Prismaflex 7.20Recalls & reported problems
Gambro AB · The Prismaflex control unit is a software controlled single patient device that performs Continuous Renal Replacement Therapy and Therapeutic Plasma Exchange Therapy · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07332414124236), received between Apr 21, 2020 and May 5, 2025 — classified as 1 death, 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
3 FDA recall records name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K131516); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 28, 2021Terminated
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Z-1279-2021 · Root cause: Process control · FDA record ↗
- Feb 6, 2020Terminated
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
Z-1832-2020 · Root cause: Under Investigation by firm · FDA record ↗
- Oct 4, 2019Open, Classified
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
Z-0279-2020 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break2
- Activation Failure1
- Device Alarm System1
- Filtration Problem1
- Improper or Incorrect Procedure or Method1
- Inadequate Ultra Filtration1
- Protective Measures Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Cardiac Arrest1
- Death1
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2025Filtration Problem; Protective Measures Problem; Filtration Problem; Protective Measures Problem
IT AS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY WITH A PRISMAFLEX MACHINE AND AN UNKNOWN TYPE OF PRISMAFLEX SET, FOAM WAS OBSERVED IN THE DEAERATION CHAMBER, AND THE MACHINE ALARMED "FILTER CLOTTED". TWO (2) SIMILAR INCIDENTS OCCURRED DURING THE SAME DAY AND IN BOTH EVENTS BLOOD WAS NOT RETURNED TO THE PATIENT. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21964594 ↗MalfunctionNov 25, 2024Inadequate Ultra Filtration; Inadequate Ultra Filtration
IT WAS REPORTED THAT AN EXCESSIVE ULTRAFILTRATION EVENT OCCURRED ON A PRISMAFLEX MACHINE DURING HEMODIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20774372 ↗MalfunctionNov 18, 2024Break; Break
IT WAS REPORTED THAT THE SCREWS ON THE BASE OF A PRISMAFLEX MACHINE WERE OBSERVED TO BE BROKEN DURING AN UNSPECIFIED PROCESS STEP. THE DEVICE WAS AT RISK OF TIPPING OVER. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20712191 ↗InjurySep 6, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT) WITH A PRISMAFLEX CONTROL UNIT, THE PATIENT EXPERIENCED A CARDIOPULMONARY ARREST. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Cardiac Arrest
MAUDE report 15363030 ↗MalfunctionJul 27, 2022Device Alarm System; Activation Failure
CRRT MACHINE ARRIVED ON UNIT AND UPON TURNING ON RN WAS UNABLE TO START DUE TO ALARM ON SCREEN STATING "ROUTINE MAINTENANCE DUE NOW."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15112063 ↗MalfunctionSep 23, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT APPROXIMATELY ONE HOUR INTO CONTINUOUS VENO-VENOUS HEMODIALYSIS USING A PRISMAFLEX CONTROL UNIT AND A PRISMAFLEX M100 SET, A 'MAIN POWER LOST ALARM' WAS GENERATED, FOLLOWED BY A ´MALFUNCTION: VOLTAGE OUT OF RANGE' ALARM. IT WAS REPORTED THAT THE PRISMAFLEX WAS PLUGGED IN SECURELY TO THE RED OUTLET. THE TREATMENT WAS ENDED WITHOUT THE EXTRACORPOREAL (EC) BLOOD BEING MANUALLY RETURNED TO THE PATIENT AS CLOTS WERE OBSERVED IN THE TUBING. THE AMOUNT OF BLOOD LOSS WAS NOT REPORTED. IT WAS REPORTED THE A BLOOD TRANSFUSION WAS ADMINISTERED TO PATIENT. FURTHER INSPECTION OBSERVED THAT THE PLUG CONNECTING THE POWER CORD TO THE PRISMAFLEX MACHINE INSIDE THE BLACK PLASTIC CONNECTOR WAS NOT ALL THE WAY PLUGGED IN TO THE BACK OF THE MACHINE. THIS WAS CORRECTED AND TREATMENT WAS RE-STARTED WITHOUT FURTHER INCIDENTS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10574908 ↗DeathJul 6, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY WITH A PRISMAFLEX CONTROL UNIT, THE PATIENT PASSED AWAY WHILE CONNECTED TO THE MACHINE. THE CAUSE OF DEATH WAS NOT REPORTED. IT WAS NOT REPORTED IF AN AUTOPSY WAS PERFORMED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Death
MAUDE report 10236307 ↗MalfunctionApr 21, 2020Break; Break
IT WAS REPORTED THAT THE WHEEL OF A PRISMAFLEX MACHINE WAS OBSERVED TO BE BROKEN. THIS EVENT HAPPENED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 9983837 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131516. One submission often covers many models, so check the product and lot details in each record.
- Z-2066-2020 · Apr 7, 2020 · TerminatedPrismaflex Control Unit
- Z-1926-2018 · Apr 24, 2018 · TerminatedPrismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilogram
- Z-1925-2018 · Apr 24, 2018 · TerminatedPrismaflex 7.XX US, Product Code 115269 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilogram
Other Dialyzer, high permeability with or without sealed dialysate system devices
- Optiflux® 0500318EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- 2008® Series Hemodialysis Machines 190713Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls37 deaths893 reports
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
- 2008® Series Hemodialysis Machines 190766Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls15 deaths370 reports
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008® Series Hemodialysis Machines 190610Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall15 deaths335 reports
- Optiflux® 0500320EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- REVACLEAR 300Gambro Renal Products, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls125 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.