Skip to content

Prismaflex 5.XXRecalls & reported problems

Gambro AB · The Prismaflex control unit is a software controlled single patient device that performs Continuous Renal Replacement Therapy and Therapeutic Plasma Exchange Therapy. · Class 2

All models of Prismaflex (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jan 28, 2021Class IITerminated

    Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.

    Z-1278-2021 · Root cause: Process control · Recall details · FDA record ↗

  2. Feb 6, 2020Class IITerminated

    Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.

    Z-1829-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Oct 4, 2019Class IIOpen, Classified

    Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.

    Z-0277-2020 · Root cause: Software design · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code KDI (dialyzer, high permeability with or without sealed dialysate system) under 21 CFR 876.5860 as a Class II device. 21 CFR 876.5860 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Common questions about the Prismaflex 5.XX

Has the Prismaflex 5.XX been recalled?

Yes. 3 FDA recall records name this device's identifier: 1 still open and 2 closed (terminated) by the FDA. The most recent, Z-1278-2021 (Class II), began on Jan 28, 2021: Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the Prismaflex 5.XX, and is it still sold?

The FDA's device database lists Gambro AB as the company responsible for it (brand Prismaflex, model 5.XX). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.