POLARSTEM 75023710Recalls & reported problems

Smith & Nephew, Inc. · POLARSTEM BALL HEAD IMPACTOR · Class 1

All models of POLARSTEM (63)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07611996114386), received between Nov 10, 2020 and Sep 23, 2025 — classified as 2 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • Break2
  • Crack2
  • Difficult to Remove2
  • Deformation Due to Compressive Stress1
  • Detachment of Device or Device Component1
  • Mechanical Jam1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2025Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING A THR SURGERY ONE (1) POLARSTEM MODULAR HEAD FOR 21000662 CAME OFF FROM ONE (1) POLARSTEM BALL HEAD IMPACTOR AND GOT WEDGED BEHIND THE ACETABULUM. THE POLARSTEM MODULAR HEAD FOR 21000662 WAS NOT RETAINING WELL ENOUGH ON THE HANDLE AND WAS RETRIEVED WITH DIFFICULTY USING LONG FORCEPS. THE PROCEDURE WAS RESUMED AFTER A SIGNIFICANT SURGICAL DELAY GREATER THAN 30 MINUTES USING A COMPETITOR DEVICE (STRYKER HEAD PUSHER). NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23123947 ↗
  • MalfunctionJan 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A THR SURGERY, DURING HANDLING ACTIVITIES, THE POLARSTEM BALL HEAD IMPACTOR CAME OFF. THE SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH A S+N BACK-UP DEVICE. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21167496 ↗
  • MalfunctionSep 9, 2022Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE POLARSTEM BALL HEAD IMPACTOR PLASTIC BROKE UP DURING IMPACTION OF THE FEMORAL HEAD. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15393561 ↗
  • MalfunctionAug 17, 2022Fracture; Fracture

    IT WAS REPORTED THAT, THE POLARSTEM BALL HEAD IMPACTOR TIP WAS BROKEN. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15241722 ↗
  • MalfunctionFeb 16, 2022Break; Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE POLARSTEM BALL HEAD IMPACTOR BROKE. NO PIECES FELL INTO PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH AND NEPHEW BACK-UP DEVICE. PATIENT WAS NOT HARMED, OR FURTHER CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13543317 ↗
  • MalfunctionJan 27, 2022Crack; Crack

    IT WAS REPORTED THAT THE POLARSTEM BALL HEAD IMPACTOR IS CRACKED. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13372183 ↗
  • MalfunctionAug 3, 2021Crack; Fracture; Crack; Fracture

    IT WAS REPORTED THAT A POLARSTEM MODULAR HEAD FOR 21000662 CRACKED. IT IS UNKNOWN WHEN THIS WAS PROBLEM WAS NOTICED. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12267473 ↗
  • MalfunctionMay 21, 2021Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE TIP OF A POLARSTEM BALL HEAD IMPACTOR IS JAMMED. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. AS THIS WAS NOTICED AFTER SURGERY, THE PATIENT WAS NO LONGER INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11862440 ↗
  • MalfunctionMay 17, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT AFTER PROCEDURE, WAS NOTICED THAT THE BLACK BUMPER TIP THAT CONNECTS TO THE POLARSTEM BALL HEAD IMPACTOR ON WAS FOUND TOO STIFF TO REMOVE ON TWO DEVICES AFTER USE FOR CLEANING (B)(4). THERE WAS NO PATIENT INVOLVED SINCE THE EVENT WAS NOTICED AFTER PROCEDURE. TWO POLARSTEM MODULAR HEAD FOR 21000662 (B)(4) WERE RECEIVED, IT IS UNKNOWN WHAT TYPE OF FAILURE DO THESE DEVICES HAVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11838434 ↗
  • MalfunctionMay 17, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT AFTER PROCEDURE, WAS NOTICED THAT THE BLACK BUMPER TIP THAT CONNECTS TO THE POLARSTEM BALL HEAD IMPACTOR ON WAS FOUND TOO STIFF TO REMOVE ON TWO DEVICES AFTER USE FOR CLEANING (B)(4). THERE WAS NO PATIENT INVOLVED SINCE THE EVENT WAS NOTICED AFTER PROCEDURE. TWO POLARSTEM MODULAR HEAD FOR 21000662 (B)(4) WERE RECEIVED, IT IS UNKNOWN WHAT TYPE OF FAILURE DO THESE DEVICES HAVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11838407 ↗
  • InjuryMar 15, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE IMPACTOR (CASE-(B)(4)) BROKE DURING USE, DUE TO WEAR AND TEAR. THE PROCEDURE WAS COMPLETED USING A SN BACKUP DEVICE. NO INJURY TO THE PATIENT WAS REPORTED. A DELAY OF 30 MIN OR LESS WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11488230 ↗
  • InjuryNov 10, 2020Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR PROCEDURE THE POLARSTEM BALL HEAD IMPACTOR BROKE WHILE IMPACTION INSIDE THE PATIENT. SURGERY WAS COMPLETED WITH A CHANGE IN THE SURGICAL TECHNIQUE. NO SURGICAL DELAYS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10815706 ↗

Other IMPACTOR devices

All IMPACTOR devices →