NA 7207003000Recalls & reported problems

Stryker Corporation · Sternum Blade Guard · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Sternum Blade Guard is used during surgery to help hold back tissue while a chest incision is worked on. The 29 reports mostly describe the blade tearing the lining tissue of the lung, with skin tears noted in 26 injuries, plus small pieces breaking off or a component coming loose; no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
26
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 07613252204336), received between Aug 12, 2022 and Jul 24, 2024 — classified as 26 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem26
  • Detachment of Device or Device Component1
  • Insufficient Device Problem Information1
  • Material Fragmentation1
  • No Apparent Adverse Event1

Patient problems

  • Skin Tears26
  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 24, 2024Material Fragmentation; Material Fragmentation

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD SMALL PIECES BREAKING OFF. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19822982 ↗
  • MalfunctionJan 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18556987 ↗
  • MalfunctionOct 31, 2023No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT AN INNOMINATE VEIN WAS HOOKED BY THE STERNUM BLADE GUARD DURING A PROCEDURE. THERE WAS NO PATIENT IMPACT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18041557 ↗
  • InjurySep 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEAR WAS TREATED BY WOUND DRAINAGE WITH A TUBE. NO FURTHER ADVERSE CONSEQUENCES REPORTED.

    Patient problems: Skin Tears

    MAUDE report 15463570 ↗
  • InjurySep 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15371584 ↗
  • InjurySep 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15334388 ↗
  • InjuryAug 31, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15328976 ↗
  • InjuryAug 31, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15328907 ↗
  • InjuryAug 31, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15328851 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15325347 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15325346 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15325218 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15325216 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15322865 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15322861 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15322340 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15322187 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15322184 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15322182 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15322041 ↗
  • InjuryAug 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15221913 ↗
  • InjuryAug 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15221911 ↗
  • InjuryAug 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15221902 ↗
  • InjuryAug 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15221900 ↗
  • InjuryAug 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE DEVICE WAS TEARING PLEURA TISSUE ON THE PATIENT DURING A PROCEDURE. THE TEARS WERE TREATED BY WOUND DRAINAGE WITH A TUBE.

    Patient problems: Skin Tears

    MAUDE report 15221895 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →