Leksell Vantage Stereotactic System recallZ-1826-2020 · Elekta Inc · initiated Mar 30, 2020

Class II Terminated All recalls posted in April 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System.

FDA root cause category: Process design

What the company is doing

Elekta Inc., notified customers on 04/01/2020 via Urgent Important Field Safety Notification letter. The letter identified the affected products and customers were instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible but no later than 30 days from receipt. Elekta will replace all defective Instrument Carriers. Replacement units will be sent to affected facilities as soon as available. The firm sent an additional letter on 05/06/2020 with the following information: All Instrument Carriers included in this IFSN will be replaced by Elekta. While you await your replacement, Elekta has formulated an additional mandatory in-field test that must be carried out on every defective device. This test must be carried out by a trained technician/engineer or equivalent at your hospital or by Elekta, and the defective device must pass the full test for the continued use of the product. Thus, if your defective Instrument Carrier should fail to pass this test, it must be immediately removed from use.

Recall details

Recall number
Z-1826-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 30, 2020
Posted by the FDA
Apr 29, 2020
Recalling firm
Elekta Inc
Quantity
25 units.
Product code
HAW
FDA submissions
K171123, K190887
Distribution
US nationwide distributions in the states of MA, NH, NC, UT, OH, FL, MN, NY, LA and MT.

Product description

Leksell Vantage Stereotactic System. Instrument Carrier, Part of Leksell Vantage Arc (1053958). Neurological stereotaxic Instrument

Code information (models, lots, serial numbers)

Serial Numbers:SH00106, SH00120, SH00149, SH00160, SH00165, SH00110, SH00143, SH00150, SH00161, SH00166, SH00111, SH00144, SH00157, SH00162, SH00167, SH00112, SH00147, SH00158, SH00163, SH00168, SH00114, SH00148, SH00159, SH00164 and SH00178. UDI 7340048304887.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

9 devices cleared under K171123, K190887 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page