Skip to content

Leksell Vantage Stereotactic System 1055007Recalls & reported problems

Elekta Instrument AB · ARC CARRIER · Class 2

All models of Leksell Vantage Stereotactic System (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07340048304887), received between Mar 24, 2020 and Mar 24, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (1 report).

1 FDA recall record name this device's identifier.

1 other recall cover devices cleared under the same FDA submission (K171123); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 30, 2020Class IITerminated

    A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System.

    Z-1826-2020 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2020Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    DURING AN INTERNAL INSPECTION OF TWO VANTAGE ARC SYSTEMS, IT WAS FOUND THAT THE ARC CARRIER WAS UNABLE TO LOCK SECURELY.

    Patient problems: No Patient Involvement

    MAUDE report 9872404 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171123. One submission often covers many models, so check the product and lot details in each record.

  • Z-2442-2020 · Jun 10, 2020 · Terminated
    Leksell Vantage Stereotactic System, UDI/GTIN 7340048304887

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page