GAMMA 1320-1120Recalls & reported problems
Stryker GmbH · Speedlock Sleeve · Class 2
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613154893669), received between Aug 26, 2025 and Mar 30, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Failure to Align” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Align2
- Device Difficult to Setup or Prepare1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Failure to Align; Material Twisted/Bent
AS REPORTED: "DURING THE ATTEMPT TO PLACE THE DISTAL LOCKING SCREW OF THE GAMMA 3 NAIL, TWO DRILL BITS BROKE BECAUSE DRILLING OCCURRED ONTO THE NAIL. THE TARGETING ARM AND SLEEVE BECAME BENT. THE MEDICAL TREATMENT COULD BE COMPLETED BECAUSE REPLACEMENT INSTRUMENTS WERE AVAILABLE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24710390 ↗MalfunctionAug 26, 2025Device Difficult to Setup or Prepare; Failure to Align; Device Difficult to Setup or Prepare; Failure to Align
AS REPORTED: "G3 200 SPEEDLOCK SLEEVE NO LONGER SITS CORRECTLY ON THE INSERTION GUIDE, RESULTING IN AN INCREASED NUMBER OF INCORRECT HOLES WHEN POSITIONING THE K-WIRE FOR THE FEMORAL NECK SCREW AND THE DISTAL LOCKING SCREW. IN SOME CASES, IT IS NO LONGER POSSIBLE TO FULLY FIX THE GUIDE SLEEVES."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22893596 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page