Skip to content

POLARCUP 75023344Recalls & reported problems

Smith & Nephew, Inc. · POLARCUP REDUCER PART FOR IMPACTOR · Class 1

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07611996112337), received between Mar 17, 2021 and Dec 3, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Fracture2
  • Deformation Due to Compressive Stress1
  • Flaked1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2024Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE TIP OF A POLARCUP REDUCER PART FOR IMPACTOR BROKE WHILE BEING HAMMERED DURING REDUCTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20833249 ↗
  • MalfunctionMar 21, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR, THE HEAD PUSHER TIP OF ONE (1) POLARCUP REDUCER PART FOR IMPACTOR BROKE IN HALF WHILE IMPACTING THE FEMORAL HEAD. ONE PIECE FELL ON THE FLOOR AND THE OTHER FELL INTO THE INCISION AND WAS RETRIEVED BY HAND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18956618 ↗
  • MalfunctionMar 21, 2023Break; Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE TIP OF THE POLARCUP REDUCER PART FOR IMPACTOR BROKE WHILE USING. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, CHANGING SURGICAL TECHNIQUE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16584053 ↗
  • MalfunctionJun 25, 2022Break; Flaked; Break; Flaked

    IT WAS REPORTED THAT, A PORTION OF THE POLARCUP REDUCER PART FOR IMPACTOR WAS CHIPPED. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14819227 ↗
  • InjuryMay 26, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, A POLARCUP REDUCER PART FOR IMPACTOR BROKE INSIDE PATIENT DURING IMPACTION. ALL PIECES WHERE RETRIEVED BY HAND. THE PROCEDURE WAS COMPLETED, WITH A DELAY LESS THAN 30 MIN, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14519789 ↗
  • MalfunctionMar 17, 2021Break; Break

    IT WAS REPORTED THAT DURING INSPECTION OF A TOTAL HIP REPLACEMENT THE POLARCUP REDUCER PART FOR IMPACTOR WAS MISSING FORM TRAY DUE TO BEING BROKEN AND DISCARDED. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11500023 ↗

Other IMPACTOR devices

All IMPACTOR devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.