Skip to content

Alaris, SmartSite 42290ERecalls & reported problems

BD Switzerland Sàrl · Gravity Set Check Valve 2 Needle-Free Valves Nonvented · Class 2

All models of Alaris, SmartSite (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613203010931), received between Jul 30, 2021 and Nov 22, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Complete Blockage” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Blockage3
  • Device Contamination with Chemical or Other Material1
  • Free or Unrestricted Flow1
  • Leak/Splash1
  • Material Separation1
  • Restricted Flow rate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Free or Unrestricted Flow; Free or Unrestricted Flow

    IT WAS REPORTED THAT BD ALARIS SMARTSITE GRAVITY SET HAD AIR IN LINE. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: THERE IS AN AIR BUBBLE IN THE TUBING WHEN THE FLUID IS CONNECTED, AND ANAESTHETIC TECH NEEDS TO USE A SYRINGE TO RID THE AIR OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20759625 ↗
  • MalfunctionAug 8, 2023Material Separation; Material Separation

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17484266 ↗
  • MalfunctionApr 10, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THE BD ALARIS SMARTSITE GRAVITY SET EXPERIENCED FOREIGN MATTER, CONTAMINATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A LARGE MOBILE FOREIGN BODY (APPEARS TO BE A PIECE OF METAL) WAS DISCOVERED IN A IV GIVING SET PRIOR TO PATIENT USE BUT AFTER IT HAD BEEN RUN THROUGH. LINE HAS FLUID IN IT BUT HAS NOT BEEN USED ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16708712 ↗
  • MalfunctionMay 5, 2022Complete Blockage; Complete Blockage

    IT WAS REPORTED 2 BD ALARIS SMARTSITE GRAVITY SETS HAD BLOCKAGE ISSUES DURING PRIMING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE SETS WERE OBSERVED TO NOT BE FLOWING DURING PRIMING".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14286043 ↗
  • MalfunctionApr 19, 2022Complete Blockage; Complete Blockage

    IT WAS REPORTED THAT THE BD ALARIS SMARTSITE GRAVITY SET WAS CLOGGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THEY WERE DOUBLE CHECKING THE VALVES IN THE GIVING SET AND FOUND ONE OF THE VALVES IS BACK TO FRONT HENCE THERE IS NO FREE FLOW OF INFUSION FLUID.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14146668 ↗
  • MalfunctionAug 19, 2021Complete Blockage; Restricted Flow rate; Complete Blockage; Restricted Flow rate

    IT WAS REPORTED THAT THE GRAVITY SET W/CV 2SS 20 DP UNV WAS DIFFICULT TO INJECT THROUGH DUE TO FLOW ISSUES/BLOCKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THERE HAVE BEEN SEVERAL ISSUES, IN THEATRE, WHERE SPECIALISTS HAVE BEEN UNABLE TO INJECT THROUGH THE BD SMART SITE CONNECTOR (2000E-04) AND BD SMART SITE GRAVITY SET (42209E). THIS HAS BEEN PREDOMINANTLY WITH THE PREFILLED JUNO SYRINGES CONTAINING EPHEDRINE AND METARAMINOL BUT HAS ALSO OCCURRED WITH BD PLASTIPAK LUER LOK TIP SYRINGES OF ALL SIZES. IT SEEMS THAT THE ISSUES RELATE ONLY TO THE BUNGS AT PRESENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12344659 ↗
  • MalfunctionJul 30, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT GRAVITY SET W/CV 2SS 20 DP UNV WAS DIFFICULT TO CONNECT AND OFTEN DISCONNECTED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN CONNECTING TO THE SMARTSITE IT IS DIFFICULT TO CONNECT AND TAKES SEVERAL ATTEMPTS, OFTEN DISCONNECTING TO FLUSH THE AND 'BREAK THE SEAL' BEFORE RECONNECTING. ISSUE IS INTERMITTENT NOT HAPPENING ALL THE TIME. WHEN CONNECTING THEY ARE DOING AND EXTRA HARD SCREW TO ENSURE THAT THE BLUE CONCERTINA DEPRESSES AND DOES NOT RESIST. THEY FEEL THE ISSUE MAYBE WITH THE BLUE PART OF THE COMPONENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12253221 ↗

Other Set, i.V. Fluid transfer devices

All Set, i.V. Fluid transfer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page