Skip to content

Navina™ Smart Control Unit 69014Recalls & reported problems

Wellspect AB · Navina Smart Transanal Irrigation. 1 Navina Smart control unit with power adapter and cable. Non Sterile Device. Rx Only. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07333387028729), received between Aug 22, 2025 and May 11, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Product Quality Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026Product Quality Problem

    A NURSE REPORTED THAT THE SMART SYSTEM DID NOT ACCOUNT CORRECTLY FOR THE VOLUME OF FLUID DELIVERED TO THE PATIENT. SPECIFICALLY, THE USER OBSERVED THAT WHEN THE CONTAINER WAS FILLED TO APPROXIMATELY 650 ML AND THE SYSTEM WAS SET TO DELIVER 400 ML, THE CONTAINER BECAME EMPTY WHILE THE DEVICE DISPLAY INDICATED THAT ONLY APPROXIMATELY 0.3 L HAD BEEN DELIVERED. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25133874 ↗
  • MalfunctionAug 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A USER OF THE NAVINA SMART ELECTRONIC CONTROL UNIT, USED IN CONJUNCTION WITH THE NAVINA SMART TRANSANAL IRRIGATION SYSTEM AND RECTAL BALLOON CATHETER, REPORTED RECURRING FUNCTIONAL ISSUES DURING USE. SPECIFICALLY, THE USER EXPERIENCED REPEATED OCCURRENCES OF ERROR CODES 501 AND 502 AT VARIOUS TIMES DURING IRRIGATION SESSIONS. ACCORDING TO THE USER, THE DEVICE BECOMES UNRESPONSIVE DURING THESE EPISODES AND CANNOT BE POWERED OFF MANUALLY. THE SYSTEM MUST BE LEFT TO FULLY DISCHARGE BEFORE IT CAN BE RESTARTED. THESE MALFUNCTIONS OCCUR WHILE THE RECTAL BALLOON IS INFLATED, PREVENTING NORMAL DEFLATION VIA THE SYSTEM. AS A RESULT, THE USER MUST MANUALLY DISCONNECT ALL TUBING IN ORDER TO REMOVE THE CATHETER FROM THE RECTUM. A REPLACEMENT NAVINA SMART UNIT WAS SHIPPED TO THE USER AND CONFIRMED DELIVERED ON AUGUST 12. NO INJURIES OR MEDICAL INTERVENTIONS WERE REPORTED. THE AFFECTED DEVICE WAS NOT RETURNED, AND THE LOT NUMBER COULD NOT BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22870599 ↗

Other Enema kit devices

All Enema kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page