PRISMAFLEX 7.21Recalls & reported problems
Gambro AB · The Prismaflex control unit is a software controlled single patient device that performs Continuous Renal Replacement Therapy and Therapeutic Plasma Exchange Therapy. · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07332414126766), received between Dec 30, 2020 and Mar 22, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
2 FDA recall records name this device's identifier.
3 other recalls cover devices cleared under the same FDA submission (K131516); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 28, 2021Terminated
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Z-1279-2021 · Root cause: Process control · FDA record ↗
- Feb 6, 2020Terminated
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
Z-1833-2020 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- No Display/Image1
- Protective Measures Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Bradycardia1
- Low Blood Pressure/ Hypotension1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 22, 2022No Display/Image
WHILE THE PATIENT WAS UNDERGOING CRRT, THE MACHINE'S SCREEN THEN TURNED ALL RED. RN UNABLE TO RETURN BLOOD. THERE WAS NO EVIDENCE OF HARM NOTED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13847993 ↗MalfunctionJan 26, 2022Break; Break
IT WAS REPORTED DURING THE SETUP OF A PRISMAFLEX MACHINE, THE DEVICE BASE WAS OBSERVED TO BE LOOSE WHEN IT CONNECTS TO THE MAIN BODY OF THE MACHINE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13359485 ↗MalfunctionDec 1, 2021Protective Measures Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY ON A PEDIATRIC PATIENT USING A PRISMAFLEX CONTROL UNIT AND FOUR (4) UNITS OF A PRISMAFLEX HF20, SELF-TEST FAILURE (4, PRESSURE POD/SENSOR ISSUE) ALARMS WERE GENERATED AND AT LEAST ONE OF THE TREATMENT WAS ENDED WITHOUT THE EXTRACORPOREAL BLOOD BEING RETURNED TO THE PATIENT. A BLOOD TRANSFUSION WAS PERFORMED ON THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12908797 ↗InjuryDec 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT RAPIDLY REOCCURRING "EXCESSIVE TMP" ALARMS WAS GENERATED ON A PRISMAFLEX MACHINE DURING PEDIATRIC CONTINUOUS RENAL REPLACEMENT THERAPY . TREATMENT WAS DISCONTINUED AND BEFORE THE THERAPY WAS RESTARTED, THE PATIENT BECAME BRADYCARDIC AND SEVERELY HYPOTENSIVE. RESUSCITATION FLUIDS WAS PROVIDED. THE PATIENT WAS RECOVERING IN THE INTENSIVE CARE UNIT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Bradycardia; Low Blood Pressure/ Hypotension
MAUDE report 11095344 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131516. One submission often covers many models, so check the product and lot details in each record.
- Z-2066-2020 · Apr 7, 2020 · TerminatedPrismaflex Control Unit
- Z-1926-2018 · Apr 24, 2018 · TerminatedPrismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilogram
- Z-1925-2018 · Apr 24, 2018 · TerminatedPrismaflex 7.XX US, Product Code 115269 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilogram
Other Dialyzer, high permeability with or without sealed dialysate system devices
- Optiflux® 0500318EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- 2008® Series Hemodialysis Machines 190713Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls37 deaths893 reports
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
- 2008® Series Hemodialysis Machines 190766Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls15 deaths370 reports
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008® Series Hemodialysis Machines 190610Fresenius Usa, Inc. · Dialyzer, high permeability with or without sealed dialysate system1 recall15 deaths335 reports
- Optiflux® 0500320EFresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- REVACLEAR 300Gambro Renal Products, Inc. · Dialyzer, high permeability with or without sealed dialysate system3 recalls125 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.