Stryker Corporation · 360 Ultra Detachable Coil · Class 2 · Implant

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613252192336), received between Jun 27, 2016 and Feb 27, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Premature Activation2
  • Adverse Event Without Identified Device or Use Problem1
  • Migration1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient3
  • Muscle Weakness/Atrophy1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A 3-MONTH FOLLOW UP, THE IMAGING REVEALED A FISHMOUTHING OF THE FLOW DIVERTER STENT AT THE DISTAL END. THE SUBJECT COIL WAS ALSO USED TO TREAT THE INTERNAL CAROTID ARTERY (ICA) ANEURYSM. THE PATIENT PREVIOUSLY EXPERIENCED WEAKNESS IN ONE LOWER LIMB, BUT THE CONDITION HAS IMPROVED AND IS NOW STABLE. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Muscle Weakness/Atrophy

    MAUDE report 21484615 ↗
  • MalfunctionMay 17, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A RIGHT VERTEBRAL ARTERY (VA) V5 DISSECTION ANEURYSM PROCEDURE THE OPERATOR USED A GUIDEWIRE TO BRING A MICROCATHETER INTO THE ANEURYSM. THE ANEURYSM WAS FRAMED WITH A COIL AND A STENT WAS DEPLOYED THROUGH THE MICROCATHETER. AFTER DELIVERING THREE COILS, THE PHYSICIAN DELIVERED THE SUBJECT COIL AS THE FOURTH COIL BUT IT WAS FOUND THAT THE TAIL OF THE DELIVERY WIRE WAS FRACTURED. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16944258 ↗
  • MalfunctionNov 21, 2022Migration; Migration

    IT WAS REPORTED DURING ANEURYSM EMBOLIZATION CASE, USED A MICROCATHETER TO DELIVER THE SUBJECT COIL. THE DETACHMENT DEVICE SHOWED THE SUCCESSFUL DETACHMENT OF THE SUBJECT COIL. THE PHYSICIAN PREPARED TO WITHDRAW THE DELIVERY WIRE. HOWEVER, UNDER THE DIGITAL SUBTRACTION ANGIOGRAPHY (DSA), THE SUBJECT COIL WAS STILL CONNECTED WITH THE MICROCATHETER TIP. THE OPERATOR ATTEMPTED TO SEPARATE THE COIL FROM THE CATHETER BUT WAS UNSUCCESSFUL. FINALLY, HE DECIDED TO WITHDRAW THE MICROCATHETER AND THE SUBJECT COIL OUT OF PATIENT'S BODY TOGETHER AND FOUND THE SUBJECT COIL ATTACHED TO THE TIP OF THE MICROCATHETER. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15832230 ↗
  • MalfunctionApr 18, 2022Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING THE EMBOLIZATION PROCEDURE OF A RIGHT POSTERIOR COMMUNICATING INTRACRANIAL ANEURYSM, PHYSICIAN FAILED TO DETACH THE SUBJECT COIL. WHILE ADVANCEMENT THERE WAS A SPONTANEOUS DETACHMENT OF THE SUBJECT COIL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT. NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14138271 ↗
  • MalfunctionAug 25, 2020Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE COIL (SUBJECT DEVICE) FAILED TO DETACH UPON MULTIPLE DETACHMENT ATTEMPTS. THE PHYSICIAN ATTEMPTED TO REMOVE AND REINSERT THE SUBJECT COIL AND FOUND THE DELIVERY WIRE TO BE KINKED. THE PHYSICIAN ATTEMPTED TO MAKE THE DELIVERY WIRE STRAIGHT BUT IT FRACTURED DURING THE ATTEMPT. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10449065 ↗
  • MalfunctionOct 27, 2016Premature Activation; Premature Activation

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE COIL PREMATURELY DETACHED INSIDE THE MICROCATHETER. THE COIL AND THE MICROCATHETER WERE REMOVED TOGETHER AS ONE UNIT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND NO ADVERSE CONSEQUENCE TO THE PATIENT WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6062647 ↗
  • MalfunctionJun 27, 2016Premature Activation; Premature Activation

    IT WAS REPORTED THAT DURING ADVANCEMENT OF THE COIL THROUGH THE MICROCATHETER, IT DETACHED UNEXPECTEDLY. THE PHYSICIAN REMOVED THE MICROCATHETER ALONG WITH THE COIL STILL INSIDE WITHOUT PATIENT COMPLICATIONS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5752638 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →