MEDPOR 80012Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · MEDPOR¿ SST? EZ? Sphere with Enucleation Implant Inserter · Class 2 · Implant
All models of MEDPOR (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613252088578), received between Mar 30, 2026 and Mar 30, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).
1 other recall cover devices cleared under the same FDA submission (K922489); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Difficult to Setup or Prepare1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Device Difficult to Setup or Prepare
IT WAS REPORTED THAT THE NEEDLES USED TO FIX THE MUSCLES IN PLACE COULD NOT BE INSERTED THROUGH THE DESIGNATED OPENINGS, AS THE OPENINGS WERE TOO NARROW AND THE SUTURE CHANNELS WERE NOT CONTINUOUS. THE SURGEON WAS REQUIRED TO APPLY SIGNIFICANT FORCE TO PENETRATE THE SPECIFIED SUTURE CHANNELS, WHICH RESULTED IN THE NEEDLE DEVIATING LATERALLY DURING THE PROCEDURE AND CAUSING AN INJURY TO THE SURGEON. NO SURGICAL DELAY OR ANY MEDICAL INTERVENTION WERE REPORTED AT INTAKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24727397 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K922489. One submission often covers many models, so check the product and lot details in each record.
- Z-1090-2015 · Dec 17, 2014 · TerminatedMEDPOR Surgical Implant Contoured Two Piece Chin 5 mm; Part Number 86001; Lot Numbers A1405048, A1404026 Product Usage: MEDPOR Implants in block, sheet and pre-formed shapes are i
Other PROSTHESIS, CHIN, INTERNAL devices
- MEDPOR 5444-0-310Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL6 reports
- MEDPOR 5444-0-510Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL4 reports
- MEDPOR 5444-0-410Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL4 reports
- MEDPOR 82007Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL4 reports
- MEDPOR 5444-0-110Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL3 reports
- MEDPOR 5444-4-310Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL3 reports
- Conform™ Extended Anatomical Chin Size: MediumImplantech Associates, Inc. · Prosthesis, Chin, Internal3 reports
- MEDPOR 80010Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page