Skip to content

GeneXpert® XPRS4PLEX-10Recalls & reported problems

Cepheid · Xpert Xpress CoV-2/Flu/RSV plus · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07332940008604), received between Dec 4, 2025 and Dec 4, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “False Negative Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 4, 2025False Negative Result; False Negative Result

    US CUSTOMER CONTACTED CEPHEID TO REPORT A QUESTIONABLE NEGATIVE RESULT FOR FLU B USING THE XPERT XPRESS COV-2/FLU/RSV PLUS ASSAY. PATIENT WAS TESTED USING XPERT XPRESS COV-2/FLU/RSV PLUS LOT 68119 WHICH RESULTED IN SARS-COV-2 NEGATIVE, FLU A NEGATIVE, FLU B NEGATIVE, RSV NEGATIVE. SAMPLE WAS PROCESSED PER IFU. THIS RESULT WAS REPORTED TO THE PHYSICIAN. TWO NASOPHARYNGEAL SWABS WERE COLLECTED FOR THIS PATIENT AT THE SAME TIME. WHEN RESULTS OF XPERT XPRESS COV-2/FLU/RSV PLUS IVD WERE REPORTED AS NEGATIVE FOR ALL TARGETS, THE CUSTOMER QUESTIONED THE FLU RESULTS SINCE THEY HIGHLY SUSPECTED THE PATIENT HAD THE FLU AS THE PATIENT PRESENTED WITH A COUGH AND FLU-LIKE SYMPTOMS. THEY USED THE SECOND SWAB TO TEST WITH THEIR SOFIA QUIDEL RAPID FLU A/FLU B TEST. SAMPLES WERE STORED AT ROOM TEMPERATURE PRIOR TO TESTING. SOFIA QUIDEL RAPID FLU A/FLU B TEST RESULTED IN FLU A NEGATIVE, FLU B POSITIVE. THIS RESULT WAS REPORTED TO THE PHYSICIAN. THE LOT NUMBER FOR SOFIA QUIDEL IS UNKNOWN. THERE WERE NO KNOWN UNDERLYING CONDITIONS AND NO KNOWN IMPACT ON THE PATIENT DUE TO THE DISCREPANCY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23720161 ↗

Other Multi-Target Respiratory Specimen Nucleic Acid Test Including Sars-Cov-2 And Other Microbial Agents devices

All Multi-Target Respiratory Specimen Nucleic Acid Test Including Sars-Cov-2 And Other Microbial Agents devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page