POLARCUP 75023344Recalls & reported problems
Smith & Nephew, Inc. · POLARCUP REDUCER PART FOR IMPACTOR · Class 1
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07611996112337), received between Mar 17, 2021 and Dec 3, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Fracture2
- Deformation Due to Compressive Stress1
- Flaked1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 3, 2024Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THE TIP OF A POLARCUP REDUCER PART FOR IMPACTOR BROKE WHILE BEING HAMMERED DURING REDUCTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20833249 ↗MalfunctionMar 21, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING A THR, THE HEAD PUSHER TIP OF ONE (1) POLARCUP REDUCER PART FOR IMPACTOR BROKE IN HALF WHILE IMPACTING THE FEMORAL HEAD. ONE PIECE FELL ON THE FLOOR AND THE OTHER FELL INTO THE INCISION AND WAS RETRIEVED BY HAND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18956618 ↗MalfunctionMar 21, 2023Break; Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress
IT WAS REPORTED THAT, DURING A THR SURGERY, THE TIP OF THE POLARCUP REDUCER PART FOR IMPACTOR BROKE WHILE USING. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, CHANGING SURGICAL TECHNIQUE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16584053 ↗MalfunctionJun 25, 2022Break; Flaked; Break; Flaked
IT WAS REPORTED THAT, A PORTION OF THE POLARCUP REDUCER PART FOR IMPACTOR WAS CHIPPED. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14819227 ↗InjuryMay 26, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING A THR SURGERY, A POLARCUP REDUCER PART FOR IMPACTOR BROKE INSIDE PATIENT DURING IMPACTION. ALL PIECES WHERE RETRIEVED BY HAND. THE PROCEDURE WAS COMPLETED, WITH A DELAY LESS THAN 30 MIN, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14519789 ↗MalfunctionMar 17, 2021Break; Break
IT WAS REPORTED THAT DURING INSPECTION OF A TOTAL HIP REPLACEMENT THE POLARCUP REDUCER PART FOR IMPACTOR WAS MISSING FORM TRAY DUE TO BEING BROKEN AND DISCARDED. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11500023 ↗
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,295 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,539 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,498 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,456 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall875 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR716 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR704 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR337 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.