Inserter Adapter OTD Straight 1002Recalls & reported problems
Ortoma AB (Publ) · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07350137521019), received between Jul 25, 2025 and Jul 25, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 28, 2025Open, Classified
Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.
Z-2615-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture2
Patient problems
- Insufficient Information2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2025Fracture; Fracture
A CUSTOMER HAS REPORTED A COMPLAINT REGARDING REF 30-201 INSERTER ADAPTER OTD STRAIGHT. THE ISSUE WAS A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER OF INSERTER ADAPTER OTD STRAIGHT DURING A SURGICAL PROCEDURE. THE ADAPTERS ARE INTENDED TO BE USED AS A PART OF OTS HIP FOR REAL TIME TRACKING OF INSTRUMENTS RELATIVE TO THE PATIENT´S ANATOMY DURING ORTHOPEDIC SURGERY. THE INSERTER ADAPTERS ARE INTENDED TO BE ATTACHED TO A CUP INSERTER TOOL. THE WELD BREAKAGE OCCURRED WHILE PERFORMING SURGICAL IMPACTIONS USING JOHNSON & JOHNSON MEDTECH EMPHASYS STRAIGHT IMPACTOR (THE CUP INSERTER TOOL), 4812-01-150. THE SURGICAL TEAM NOTICED THE BREAKAGE AND THE PIN COULD BE PULLED OUT. TO PROCEED WITH THE SURGERY, THE TEAM STARTED TO USE A SECOND REF 30-201 INSERTER ADAPTER OTD STRAIGHT (PREPARED AS A BACK-UP ITEM, AVAILABLE IN OR). DURING THE SAME SURGICAL PROCEDURE, THE SECOND ADAPTER FAILED DURING IMPACTIONS, USING THE CUP INSERTER TOOL. ALSO THIS FAILURE CONSISTED OF A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER. THE SURGICAL PROCEDURE COULD CONTINUE BY REVERTING TO CONVENTIONAL SURGERY. NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION WERE REPORTED. IMAGES WERE PROVIDED T
Patient problems: Insufficient Information
MAUDE report 22613203 ↗MalfunctionJul 25, 2025Fracture; Fracture
A CUSTOMER HAS REPORTED A COMPLAINT REGARDING REF 30-201 INSERTER ADAPTER OTD STRAIGHT. THE ISSUE WAS A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER OF INSERTER ADAPTER OTD STRAIGHT DURING A SURGICAL PROCEDURE. THE ADAPTERS ARE INTENDED TO BE USED AS A PART OF OTS HIP FOR REAL TIME TRACKING OF INSTRUMENTS RELATIVE TO THE PATIENT´S ANATOMY DURING ORTHOPEDIC SURGERY. THE INSERTER ADAPTERS ARE INTENDED TO BE ATTACHED TO A CUP INSERTER TOOL. THE WELD BREAKAGE OCCURRED WHILE PERFORMING SURGICAL IMPACTIONS USING JOHNSON & JOHNSON MEDTECH EMPHASYS STRAIGHT IMPACTOR (THE CUP INSERTER TOOL), 4812-01-150. THE SURGICAL TEAM NOTICED THE BREAKAGE AND THE PIN COULD BE PULLED OUT. TO PROCEED WITH THE SURGERY, THE TEAM STARTED TO USE A SECOND REF 30-201 INSERTER ADAPTER OTD STRAIGHT (PREPARED AS A BACK-UP ITEM, AVAILABLE IN OR). DURING THE SAME SURGICAL PROCEDURE, THE SECOND ADAPTER FAILED DURING IMPACTIONS, USING THE CUP INSERTER TOOL. ALSO THIS FAILURE CONSISTED OF A WELD BREAKAGE BETWEEN THE PIN AND INSERTER HOLDER. THE SURGICAL PROCEDURE COULD CONTINUE BY REVERTING TO CONVENTIONAL SURGERY. NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION WERE REPORTED. IMAGES WERE PROVIDED T
Patient problems: Insufficient Information
MAUDE report 22613017 ↗
Other Orthopedic Stereotaxic Instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page