Alaris, SmartSite 11171447Recalls & reported problems

BD Switzerland Sàrl · Pump Infusion Set, 2 Back Check Valves, 3 Y-Sites 320 cm (126 in) · Class 2

All models of Alaris, SmartSite (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This BD Alaris SmartSite pump infusion set is tubing used with infusion pumps. Of its 78 reports, 74 are malfunctions and 4 are injuries, with no deaths. Many describe leaks and connections that come apart, and some describe backflow. There are 4 recalls: 3 open, on pump performance and a labeling error, and 1 closed, on incomplete occlusion. Reports fell to 8 in the last 12 months from 24 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
77
Deaths reported
0
Injuries reported
3
Recalls naming it
4

Summary

The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 07613203021234), received between Apr 25, 2018 and Sep 28, 2026 — classified as 3 injuries, 74 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (27 reports).

4 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 3 reportsApr 2025: 2 reportsMay 2025: 5 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 19, 2026Class IOpen, Classified

    Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance.

    Z-3080-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 8, 2025Class IOpen, Classified

    Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

    Z-2180-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Oct 11, 2023Class IIOpen, Classified

    Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).

    Z-0349-2024 · Root cause: Error in labeling · Recall details · FDA record ↗

  4. May 6, 2019Class ITerminated

    An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running status or can result in faster than expected drug delivery flow when the pump is infusing.

    Z-1770-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash27
  • Fluid/Blood Leak21
  • Connection Problem10
  • Backflow7
  • Complete Blockage4
  • Air/Gas in Device3
  • Defective Component3
  • Disconnection3

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • No Consequences Or Impact To Patient19
  • No Known Impact Or Consequence To Patient5
  • No Information3
  • Blood Loss1
  • Insufficient Information1
  • Underdose1

Recent adverse event reports

The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Air/Gas in Device

    It was reported that BD alaris smartsite pump infusion set had air in line.The following information was received by the initial reporter with the following Verbatim.It was reported by customer that failure to alarm for air-in-line.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917203 ↗
  • MalfunctionMar 19, 2026Activation, Positioning or Separation Problem

    IT WAS REPORTED BY THE CUSTOMER THAT AN ALARIS SMARTSITE BONDED SITE CAME DISCONNECTED; DISTAL END OF TUBING JUST PRIOR TO END OF LINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24641575 ↗
  • MalfunctionFeb 2, 2026Material Split, Cut or Torn

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24240335 ↗
  • MalfunctionDec 30, 2025Leak/Splash; Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23927657 ↗
  • MalfunctionDec 17, 2025Complete Blockage; Complete Blockage

    IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET HAD FLOW ISSUES. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT THEY ARE UNABLE TO INJECT THROUGH PORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23835087 ↗
  • MalfunctionNov 23, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ONCE 1L BAG OF FLUID WAS SPIKED, THE LINE STARTED LEAKING AT THE BOTTOM OF THE DRIP CHAMBER. DESCRIPTION OF PRODUCT: SET ALARIS 126IN 20 DR 3-Y 2 LUER ST 20EA/CA. LOT OR S/N: (B)(6). INCIDENT DATE: UNKNOWN. COMPLAINT NOTICED: DURING / AFTER USE. PROBLEM FREQUENCY: 1ST TIME. PATIENT INJURY: NO. QTY AFFECTED: 1 EA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23627954 ↗
  • MalfunctionNov 21, 2025Complete Blockage; Complete Blockage

    IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS OCCLUDED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT THEY UNABLE TO INJECT MEDICATION THROUGH PORT..

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23619317 ↗
  • MalfunctionAug 4, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT LEAKING FROM THE PLIABLE TUBING PORTION THAT GOES INTO THE ALARIS PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22707240 ↗
  • MalfunctionAug 4, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING . IT WAS REPORTED BY CUSTOMER THAT LEAKING OCCURRED AT TEXIUM/SMARTSITE PORT CONNECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22701771 ↗
  • MalfunctionJul 28, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKING OCCURRED AT TEXIUM/SMARTSITE PORT CONNECTION. PATIENT AND EMPLOYEES WERE NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22628452 ↗
  • MalfunctionJul 8, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING. IT WAS REPORTED BY CUSTOMER THAT CHEMO FOUND DRIPPING FROM CONNECTION IN THE IV TUBING. ABLE TO RESECURE AND CONTINUE THE INFUSION. THE SECOND EVENT THE NEXT MORNING AT 0800. WITH THIS MEDICATION, AIR BUBBLES ARE VERY COMMON, SO THEY OFTEN HAVE TO REMOVE THE TUBING FROM THE PUMP AND FLICK IT TO MOVE THE BUBBLES UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22435391 ↗
  • MalfunctionJul 8, 2025Defective Component; Defective Component

    IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT CHEMO FOUND DRIPPING FROM CONNECTION IN THE IV TUBING. ABLE TO RESECURE AND CONTINUE THE INFUSION. THE SECOND EVENT THE NEXT MORNING AT 0800. WITH THIS MEDICATION, AIR BUBBLES ARE VERY COMMON, SO THEY OFTEN HAVE TO REMOVE THE TUBING FROM THE PUMP AND FLICK IT TO MOVE THE BUBBLES UP

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22435389 ↗
  • MalfunctionJun 30, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD ALARIS PUMP MODULE SMARTSITE INFUSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKING OCCURRED AT TEXIUM/SMARTSITE PORT CONNECTION. PATIENT AND EMPLOYEES WERE NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22364973 ↗
  • MalfunctionJun 3, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT CHEMO RN CAME TO THE PHARMACY STATING WHEN INFUSION WAS STARTED ON HER KEYTRUDA 400MG PIGGYBACK DROPS OF FLUID BEGAN TO LEAK WHERE TEXIUM MET PATIENT LINE (ATTACHED WITH ARROW POINTING TO AREA OF INTEREST).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22133959 ↗
  • MalfunctionMay 23, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CHEMO RN CAME TO THE PHARMACY STATING KEYTRUDA IV PIGGYBACK LEAKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22081616 ↗
  • MalfunctionMay 17, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE INFUSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT WHILE PRIMING THE LINE THE NURSE NOTED THAT THERE WAS QUITE A BIT OF LEAKAGE FROM THE SITE WHERE THE SECONDARY TUBING MEETS THE PRIMARY TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22041242 ↗
  • MalfunctionMay 13, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING SECONDARY SET WAS LOOSE THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. CHEMO RN CAME TO THE PHARMACY STATING WHEN INFUSION WAS STARTED ON HER ZIRABEV PIGGYBACK DROPS OF FLUID BEGAN TO LEAK WHERE TEXIUM MET PATIENT LINE (ATTACHED WITH ARROW POINTING TO AREA OF INTEREST). RN STOPPED INFUSION AND TRIED RE-TIGHTENING CONNECTION. WHEN INFUSION WAS STARTED AGAIN FLUID BEGAN TO LEAK FROM CONNECTION SEAM OF TEXIUM AND PATIENT LINE AGAIN. MEDICATION, LINES, AND ORIGINAL PACKAGING WERE PLACED INTO CHEMOTHERAPY HAZARD BAG AND SENT TO THE PHARMACY. PHARMACY TECHS SUPPLIED ORIGINAL PACKAGING USED FOR CHEMO PIGGY BACK AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22012830 ↗
  • MalfunctionMay 6, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT THE LEAKAGE AT THE TEXIUM TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21972979 ↗
  • MalfunctionMay 6, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT THE LEAKAGE IS AT THE TEXIUM TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21972962 ↗
  • MalfunctionApr 23, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT NOTICED SEVERAL IV TUBING SETS LEAKING BELOW THE DRIP CAMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21896596 ↗
  • MalfunctionApr 4, 2025Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    THE PATIENT RECEIVED APPROX.. 300ML OF FLUID (3KG PATIENT), RESULTING IN FLUID OVERLOAD. THE PATIENT REQUIRED ADDITIONAL MEDICATIONS AND SUPPORTS TO MANAGE THE FLUID. IT IS UNKNOWN WHAT THE LONG-TERM EFFECTS ARE FOR THE PATIENT AS THE PATIENT HAS OTHER MEDICAL COMPLEXITIES THAT THEY ARE ALSO MANAGING. NO ADDITIONAL INFORMATION PROVIDED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21775384 ↗
  • MalfunctionMar 21, 2025Disconnection; Disconnection

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21668201 ↗
  • MalfunctionMar 17, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM (B)(6) 2025 PATIENT HERE FOR TAXOL, HERCEPTIN, 3RD TREATMENT. PIV WAS PLACED IN PATIENTS L HAND. PAPER TAPE WAS APPLIED TO THE CSTD AND AT THE PIV SITE. ABOUT HALF WAY THROUGH THE 3-HOUR TAXOL, PATIENT HAD TO USE THE BATHROOM. WHEN PATIENT CAME BACK FROM THE BATHROOM, BEFORE SITTING DOWN, THE RN CHECKED THE PIV SITE AND CONNECTION. EVERYTHING LOOKED GOOD. RIGHT AFTER THE PATIENT SAT DOWN, THE RN NOTED DROPS OF FLUID COMING FROM THE CONNECTION (N35-0) TO REGULAR TUBING. CHEMOTHERAPY DID NOT GET ON THE PATIENT. IT FELL TO THE FLOOR. APPROXIMATELY LESS THAN 5 ML WERE SPILLED. RN NOTED THAT THE TUBING WAS KINKED JUST AFTER THE HARD PLASTIC CONNECTION SITE. RN RECONNECTED THE TUBING AND APPLIED A 3RD PIECE OF TAPE ABOVE, ON THE PATIENT¿S DRESSING TO PREVENT FURTHER KINKING. RN NOTIFIED PHARMACY AND PHYSICIAN. NO NEW ORDERS GIVEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21617545 ↗
  • MalfunctionMar 17, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS KINKED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT THE RN NOTED DROPS OF FLUID COMING FROM THE CONNECTION (N35-0) TO REGULAR TUBING. CHEMOTHERAPY DID NOT GET ON THE PATIENT. IT FELL TO THE FLOOR. APPROXIMATELY LESS THAN 5 ML WERE SPILLED. RN NOTED THAT THE TUBING WAS KINKED JUST AFTER THE HARD PLASTIC CONNECTION SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21617515 ↗
  • MalfunctionFeb 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY THE CUSTOMER THAT "WHILE TITRATING PT'S RITUXIMAB WRITE NOTICED THAT WHEN THE RATE WAS BUMPED UP THE DRUG WAS DRIPPING FROM ALARIS CHANNEL. TUBING SHOWED A PIN HOLE LEAK NEAR CONNECTION FOR PUMP. NOT MUCH DRUG WAS LOST AND LYM TEAM MADE AWARE AND INSTRUCTED TO CHANGE PRIMARY TUBING AND PROCEED WITH TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21413147 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →