RayStation Software Version Numbers version numbers 8B, 9A… recallZ-1482-2024 · Raysearch Laboratories AB · initiated Feb 23, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.

FDA root cause category: Under Investigation by firm

What the company is doing

On February 26, 2024, the firm notified customers of the issue via Urgent Field Safety Notice: Medical Device Correction letters via email. Actions to be taken by the user: 1. Educate planning staff and all users about the possibility of the described use error. 2. If the use error is deemed likely to occur at your clinic, take actions to ensure that proper review of assigned materials is performed before a patient model is used for treatment planning. 3. If you wish to have the Silicon chemical element removed from the material list, contact RaySearch support for assistance. This is especially recommended for clinics that perform ion beam treatments, where the harm caused by the use error could be significant. 4. Inspect your product and identify all installed units with the above software version number(s). 5. Confirm you have read and understood this notice by replying to the notification email. The Silicon chemical element will not be included in the next version of RayStation/RayPlan, scheduled for market release in April 2024 (subject to market clearance in some markets). If requested by the customer, it can be removed from the material list of existing installations. This notice needs to be passed on to all those who need to be aware within your organization. Maintain awareness of this notice as long as any affected version is in use. Thank you for your cooperation, and we apologize for any inconvenience. For regulatory information, please contact [email protected].

Recall details

Recall number
Z-1482-2024
Status
Open, Classified
Initiated by the company
Feb 23, 2024
Posted by the FDA
Apr 5, 2024
Recalling firm
Raysearch Laboratories AB
Quantity
154 units
Product code
MUJ
FDA submissions
K190387, K200569, K210645, K211867, K220141, K222312
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hong Kong S.A.R., Hungary, India, Indonesia, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, P.R. China, Qatar, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom.

Product description

RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Version Number (Build Number/UDI): 8B (8.1.0.47/07350002010129), 8B SP1 (8.1.1.8/07350002010204), 8B SP2 (8.1.2.5/07350002010235), 9A (9.0.0.113/07350002010174), 9A SP1 (9.0.1.142/07350002010488), 9B (9.1.0.933/07350002010266), 9B SP1 (9.2.0.483/07350002010297), 10A (10.0.0.1154/07350002010303), 10A SP1 (10.0.1.52/07350002010365), 10A SP2 (10.0.2.10/07350002010655), 10B (10.1.0.613/07350002010310), 10B SP1 (10.1.1.54/07350002010471), 11A (11.0.0.951/07350002010389), 11A SP1 (11.0.1.29/07350002010433), 11A SP2 (11.0.3.116/07350002010440), 11A SP3 (11.0.4.15/07350002010631), 11B (12.0.0.932/07350002010426), 11B SP1 (12.1.0.1221/07350002010495), 11B SPC1 (12.1.1.41/07350002010587), 11B SP2 (12.0.3.68/07350002010501), 11B SPC2 (12.1.2.91/07350002010617), 11B SP3 (12.0.4.12/07350002010600), 11B SPC3 (12.1.3.162/07350002010662), 11B SPC4 (12.1.4.113/07350002010709), 11B SPC5 (12.1.5.60/07350002010747), 11B SPT1 (12.3.0.119/07350002010570), 12A (13.0.0.1547/07350002010549), 12A SP1 (13.1.0.144/07350002010679), 12A SP2 (13.1.1.89/07350002010730), 2023B (14.0.0.3338/07350002010556), 2024A (15.0.0.430/07350002010723).

Devices named in this recall

20 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page