LoFric® 40012 (REF 4001240)Recalls & reported problems
Wellspect AB · Single Use Urinary Catheter LoFric Nelaton 16" 12FR · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07333387033433), received between Mar 21, 2025 and Feb 27, 2026 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product1
- Device Handling Problem1
- Material Deformation1
- Patient-Device Incompatibility1
Patient problems
- Alteration in Body Temperature1
- Burning Sensation1
- Chills1
- Genital Bleeding1
- No Clinical Signs, Symptoms or Conditions1
- Urinary Tract Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2026Delivered as Unsterile Product
THE ADVERSE EVENT OCCURRED OUTSIDE US, IN ITALY. A MALE PATIENT HAS BEEN USING, A STERILE, SINGLE-USE INTERMITTENT URINARY CATHETERS LOFRIC NELATON (40CM, CH12). ON (B)(6) 2026, THE PATIENT CONTACTED THE MANUFACTURER REPRESENTATIVE AND REPORTED THAT THE END OF THE PRIMARY PACKAGE WAS NOT SEALED COMPLETELY, COMPROMISING THE STERILE BARRIER OF THE PRODUCT. THE PATIENT REPORTED THAT HE HAD USED THREE PRODUCTS WITH THIS MAL-FUNCTION. FURTHER THE PATIENT REPORTED THAT HE HAD DISCARDED THE USED PRODUCTS AND THAT HE WILL RETURN THE REMAINING PRODUCTS TO THE MANUFACTURER. THE PATIENT DID NOT SUFFER ANY HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24468581 ↗InjuryDec 11, 2025Material Deformation; Material Deformation
ADVERSE EVENT OCCURRED OUTSIDE US, IN FINLAND, WHERE A MALE PATIENT OF 76 YEARS OLD CATHETERIZED HIMSELF IN THE EVENING (8:15 PM) ON (B)(6). WHEN THE CATHETER WAS INSERTED INTO THE URETHRA, A STABBING PAIN WAS FELT. THE PATIENT EXAMINED THE CATHETER AFTER IT HAD BEEN PULLED OUT. THE PATIENT NOTED A SHARP EDGE ON THE CATHETER. AFTER THE EVENT, HE EXPERIENCED A HINT OF REDNESS AND AN INTENSIFIED PAIN. AFTER 3.5 HOURS, THE PATIENT BEGAN TO FEEL UNWELL AND HAD CHILLS. THE PATIENT MEASURED THE FEVER AT 38.4° C AND AN HOUR LATER IT WAS 38.8° C. THE PATIENT ALSO MEASURED HIS BLOOD PRESSURE WHICH SHOWED AN UPPER VALUE OF 198, A LOWER VALUE OF 154, A PULSE OF 205 BEATS/MINUTE. HIS NORMAL VALUE IS AN UPPER VALUE OF 100-120, AND A LOWER VALUE OF 50-70, WITH A PULSE OF 45-60 BEATS/MINUTE. THE FOLLOWING MORNING, THE PATIENT CALLED HIS HEALTH CENTER AND WAS INSTRUCTED TO GO TO THE HOSPITAL. AT THE HOSPITAL, IT WAS CONCLUDED THAT HIS CRP VALUE WAS ELEVATED, AND HE HAD AN UTI (URINARY TRACT INFECTION). THE PATIENT WAS DISCHARGED FROM THE ER SAME DAY WITH ANTIBIOTICS AND PAIN MEDICATION. AFTER 1,5 DAY THE PATIENT FELT BETTER AND HIS CONDITION IS NOW GOOD. THE PATIENT HAS BEEN CATHETERIZING HIMSELF
Patient problems: Alteration in Body Temperature; Chills; Urinary Tract Infection
MAUDE report 23777245 ↗InjuryMar 21, 2025Patient-Device Incompatibility; Device Handling Problem; Patient-Device Incompatibility; Device Handling Problem
ADVERSE EVENT OCCURRED OUTSIDE US, IN GREAT BRITAIN. A USER REPORT FROM THE YELLOW CARD TEAM WAS RECEIVED BY MANUFACTURER ON (B)(6). THE INFORMATION IN THE USER REPORT WAS LIMITED TO "THE PATIENT STATES THAT THE DEVICE CAUSED IRRITATION TO THE URETHRA". AFTER SEVERAL ATTEMPTS, WELLSPECT WAS ABLE TO GET IN CONTACT WITH THE PATIENT IN QUESTION. HE IS A 71-YEAR-OLD MALE WITH DEMENTIA AND RECURRENT UTIS. HE ALSO PASSED KIDNEY STONES 4-5 WEEKS AGO. THE PATIENT HAS BEEN PERFORMING INTERMITTENT SELF CATHETERIZATION (ISC) FOR ABOUT 12 YEARS. THE PATIENT SAID THAT THE CATHETER FELT ROUGH AND HE EXPERIENCED BLEEDING ON WITHDRAWAL, HE NOTICED BLOOD IN THE TOILET, BUT NO BLOOD ON THE CATHETER. THE BLEEDING HAS NOW DECREASED. HE SPOKE WITH HIS GP BUT WAS NOT HOSPITALIZED NOT TREATED. CURRENTLY THE PATIENT IS WASHING AND REUSING TWO CATHETERS THAT HE THINKS IS SMOOTH, LIKELY FROM THE SAME LOT AS THE ONE THAT FELT ROUGH. THE PATIENT HAS DISCARDED THE ROUGH CATHETER.
Patient problems: Burning Sensation; Genital Bleeding
MAUDE report 21661315 ↗
Other Catheter, straight devices
- Magic3 Go Male Catheter with SureGrip 53814GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall490 reports
- Magic3 Go Male Coude Catheter with SureGrip 50814GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall3 deaths416 reports
- Magic3 Go™ Intermittent Urinary Catheter 51814C. R. Bard, Inc. · CATHETER, STRAIGHT1 recall327 reports
- Magic3 Go Male Coude Catheter with SureGrip 50816GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall239 reports
- Magic3 Go Male Catheter with SureGrip 53816GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall164 reports
- Magic3 Go Male Catheter with SureGrip 53812GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall120 reports
- Magic3 Go™ Intermittent Urinary Catheter 51812C. R. Bard, Inc. · CATHETER, STRAIGHT1 recall109 reports
- CURE CATHETER HM14Convatec, Purchasing Department · Catheter, Straight154 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page