Alaris, SmartSite C24105ERecalls & reported problems

BD Switzerland Sàrl · Pump Infusion Set, Back Check Valve, 5 Y-Sites, 281 cm (111 in) · Class 2

All models of Alaris, SmartSite (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613203019040), received between Jan 24, 2019 and Jan 24, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “No Audible Alarm” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Audible Alarm1

Patient problems

  • Respiratory Distress1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 24, 2019No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED THAT THE DEVICE ALLOWED AIR BUBBLES THROUGH THE AIR-IN-LINE DETECTOR WITHOUT ALARMING DURING A D5%NS INFUSION AT 150ML/HR, AND A SECONDARY ZOFRAN 16MG INFUSION OVER 15 MINUTES. THE AIR-IN-LINE DETECTION WAS SET TO 75 MICROLITERS AND THE ACCUMULATED AIR SETTINGS WERE ENABLED. ALTHOUGH THERE WAS NO LASTING INJURY, THE PATIENT EXPERIENCED SHORTNESS OF BREATH AND THE NURSE PRACTITIONER WAS NOTIFIED. THE DEVICE WAS SEQUESTERED WITH THE IV ADMINISTRATION SET DISCONNECTED. THE BIOMED DEPARTMENT TESTED THE DEVICE ON 12/21/2018 AND FOUND NO DISCREPANCIES USING THE CALIBRATED ADMINISTRATION SET AND A CALIBRATED DRYSET (ASM CHECK-IN TEST). THEY ALSO PERFORMED AN AIR BUBBLE DETECTION TEST AND NO AIR WAS NOTED IN THE LINE. THE CUSTOMER WOULD LIKE SECOND OPINION REGARDING THE ADMINISTRATION SET THAT WAS USED ON THE PATIENT BECAUSE THERE WERE 2 LINES CONNECTED TO IT, AS WELL AS THE LVP UNIT.

    Patient problems: Respiratory Distress

    MAUDE report 8274213 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →