LoFric® Hydro-Kit™ 42512Recalls & reported problems

Wellspect AB · Single Use Urinary Catheter LoFric Hydro-Kit Coudé 16" 12FR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07333387042947), received between Dec 8, 2025 and Jan 12, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product2
  • Unsealed Device Packaging2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 12, 2026Delivered as Unsterile Product; Unsealed Device Packaging

    PRODUCT PROBLEM OCCURRED OUTSIDE US, IN FRANCE. PATIENT REPORTS THAT THE WELD ON THE TIP OF THE STERILE PACKAGING IS OPEN AND THAT IN A CUSTOMER BOX CONTAINING 30 PRODUCTS, ALMOST ALL WERE DEFECTIVE. THE IMAGE RECEIVED WITH THE COMPLAINT SHOWS THAT THE WELD AT THE CATHETER TIP ON THE STERILE PACKAGE IS MISSING. WHEN THE INCLUDED WATER SACHET IS ACTIVATED, THE WETTING SOLUTION LEAKS OUT. LOFRIC HYDRO-KIT IS A STERILE SINGLE USE URINARY CATHETER. EACH PRIMARY STERILE PACKAGE CONSISTS OF A CATHETER, A WATER SACHET (WETTING SOLUTION BAG) AND A URINE COLLECTION BAG. THE WETTING SOLUTION BAG IS USED TO ACTIVATE THE HYDROPHILIC CATHETER COATING BEFORE USE. THE DEFECTIVE PRODUCTS WERE NOT USED AND THE PATIENT DID NOT REPORT ANY HARM OR HEALTH IMPACT. DEFECTIVE PRODUCTS WERE DISCARDED BY THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24035699 ↗
  • MalfunctionDec 8, 2025Delivered as Unsterile Product; Unsealed Device Packaging; Delivered as Unsterile Product; Unsealed Device Packaging

    MALFUNCTION OCCURRED OUTSIDE US, IN FRANCE, WHERE A PATIENT REPORTS THAT THE WELD ON THE STERILE PACKAGE OPENS WHEN ACTIVATING THE URINARY CATHETER. AT LEAST 10 CATHETERS FROM THE SAME CUSTOMER BOX AND LOT WERE AFFECTED, OF WHICH THE PATIENT DISCARDED 4 AND 6 CATHETERS WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE IMAGE RECEIVED WITH THIS COMPLAINT SHOWED THAT THE WELD AT THE CATHETER TIP ON THE STERILE PACKAGE IS MISSING. WHEN THE INCLUDED WATER SACHET IS ACTIVATED, THE WETTING SOLUTION LEAKS OUT AT THE BOTTOM OF THE STERILE PACKAGE. PATIENT IS A 54-YEAR-OLD MAN WITH SEVERAL YEARS OF EXPERIENCE WITH INTERMITTENT CATHETERIZATION AND HAVE BEEN USING LOFRIC HYDRO-KIT FOR MANY YEARS. THE DEFECTIVE PRODUCTS WERE NOT USED. NO INFORMATION ABOUT PATIENTS' HEALTH CONDITION OF REASON FOR USING INTERMITTENT CATHETERS HAS BEEN COMMUNICATED. LOFRIC HYDRO-KIT IS A STERILE SINGLE USE URINARY CATHETER. EACH PRIMARY STERILE PACKAGE CONSISTS OF A CATHETER, A WATER SACHET (WETTING SOLUTION BAG) AND A URINE COLLECTION BAG. THE WETTING SOLUTION BAG IS USED TO ACTIVATE THE HYDROPHILIC CATHETER COATING BEFORE USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23741324 ↗

Other Catheter, straight devices