Navina™ Smart System Regular 69009Recalls & reported problems

Wellspect AB · Navina Smart Transanal Irrigation. 1 Navina Smart control unit, 1 Water container, 1 Tube set, 2 Rectal catheters regular, 1 Accessory set, 1 Navina case. Non Sterile Device. Rx Only. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07333387027364), received between Dec 12, 2024 and Dec 13, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Abdominal Pain1
  • Bowel Perforation1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT WITH A SPINAL CORD INJURY AT THE C5 LEVEL AND NEUROGENIC BOWEL, FOLLOWED BY CHRONIC CONSTIPATION, HAS BEEN PRESCRIBED THE NAVINA CLASSIC RECTAL IRRIGATION SYSTEM WITH A BALLOON CATHETER. THE PATIENT HAS BEEN USING THE NAVINA SYSTEM SINCE 2022, WITH A WATER IRRIGATION VOLUME OF 1000 ML. RECENTLY, THE PATIENT NOTICED SEVERAL BALLOON BURSTS IN THE RECTUM, ALL INVOLVING CATHETERS FROM THE SAME LOT NUMBER. ALTHOUGH THERE WERE NO MEDICAL COMPLICATIONS FROM THE BALLOON BURSTS, THE PATIENT EXPERIENCED A PRESSURE IN THE RECTUM WHEN THE BALLOON BURST.

    Patient problems: Abdominal Pain

    MAUDE report 20922725 ↗
  • InjuryDec 12, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    A FOREIGN INCIDENT OCCURRED OUTSIDE UNITED STATES. A 65-YEAR-OLD MALE PATIENT, WHO UNDERWENT AN INTESTINAL ANASTOMOSIS OPERATION MORE THAN TEN YEARS AGO, WAS PRESCRIBED A NAVINA SMART TRANSANAL IRRIGATION (TAI) SYSTEM WITH A RECTAL BALLOON CATHETER IN 2022. ACCORDING TO THE COMPLAINT, THE PATIENT HAD BEEN USING TRANSANAL IRRIGATION SINCE HIS INTESTINAL ANASTOMOSIS OPERATION TEN YEARS AGO, BUT IT IS UNKNOWN WHICH SYSTEM (BRAND) WAS USED BEFORE HE WAS PRESCRIBED THE NAVINA SMART TAI SYSTEM. ACCORDING TO THE INFORMATION COMMUNICATED TO THE MANUFACTURER, THE PATIENT RECEIVED AT-HOME TRAINING FROM THE THIRD-PARTY MEDICAL SUPPLY REPRESENTATIVE AND BECAME FAMILIAR WITH THE DEVICE INSTRUCTIONS FOR USE. THE PATIENT USED THE DEVICE WITHOUT ANY PROBLEM, BUT ON (B)(6), HE WAS ADMITTED TO THE HOSPITAL DUE TO A BOWEL PERFORATION. THE PERFORATION RESULTED IN AN EMERGENCY OPERATION WITH FECAL PERITONITIS, NECESSITATING RESECTION AND OSTOMY FORMATION, FOLLOWED BY THE DAILY USE OF OSTOMY AIDS AFTER THE OPERATION. AT THE TIME OF THE INCIDENT, THE RECTAL BALLOON CATHETER WAS SET TO SETTING 5, WHICH IS THE MAXIMUM BALLOON SIZE (VOLUME), WITH A WATER FLOW OF 500 ML.

    Patient problems: Bowel Perforation

    MAUDE report 20910758 ↗