The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07333387043456), received between Jan 9, 2025 and Dec 11, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information1
- Product Quality Problem1
Patient problems
- Inflammation1
- Urinary Tract Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 11, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THIS ADVERSE EVENT OCCURRED IN THE USA. A MALE PATIENT WITH THE MEDICAL CONDITION URINARY RETENTION, BPH, CONSTIPATION AND DIABETES, HAS BEEN USING A STERILE, SINGLE-USE, INTERMITTENT URINARY CATHETER, LOFRIC ORIGO COUDÉ (16'' 16FR). ON (B)(6) 2025, THE PATIENT INFORMED THE MANUFACTURER REPRESENTATIVE THAT HE HAD AN INCIDENT WITH THE CATHETER 2 DAYS AGO. THE PATIENT, WHO IS NEW TO THE THERAPY INTERMITTENT CATHETERIZATION, HAD PROBLEMS INSERTING THE CATHETER AND HE EXPERIENCED IT VERY RESISTANT TO INSERT. HE STATED HE DID EXPERIENCE PAIN WHEN HE MET THE RESISTANCE. AFTER HE HAD REMOVED THE CATHETER, SEVERE BLEEDING OCCURRED (PROBABLY A FRUIT JUICE GLASS SIZE OF BLOOD, ACCORDING TO THE PATIENT). THE PATIENT WENT TO THE HOSPITAL EMERGENCY DUE TO THE BLEEDING. HE STATED THAT THE HOSPITAL PLACED A FOLEY CATHETER AND TOLD HIM THAT HIS PROSTATE WAS SWOLLEN AND "PINCHED OFF THE URETHRA." HE WAS THEN RELEASED HOME. THE ONLY TREATMENT HE HAD WAS A FOLEY CATHETER INSERTED. HE STATED HE WILL HAVE THE FOLEY UNTIL ON (B)(6) AS THAT'S THE SOONEST THE UROLOGIST COULD SEE HIM. THE MANUFACTURER REPRESENTATIVE PROVIDED EDUCATION FOR IC TECHNIQUE, WAYS TO RELAX THE SPHINCTER AND EASE THE CATHETER IN,
Patient problems: Inflammation
MAUDE report 23776510 ↗MalfunctionJan 9, 2025Product Quality Problem; Product Quality Problem
A 73-YEAR-OLD MAN HAS BEEN USING LOFRIC ORIGO WITH TIEMANN (COUDE) TIP DUE TO PROBLEMS WITH URINARY RETENTION (NEUROGENIC BLADDER) AS WELL AS ENLARGED PROSTATE. THE USER EXPERIENCED BLOOD IN THE URINE AND ON THE TIP OF THE CATHETER AFTER USE. HE SOUGHT MEDICAL ATTENTION AND HAD A URINE CULTURE DONE TO DETERMINE THE BACTERIA COUNT. TO TREAT A URINARY TRACT INFECTION (UTI), HE WAS PRESCRIBED ANTIBIOTICS 3 TIMES DAILY FOR 14 DAYS. IN ADDITION, HE WILL TAKE LOW DOSE ANTIBIOTIC 1 TIMES DAILY INDEFINITELY TO FURTHER REDUCE RISKS OF UTI'S. NO OTHER TESTING, PROCEDURES, OR TREATMENTS WERE PERFORMED. ACCORDING TO THE PATIENT THE HARM IS RELATED DUE TO MISALIGNMENT BETWEEN THE CATHETER BENT TIP AND THE INDICATOR ON THE CATHETER FUNNEL WHICH HAS SCORED HIS URETHRA AND LET TO UTI.
Patient problems: Urinary Tract Infection
MAUDE report 21109464 ↗
Other Catheter, straight devices
- Magic3 Go Male Catheter with SureGrip 53814GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall490 reports
- Magic3 Go Male Coude Catheter with SureGrip 50814GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall3 deaths416 reports
- Magic3 Go™ Intermittent Urinary Catheter 51814C. R. Bard, Inc. · CATHETER, STRAIGHT1 recall327 reports
- Magic3 Go Male Coude Catheter with SureGrip 50816GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall239 reports
- Magic3 Go Male Catheter with SureGrip 53816GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall164 reports
- Magic3 Go Male Catheter with SureGrip 53812GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall120 reports
- Magic3 Go™ Intermittent Urinary Catheter 51812C. R. Bard, Inc. · CATHETER, STRAIGHT1 recall109 reports
- CURE CATHETER HM14Convatec, Purchasing Department · Catheter, Straight154 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page