Inserter Adapter OTD Straight 1002Recalls & reported problems

Ortoma AB (Publ) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07350137521019), received between Jul 25, 2025 and Jul 25, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 28, 2025Open, Classified

    Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.

    Z-2615-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2025Fracture; Fracture

    A CUSTOMER HAS REPORTED A COMPLAINT REGARDING REF 30-201 INSERTER ADAPTER OTD STRAIGHT. THE ISSUE WAS A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER OF INSERTER ADAPTER OTD STRAIGHT DURING A SURGICAL PROCEDURE. THE ADAPTERS ARE INTENDED TO BE USED AS A PART OF OTS HIP FOR REAL TIME TRACKING OF INSTRUMENTS RELATIVE TO THE PATIENT´S ANATOMY DURING ORTHOPEDIC SURGERY. THE INSERTER ADAPTERS ARE INTENDED TO BE ATTACHED TO A CUP INSERTER TOOL. THE WELD BREAKAGE OCCURRED WHILE PERFORMING SURGICAL IMPACTIONS USING JOHNSON & JOHNSON MEDTECH EMPHASYS STRAIGHT IMPACTOR (THE CUP INSERTER TOOL), 4812-01-150. THE SURGICAL TEAM NOTICED THE BREAKAGE AND THE PIN COULD BE PULLED OUT. TO PROCEED WITH THE SURGERY, THE TEAM STARTED TO USE A SECOND REF 30-201 INSERTER ADAPTER OTD STRAIGHT (PREPARED AS A BACK-UP ITEM, AVAILABLE IN OR). DURING THE SAME SURGICAL PROCEDURE, THE SECOND ADAPTER FAILED DURING IMPACTIONS, USING THE CUP INSERTER TOOL. ALSO THIS FAILURE CONSISTED OF A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER. THE SURGICAL PROCEDURE COULD CONTINUE BY REVERTING TO CONVENTIONAL SURGERY. NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION WERE REPORTED. IMAGES WERE PROVIDED T

    Patient problems: Insufficient Information

    MAUDE report 22613203 ↗
  • MalfunctionJul 25, 2025Fracture; Fracture

    A CUSTOMER HAS REPORTED A COMPLAINT REGARDING REF 30-201 INSERTER ADAPTER OTD STRAIGHT. THE ISSUE WAS A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER OF INSERTER ADAPTER OTD STRAIGHT DURING A SURGICAL PROCEDURE. THE ADAPTERS ARE INTENDED TO BE USED AS A PART OF OTS HIP FOR REAL TIME TRACKING OF INSTRUMENTS RELATIVE TO THE PATIENT´S ANATOMY DURING ORTHOPEDIC SURGERY. THE INSERTER ADAPTERS ARE INTENDED TO BE ATTACHED TO A CUP INSERTER TOOL. THE WELD BREAKAGE OCCURRED WHILE PERFORMING SURGICAL IMPACTIONS USING JOHNSON & JOHNSON MEDTECH EMPHASYS STRAIGHT IMPACTOR (THE CUP INSERTER TOOL), 4812-01-150. THE SURGICAL TEAM NOTICED THE BREAKAGE AND THE PIN COULD BE PULLED OUT. TO PROCEED WITH THE SURGERY, THE TEAM STARTED TO USE A SECOND REF 30-201 INSERTER ADAPTER OTD STRAIGHT (PREPARED AS A BACK-UP ITEM, AVAILABLE IN OR). DURING THE SAME SURGICAL PROCEDURE, THE SECOND ADAPTER FAILED DURING IMPACTIONS, USING THE CUP INSERTER TOOL. ALSO THIS FAILURE CONSISTED OF A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER. THE SURGICAL PROCEDURE COULD CONTINUE BY REVERTING TO CONVENTIONAL SURGERY. NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION WERE REPORTED. IMAGES WERE PROVIDED T

    Patient problems: Insufficient Information

    MAUDE report 22613017 ↗