Skip to content

LoFric® Primo™ 41012Recalls & reported problems

Wellspect AB · Single Use Urinary Catheter LoFric Primo Nelaton 16" 12FR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07333387043005), received between Oct 17, 2024 and Oct 22, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Delivered as Unsterile Product1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 22, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    THE ADVERSE EVENT OCCURRED OUTSIDE US, IN GREAT BRITAIN. A PATIENT REPORTED ON (B)(6) 2025, THAT SEVERAL DEVICES WERE RECEIVED BY THE PATIENT WITH BROKEN PRIMARY PACKAGING SEAL, COMPROMISING THE STERILE BARRIER OF THE DEVICE. THE DEVICE IS A STERILE, SINGLE-USE INTERMITTENT URINARY CATHETER, LOFRIC PRIMO NELATON MALE 40CM CH12. PRODUCTS WITH TWO DIFFERENT LOT NUMBERS WERE INVOLVED. LOT: 664407, MANUFACTURING DATE: 17 JANUARY 2025, EXPIRY DATE: 17 JULY 2028, LOT: 669391, MANUFACTURING DATE: 13 JUNE 2025, EXPIRY DATE: 11 DECEMBER 2028, IT IS UNKNOWN IF THE PATIENT USED ANY OF THE DEVICES. NO HARM IS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23355811 ↗
  • MalfunctionOct 31, 2024Break; Break

    THE ADVERSE EVENT OCCURRED OUTSIDE US, IN (B)(6). THE DEVICE LOFRIC PRIMO INCLUDES A WETTING SOLUTION BAG (SACHET) THAT IS USED TO ACTIVATE THE HYDROPHILIC CATHETER COATING BEFORE USE. THIS IS DONE BY SQUEEZING THE INTEGRATED BAG TO RELEASE THE WETTING SOLUTION WITH THE CATHETER IN ITS PACKAGE. WHEN THE USER SQUEEZED THE SACHET TO ACTIVATE THE COATING OF THE CATHETER, THE WETTING SOLUTION CAME OUT FROM THE SIDE OF THE CATHETER PACKAGE, INDICATING THAT THE PACKAGE'S STERILE BARRIER WAS COMPROMISED. THE CATHETERS WERE NOT USED, AND THERE WAS NO HARM OR CLINICAL IMPACT ON THE USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20579971 ↗
  • MalfunctionOct 17, 2024Break; Product Quality Problem; Break; Product Quality Problem

    ADVERSE EVENT OCCURRED OUTSIDE US, IN ITALY. LOFRIC PRIMO INCLUDES A WETTING SOLUTION BAG (SACHET) THAT IS USED TO ACTIVATE THE HYDROPHILIC CATHETER COATING BEFORE USE. THIS IS DONE BY SQUEEZING THE INTEGRATED BAG TO RELEASE THE WETTING SOLUTION WITH THE CATHETER IN ITS PACKAGE. WHEN THE PATIENT SQUEEZED THE SACHET TO ACTIVATE THE COATING OF THE CATHETER, THE WETTING SOLUTION CAME OUT FROM THE SIDE OF THE CATHETER PACKAGE, INDICATING THAT THE PACKAGE'S STERILE BARRIER WAS COMPROMISED. THIS OCCURRED WITH 60 CATHETERS. THE CATHETERS WERE NOT USED, AND THERE WAS NO HARM OR CLINICAL IMPACT ON THE USER. THE PATIENT COULD PERFORM THE INTERMITTENT CATHETERIZATION (IC) USING PRODUCTS FROM ANOTHER BOX. THE PATIENT HAS BEEN CATHETERISING WITH THE SAME CATHETER FOR SEVERAL YEARS AND HAS NEVER EXPERIENCED ANY PROBLEMS BEFORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20469259 ↗

Other Catheter, straight devices

All Catheter, straight devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page