMOVED. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14521419 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page