Skip to content

RayStation 11B SP2Recalls & reported problems

Raysearch Laboratories AB (Publ) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
4

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Nov 28, 2025Class IIOpen, Classified

    Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

    Z-1104-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jan 16, 2025Class IIOpen, Classified

    Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

    Z-1190-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Mar 28, 2024Class IIOpen, Classified

    Potential for reported SSD to be too high.

    Z-1749-2024 · Root cause: Software design · Recall details · FDA record ↗

  4. Feb 23, 2024Class IIOpen, Classified

    A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.

    Z-1482-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other System, Planning, Radiation Therapy Treatment devices

All System, Planning, Radiation Therapy Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page