Prismaflex 7.XXRecalls & reported problems
Gambro AB · The Prismaflex control unit is a software controlled single patient device that performs Continuous Renal Replacement Therapy and Therapeutic Plasma Exchange Therapy · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 2
- Injuries reported
- 1
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07332414117702), received between Jan 9, 2020 and Feb 11, 2021 — classified as 2 deaths, 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Component Missing” (1 report).
3 FDA recall records name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K131516); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 28, 2021Terminated
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Z-1279-2021 · Root cause: Process control · FDA record ↗
- Feb 6, 2020Terminated
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
Z-1831-2020 · Root cause: Under Investigation by firm · FDA record ↗
- Oct 4, 2019Open, Classified
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
Z-0279-2020 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Component Missing1
Patient problems
- Death2
- Blood Loss1
- Cardiopulmonary Arrest1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 11, 2021Component Missing; Component Missing
IT WAS REPORTED THAT DURING TREATMENT WITH A PRISMAFLEX MACHINE, THE DEVICE HAD A MISSING BLOOD PUMP CRANK. IT WAS REPORTED THAT BLOOD WAS NOT RETURNED TO THE PATIENT A TOTAL OF THREE TIMES. THE AMOUNT OF BLOOD LOSS WAS NOT REPORTED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11315064 ↗InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A ¿SELF-TEST FAILURE, CODE 4¿ ALARM WAS TRIGGERED TWICE AND A TRANSMEMBRANE PRESSURE WAS OBSERVED TO INCREASE DURING CONTINUOUS VENO-VENOUS HEMODIAFILTRATION WITH A PRISMAFLEX CONTROL UNIT AND A PRISMAFLEX SET. TROUBLESHOOTING WAS UNSUCCESSFUL AND THE TREATMENT WAS ENDED WITHOUT THE EXTRACORPOREAL BLOOD BEING RETURNED TO THE PATIENT. AS A RESULT, THE PATIENT NEEDED AN EMERGENCY INTUBATION. THE PATIENT OUTCOME WAS NOT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Blood Loss
MAUDE report 9981157 ↗DeathMar 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY, A PATIENT PASSED AWAY WHILE CONNECTED TO A PRISMAFLEX CONTROL UNIT. THE CAUSE OF DEATH WAS NOT REPORTED. IT WAS NOT REPORTED IF AN AUTOPSY WAS PERFORMED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Death
MAUDE report 9889236 ↗DeathJan 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT), THE PATIENT ¿CODED¿ AND PASSED AWAY WHILE CONNECTED TO A PRISMAFLEX CONTROL UNIT. THE CAUSE OF DEATH WAS NOT REPORTED. IT WAS NOT REPORTED IF AN AUTOPSY WAS PERFORMED. AT THE TIME OF THIS REPORT, NO FURTHER DETAILS WERE PROVIDED REGARDING PATIENT TREATMENT OR MEDICAL INTERVENTIONS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Cardiopulmonary Arrest; Death
MAUDE report 9569459 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131516. One submission often covers many models, so check the product and lot details in each record.
- Z-2066-2020 · Apr 7, 2020 · TerminatedPrismaflex Control Unit
- Z-1926-2018 · Apr 24, 2018 · TerminatedPrismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilogram
- Z-1925-2018 · Apr 24, 2018 · TerminatedPrismaflex 7.XX US, Product Code 115269 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilogram
Other Dialyzer, high permeability with or without sealed dialysate system devices
- Optiflux® 0500318EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- 2008® Series Hemodialysis Machines 190713Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls37 deaths893 reports
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
- 2008® Series Hemodialysis Machines 190766Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls15 deaths370 reports
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008® Series Hemodialysis Machines 190610Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall15 deaths335 reports
- Optiflux® 0500320EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- REVACLEAR 300Gambro Renal Products, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls125 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.