Hemodialysis Bloodlines SL-2000M2095DRecalls & reported problems

B. Braun Medical Inc. · STREAMLINE FRESENIUS FOR DAVITA · Class 2

All models of Hemodialysis Bloodlines (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This B. Braun hemodialysis bloodline is tubing that carries blood during dialysis treatment. Many reports describe air getting into the lines, leaks, and holes in the tubing, and some describe parts coming apart. The 158 reports include no deaths and no injuries. One open recall, begun in August 2024, concerns a patient connector that may not fully insert, which could block flow when blood is returned to the patient.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
158
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 158 adverse event reports that list this exact device (DI 04046955348763), received between Jun 11, 2021 and Jan 9, 2024 — classified as 158 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (66 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 14 reportsJan 2024: 6 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2024Class IIOpen, Classified

    Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

    Z-3174-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Air/Gas in Device66
  • Leak/Splash50
  • Material Puncture/Hole32
  • Detachment of Device or Device Component23
  • Material Deformation15
  • Crack3
  • Contamination /Decontamination Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions142
  • Hemorrhage/Blood Loss/Bleeding16

Recent adverse event reports

The 25 most recent of 158 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 3 WE REALIZED IN A 2-3 DAY PERIOD THAT THE WHOLE SYSTEM WOULD FILL WITH AIR ONLY WHEN THE BLOOD WENT INTO THE LINES. WHEN IT ACTUALLY HAPPENED TO TWO DIFFERENT PATIENTS ON THE SAME SHIFT IS WHEN THEY REALIZED THAT IT PROBABLY WAS THE LINES SO WE PULLED ALL THAT LOT NUMBER WHICH ENDED UP BEING ONLY A CASE (36) NO. PATIENTS INVOLVED: 5-6. DATE OF INCIDENT: ON (B)(6) 2023, ON (B)(6)2023, ON (B)(6)2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475314 ↗
  • MalfunctionJan 9, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 6. WE REALIZED IN A 2-3 DAY PERIOD THAT THE WHOLE SYSTEM WOULD FILL WITH AIR ONLY WHEN THE BLOOD WENT INTO THE LINES. WHEN IT ACTUALLY HAPPENED TO TWO DIFFERENT PATIENTS ON THE SAME SHIFT IS WHEN THEY REALIZED THAT IT PROBABLY WAS THE LINES SO WE PULLED ALL THAT LOT NUMBER WHICH ENDED UP BEING ONLY A CASE (36) NO. PATIENTS INVOLVED: 5-6 DATE OF INCIDENT: (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475305 ↗
  • MalfunctionJan 9, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 1 WE REALIZED IN A 2-3 DAY PERIOD THAT THE WHOLE SYSTEM WOULD FILL WITH AIR ONLY WHEN THE BLOOD WENT INTO THE LINES. WHEN IT ACTUALLY HAPPENED TO TWO DIFFERENT PATIENTS ON THE SAME SHIFT IS WHEN THEY REALIZED THAT IT PROBABLY WAS THE LINES SO WE PULLED ALL THAT LOT NUMBER WHICH ENDED UP BEING ONLY A CASE (36) NO. PATIENTS INVOLVED: 5-6 DATE OF INCIDENT: 12/23/23, 12/24/23, 12/26/23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475304 ↗
  • MalfunctionJan 9, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 2. WE REALIZED IN A 2-3 DAY PERIOD THAT THE WHOLE SYSTEM WOULD FILL WITH AIR ONLY WHEN THE BLOOD WENT INTO THE LINES. WHEN IT ACTUALLY HAPPENED TO TWO DIFFERENT PATIENTS ON THE SAME SHIFT IS WHEN THEY REALIZED THAT IT PROBABLY WAS THE LINES SO WE PULLED ALL THAT LOT NUMBER WHICH ENDED UP BEING ONLY A CASE (36) NO. PATIENTS INVOLVED: 5-6 DATE OF INCIDENT: 12/23/23, 12/24/23, 12/26/23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475303 ↗
  • MalfunctionJan 9, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 5. WE REALIZED IN A 2-3 DAY PERIOD THAT THE WHOLE SYSTEM WOULD FILL WITH AIR ONLY WHEN THE BLOOD WENT INTO THE LINES. WHEN IT ACTUALLY HAPPENED TO TWO DIFFERENT PATIENTS ON THE SAME SHIFT IS WHEN THEY REALIZED THAT IT PROBABLY WAS THE LINES SO WE PULLED ALL THAT LOT NUMBER WHICH ENDED UP BEING ONLY A CASE (36). NO. PATIENTS INVOLVED: 5-6. DATE OF INCIDENT: (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475292 ↗
  • MalfunctionJan 9, 2024Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: EVENT 4: WE REALIZED IN A 2-3 DAY PERIOD THAT THE WHOLE SYSTEM WOULD FILL WITH AIR ONLY WHEN THE BLOOD WENT INTO THE LINES. WHEN IT ACTUALLY HAPPENED TO TWO DIFFERENT PATIENTS ON THE SAME SHIFT IS WHEN THEY REALIZED THAT IT PROBABLY WAS THE LINES SO WE PULLED ALL THAT LOT NUMBER WHICH ENDED UP BEING ONLY A CASE (36) NO. PATIENTS INVOLVED: 5-6 DATE OF INCIDENT: (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475291 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 3: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D. LOT # 20255031. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301398 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 7: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20255031. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301397 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 6: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20255031. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301396 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 5: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20255031. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301395 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 2: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20255031. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301384 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 4: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20255031. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301377 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 1: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20255031. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301376 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 1: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20955010. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301276 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 2: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20955010. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301273 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 5: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20955010. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301270 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 3: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D, LOT # 20955010. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301269 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 7: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20955010. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301268 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 4: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20955010. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301267 ↗
  • MalfunctionDec 11, 2023Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE USER FACILITY: EVENT 6: BRIEF INQUIRY DESCRIPTION: SLICES IN BLOOD PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: SLICES NOTED IN BLOOD PUMP SEGMENTS OF APPROXIMATELY 7 ADDITIONAL BLOODLINES FROM SL-2000M2095D LOT # 20955010. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301266 ↗
  • MalfunctionSep 12, 2023Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: LARGE ARTERIAL CLAMP DOES NOT CLOSE COMPLETELY, CAUSING BLOOD TO LEAK DURING TAKE-OFF. NO INJURY REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17728369 ↗
  • MalfunctionJun 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED BY THE USER FACILITY: THE ARTERIAL POD IS BREAKING OFF FROM THE TUBING WHEN THEY ARE BEING PUT IN THE MACHINE (BREAKING RIGHT IN TWO) BEFORE THEY ARE EVEN IN THE ROLLER CLAMP. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17161816 ↗
  • MalfunctionJun 18, 2023Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: SCREW ON THE TUBING LEAKS FROM THE CONNECTION. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17149361 ↗
  • MalfunctionJun 14, 2023Leak/Splash; Material Deformation; Leak/Splash; Material Deformation

    AS REPORTED BY THE USER FACILITY: EVENT 3 : DESCRIPTION: VENOUS SIDE OF LINES ARE MOLDED SHUT AND WERE PARTLY OPEN WHEN PUT ON DIALYZER WILL LEAK BLOOD. THE VENOUS PORT IS THE PROBLEM. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17124229 ↗
  • MalfunctionJun 14, 2023Leak/Splash; Material Deformation; Leak/Splash; Material Deformation

    AS REPORTED BY THE USER FACILITY: EVENT 1 : DESCRIPTION: VENOUS SIDE OF LINES ARE MOLDED SHUT AND WERE PARTLY OPEN WHEN PUT ON DIALYZER WILL LEAK BLOOD. THE VENOUS PORT IS THE PROBLEM. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17124218 ↗

Other Set, tubing, blood, with and without anti-regurgitation valve devices

All Set, tubing, blood, with and without anti-regurgitation valve devices →