Alaris, SmartSite C24103ERecalls & reported problems
BD Switzerland Sàrl · Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 Y-Site, 298 cm (117 in) · Class 2
All models of Alaris, SmartSite (160)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613203021371), received between Sep 20, 2018 and Dec 17, 2019 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Disconnection” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Disconnection6
- Break1
- Crack1
- Fluid/Blood Leak1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient7
- No Code Available1
- No Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 17, 2019Stretched; Stretched
IT WAS REPORTED THAT THERE WAS AN UNSPECIFIED TUBING CONCERN. IT WAS NOT KNOWN IF THE EVENT OCCURRED DURING INFUSION OR PRIMING OF THE IV SET. THE CUSTOMER PROVIDED A PHOTO OF THE IV TUBING. ALTHOUGH REQUESTED, THERE IS NO FURTHER PATIENT OR EVENT INFORMATION AVAILABLE.
Patient problems: No Information
MAUDE report 9480918 ↗MalfunctionApr 4, 2019Disconnection; Disconnection
IT WAS REPORTED THAT AN UNINTENDED DISCONNECTION OF THE PUMP TUBING TO THE CHILD'S IV ACCESS DEVICE OCCURRED DESPITE THOROUGHLY TIGHTENING THE TUBING. THE NURSE STATED THAT THE PERIPHERAL CATHETERS ARE OFTEN PLACED IN THE CREASE OF THE LEG WHERE THE UPPER LEG MEETS THE GROIN/UPPER ABDOMEN. NO PATIENT HARM WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8483518 ↗MalfunctionApr 4, 2019Disconnection; Disconnection
IT WAS REPORTED THAT DURING A CHILD¿S INFUSION WITH TWO MEDICATIONS AND TWO PUMP MODULES, THE SODIUM CHLORIDE SET DISCONNECTED AT THE LUER CONNECTION TO THE MILRINONE SET. THE EVENT SET WAS CONNECTED TO THE MILRINONE SET LUER CLOSEST TO THE PATIENT'S IV CATHETER. THE EVENT SET INFUSED MAINTENANCE FLUID (500 ML BAG OF 0.45% SODIUM CHLORIDE) AT AN UNKNOWN RATE. THE SET ATTACHED TO THE IV CATHETER CONTAINED MILRINONE (100 ML BAG INFUSING AT 1.82 ML/HR). THE CLINICIAN STATED THAT NO INFECTION OR PATIENT HARM OCCURRED AS A RESULT OF THE DISCONNECTION, AND THE SETS WERE SEQUESTERED FOR INVESTIGATION. ALTHOUGH REQUESTED, NO OTHER EVENT OR PATIENT DETAILS WERE PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8483381 ↗MalfunctionApr 4, 2019Disconnection; Disconnection
IT WAS REPORTED THAT DURING A MILRINONE INFUSION (100 ML BAG INFUSING AT 1.82 ML/HR), AN UNINTENDED DISCONNECTION OF THE PUMP TUBING TO THE CHILD'S IV ACCESS DEVICE OCCURRED DESPITE THOROUGHLY TIGHTENING THE TUBING PRIOR TO THE START OF THE INFUSION. THE NURSE STATED THAT THE PERIPHERAL CATHETERS ARE OFTEN PLACED IN THE CREASE OF THE LEG WHERE THE UPPER LEG MEETS THE GROIN/UPPER ABDOMEN. NO PATIENT HARM WAS REPORTED. ALSO CONNECTED TO THE PATIENT AT THE DISTAL Y-SITE CLOSEST TO THE PATIENT'S CATHETER WAS AN ADDITIONAL PRIMARY ADMINISTRATION SET INFUSING A 500 ML BAG OF 0.45% SODIUM CHLORIDE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8483066 ↗MalfunctionApr 4, 2019Disconnection; Disconnection
IT WAS REPORTED THAT AN UNINTENDED DISCONNECTION OF THE PUMP TUBING TO THE CHILD'S IV ACCESS DEVICE OCCURRED DESPITE THOROUGHLY TIGHTENING THE TUBING. THE NURSE STATED THAT THE PERIPHERAL CATHETERS ARE OFTEN PLACED IN THE CREASE OF THE LEG WHERE THE UPPER LEG MEETS THE GROIN/UPPER ABDOMEN. NO PATIENT HARM WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8482918 ↗MalfunctionApr 3, 2019Disconnection; Disconnection
IT WAS REPORTED THAT AN UNINTENDED DISCONNECTION OF THE PUMP TUBING TO THE CHILD'S IV ACCESS DEVICE OCCURRED DESPITE THOROUGHLY TIGHTENING THE TUBING. THE NURSE STATED THAT THE PERIPHERAL CATHETERS ARE OFTEN PLACED IN THE CREASE OF THE LEG WHERE THE UPPER LEG MEETS THE GROIN/UPPER ABDOMEN. NO PATIENT HARM WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8481052 ↗MalfunctionApr 3, 2019Disconnection; Disconnection
THE CUSTOMER REPORTED UNINTENDED DISCONNECTION OCCURRED DESPITE THOROUGHLY TIGHTENING THE TUBING. THE DISCONNECTION WAS BETWEEN THE SECONDARY SET AND THE UPPER Y SITE OF THE PRIMARY TUBING. NO PATIENT HARM WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8480819 ↗MalfunctionOct 24, 2018Crack; Crack
THE CUSTOMER REPORTED THAT THE MALE LUER CRACKED AT THE CONNECTION TO THE MICROCLAVE. THERE WAS NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8001356 ↗InjurySep 20, 2018Break; Fluid/Blood Leak; Break; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE TUBING BROKE OFF AT A CONNECTION POINT AND DRIPPED BLOOD WHILE IT WAS CONNECTED TO A PATIENT FOR NO MORE THAN 24 HOURS. THE PATIENT EXPERIENCED A SMALL AMOUNT OF BLOOD LOSS AS THE LINE BACKED UP ONTO THE PATIENT'S BED. THE PATIENT WAS ALREADY RECEIVING DAILY BLOOD WORK. THE BLOOD WORK WAS DONE 6 HOURS EARLIER THAN ITS USUAL TIME. (B)(6) 06:55 AM HGB: 92, (B)(6) 00:23 AM HGB: 88. THERE WAS NO WRITTEN INDICATION IN THE PATIENT'S CHART THAT THIS DROP IN HGB WAS FROM THE BLOOD LOSS DUE TO THE LINE DISCONNECTION. THE TUBING WAS REPLACED AND NO LASTING HARM WAS REPORTED.
Patient problems: No Code Available
MAUDE report 7894438 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page