Cartridge Blood Set 101025Recalls & reported problems
Gambro Renal Products, S.A. de C.V. · The Gambro Cartridge Blood Set is intended for single use in a hemodialysis treatment using the Phoenix Dialysis Delivery Systems. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Gambro Cartridge Blood Set is a single-use hemodialysis blood tubing set. Reports mostly describe leaks, cracks, disconnections, or breaks during priming or therapy, usually with no patient injury. There are 80 adverse event reports, no deaths and 2 injuries. There are 3 recalls, 1 still open about kinked access lines that began December 2020. Reports fell from 7 to 2 over the last 12 months compared with the prior year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 80
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 80 adverse event reports that list this exact device (DI 07332414007836), received between May 29, 2018 and May 6, 2026 — classified as 2 injuries, 78 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (62 reports).
3 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 14, 2021Class IITerminated
There is a potential leak between the venous patient connector and patient's hemodialysis access.
Z-1684-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 30, 2020Class IIOpen, Classified
complaints of kinked access lines observed during treatment using cartridge sets for hemodialysis treatments were received.
Z-1118-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 17, 2020Class IITerminated
Potential disconnection of tubing set.
Z-2193-2020 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak62
- Crack11
- Disconnection10
- Break8
- Leak/Splash8
- Material Split, Cut or Torn5
- Adverse Event Without Identified Device or Use Problem2
- Material Rupture2
Patient problems
- No Clinical Signs, Symptoms or Conditions51
- No Consequences Or Impact To Patient21
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Hemolysis2
- No Known Impact Or Consequence To Patient2
- Chest Pain1
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 80 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2026Break; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE BROKEN CASSETTE OF A GAMBRO CARTRIDGE LINE DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25101375 ↗MalfunctionApr 13, 2026Break; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE "BROKEN LINE" OF A GAMBRO CARTRIDGE SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865287 ↗MalfunctionAug 27, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK ALONG THE VENOUS LINE OF THE CASSETTE WAS OBSERVED FROM A GAMBRO CARTRIDGE SET AT THE TIME OF PATIENT CONNECTION FOR HEMODIALYSIS THERAPY. THE SET WAS REPLACED TO CONTINUE TREATMENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22903589 ↗MalfunctionMay 16, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE BACK OF THE CASSETTE OF A GAMBRO CARTRIDGE SET DURING HEMODIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22038475 ↗MalfunctionFeb 5, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL BLOOD LEAK WAS OBSERVED FROM THE ACCESS SAMPLE SITE OF A GAMBRO CARTRIDGE SET FIVE (5) MINUTES AFTER THE START OF HEMODIALYSIS THERAPY. AIR WAS ALSO PRESENT IN THE ACCESS LINE. THE VOLUME OF BLOOD LOSS WAS APPROXIMATELY 190ML. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21314513 ↗MalfunctionFeb 3, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM A GAMBRO CARTRIDGE SET DURING HEMODIALYSIS THERAPY. THE ABSENCE OF A ¿SAMPLE COLLECTION PORT¿ ON THE SET WAS OBSERVED. THE BLOODLINES WERE REPLACED TO CONTINUE TREATMENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21298025 ↗MalfunctionJan 27, 2025Break; Fluid/Blood Leak; Break; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL LEAK OF SALINE SOLUTION WAS OBSERVED FROM A BREAKAGE IN A GAMBRO CARTRIDGE SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21245750 ↗MalfunctionJan 22, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL BLOOD LEAK WAS OBSERVED FROM THE VENOUS LINE OF GAMBRO CARTRIDGE SET DURING HEMODIALYSIS THERAPY. THE VOLUME OF BLOOD LOSS WAS LES THAN 150ML. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21214461 ↗MalfunctionJan 7, 2025Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT AN EXTERNAL BLOOD LEAK WAS OBSERVED FROM THE VENOUS CONNECTION PORT OF THE PATIENT LINE OF A GAMBRO CARTRIDGE SET. THE VENOUS LINE THREAD DID NOT TIGHTEN PROPERLY. THE VOLUME OF BLOOD LOSS WAS APPROXIMATELY 10ML. THE BLOOD WAS RETURNED TO THE PATIENT AND THE LINES WERE REPLACED TO CONTINUE TREATMENT. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21096191 ↗MalfunctionOct 30, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL BLOOD LEAK WAS OBSERVED FROM A "SMALL FISSURE" IN THE BLOOD PUMP SEGMENT OF A GAMBRO CARTRIDGE SET DURING HEMODIALYSIS THERAPY. THE TREATMENT WAS ENDED WITHOUT BLOOD RESTITUTION. THE VOLUME OF BLOOD LEAKAGE WAS NOT GREATER THAN 20CC. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20566847 ↗MalfunctionMay 7, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL LEAK OF SALINE SOLUTION WAS OBSERVED FROM A CRACK IN A GAMBRO CARTRIDGE SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19259873 ↗MalfunctionFeb 5, 2024Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT AN EXTERNAL LEAK OF SALINE SOLUTION WAS OBSERVED FROM THE ¿PERFORATED¿ LINE OF A GAMBRO CARTRIDGE SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18646543 ↗MalfunctionDec 27, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS A CRACK IN THE CASSETTE OF A GAMBRO CARTRIDGE WHICH RESULTED IN AN EXTERNAL SALINE LEAK. THE LEAK WAS DISCOVERED WHILE PRIMING THE SET PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18400311 ↗MalfunctionNov 1, 2023Fluid/Blood Leak; Fluid/Blood Leak
UPON FURTHER INFORMATION, THE CUSTOMER CONFIRMED THERE WAS NO LEAKAGE FROM THE CARTRIDGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18046924 ↗MalfunctionOct 30, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL FLUID LEAK WAS OBSERVED ORIGINATING FROM THE VENOUS CHAMBER OF A GAMBRO CARTRIDGE SET DURING PRIME. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18030435 ↗MalfunctionOct 2, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL FLUID LEAK (SALINE SOLUTION) WAS OBSERVED ORIGINATING FROM THE BLOOD PUMP SEGMENT TUBING OF A GAMBRO CARTRIDGE DN PRIME LINE DURING PRIME. UPON FURTHER INSPECTION, ¿A CASSETTE FISSURE¿ WAS NOTED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17858294 ↗MalfunctionOct 2, 2023Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS REPORTED AN EXTERNAL FLUID LEAK (SALINE SOLUTION) WAS OBSERVED ORIGINATING FROM A CUT BLOOD PUMP SEGMENT TUBING OF A GAMBRO CARTRIDGE DN PRIME LINE DURING PRIME. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17858119 ↗MalfunctionSep 26, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL SALINE LEAK WAS OBSERVED ORIGINATING IN THE VENOUS CHAMBER OF A GAMBRO CARTRIDGE. THE UNSPECIFIED DIALYSIS MACHINE WAS TURNED OFF AND THE SET WAS REPLACED. THE REPORTER STATED ¿MANUAL BLOOD RETURN WAS PERFORMED¿; HOWEVER, THE EVENT OCCURRED DURING PRIME. THEREFORE, THE EVENT OCCURRED DURING AN UNKNOWN PROCESS SET. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17817683 ↗MalfunctionSep 22, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL FLUID LEAK WAS OBSERVED ORIGINATING FROM THE VENOUS CHAMBER OF A GAMBRO CARTRIDGE SET DURING PRIME. PRIMING WAS STOPPED, THE CASSETTE WAS CHANGED, AND PRIMING WAS RESUMED WITHOUT FURTHER INCIDENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17803763 ↗MalfunctionAug 22, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL FLUID LEAK WAS OBSERVED IN A GAMBRO CARTRIDGE SET DURING PRIME. WHEN THE CASSETTE WAS REMOVED ¿FROM THE HYDROPHOBIC ONES, IT WAS NOTICED THAT A PIECE OF THE CASSETTE WAS ADHERED TO THE WALL OF THE MACHINE¿ AND ¿THE SALINE SOLUTION BEGAN TO SPREAD ON THE PART OF THE CASSETTE OF THE GAMBRO LINE¿. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17595734 ↗MalfunctionAug 21, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL BLOOD LEAK WAS OBSERVED ORIGINATING FROM THE CASSETTE TUBING (BLOOD PUMP SEGMENT) OF A GAMBRO CARTRIDGE DURING HEMODIALYSIS THERAPY. TREATMENT WAS ENDED WITHOUT BLOOD RESTITUTION AND THE ACCUMULATED BLOOD LOST WAS MORE THAN 150ML. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17585833 ↗MalfunctionAug 14, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL FLUID LEAK WAS OBSERVED ORIGINATING FROM THE CASSETTE OF A GAMBRO CARTRIDGE DN PRIME LINE DURING PRIME. THE EVENT OCCURRED WHEN REMOVING THE CASSETTE FROM THE HYDROPHOBES AND A PIECE OF THE CASSETTE WAS STUCK TO THE WALL OF THE MACHINE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17542303 ↗InjuryJul 17, 2023Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS REPORTED THAT 15 MINUTES AFTER STARTING TREATMENT WITH A CARTRIDGE LINE, AN EXTERNAL BLOOD LEAK BETWEEN THE ARTERIAL PATIENT LINE AND THE ARTERIAL INJECTION PORT WAS OBSERVED. IT WAS FURTHER REPORTED AN ALARM WAS GENERATED. ACCORDING TO THE REPORTER IT WAS ¿A CLEAN CUT¿ AND ¿THE TUBING JUST CUT OFF AND THE PATIENT STARTED BLEEDING OUT". THE BLOOD LOSS WAS 700-1000 ML. NO CLINICAL SYMPTOMS WERE REPORTED; HOWEVER, ¿EXTRA FLUID¿ WAS ADMINISTERED TO THE PATIENT PRIOR TO LEAVING. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 17333452 ↗MalfunctionJun 22, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED AN EXTERNAL SALINE LEAK WAS OBSERVED ORIGINATING FROM THE ¿WALLS OF THE CASSETTE¿ OF A GAMBRO CARTRIDGE DN PRIME LINE DURING PRIME. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17182245 ↗MalfunctionMay 30, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS FLUID LEAK FROM THE BACK OF THE VENOUS CHAMBER OF A GAMBRO CATRIDGE. THE LEAK WAS FURTHER DESCRIBED AS, ¿WHEN PLACING THE CASSETTE AND PRIMING IT, A VERY FINE AND CONTINUOUS LIQUID JET WAS OBSERVED COMING OUT FROM THE BACK OF THE VENOUS CHAMBER, AROUND THE CIRCLE FORMED WHERE THE MACHINE'S SENSOR IS INSERTED¿. THIS ISSUE WAS OBSERVED DURING PRIMING PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17020252 ↗
Other SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE devices
- Hemodialysis Bloodlines SL-2010M2096B. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve3 recalls3 deaths1,000 reports
- Hemodialysis Bloodlines SL-2010M2096AB. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve2 recalls1 death483 reports
- Hemodialysis Bloodlines SL-2000M2095DAB. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve2 recalls1 death214 reports
- CombiSet 03-2722-9Fresenius Medical Care Holdings, Inc. · SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE8 deaths311 reports
- Hemodialysis Bloodlines SL-2000M2095B. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve1 recall219 reports
- CombiSet 03-2522-1Fresenius Medical Care Holdings, Inc. · Set, Tubing, Blood, With And Without Anti-Regurgitation Valve1 recall214 reports
- Hemodialysis Bloodlines SL-2000M2095DB. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve1 recall158 reports
- Hemodialysis Bloodlines SL-2000M2095LB. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve2 recalls97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page